Effect of MR head coil geometry on deep‐learning‐based MR image reconstruction
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30130 ↗At a glance
Journal article in Magnetic Resonance in Medicine, 2024. Written by Natalia Dubljevic and 4 co-authors. Institution: Department of Biomedical Engineering, University of Calgary, Calgary, Alberta, Canada. Major MEDLINE topics: Deep Learning, Magnetic Resonance Imaging, Image Processing, Computer-Assisted, Signal-To-Noise Ratio. Cited 1 times so far (Crossref); 33 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: Natural Sciences and Engineering Research Council of Canada, Alberta Innovates.
Abstract (original language)
Purpose
To investigate whether parallel imaging‐imposed geometric coil constraints can be relaxed when using a deep learning (DL)‐based image reconstruction method as opposed to a traditional non‐DL method. Theory and Methods Traditional and DL‐based MR image reconstruction approaches operate in fundamentally different ways: Traditional methods solve a system of equations derived from the image data whereas DL methods use data/target pairs to learn a generalizable reconstruction model. Two sets of head coil profiles were evaluated: (1) 8‐channel and (2) 32‐channel geometries. A DL model was compared to conjugate gradient SENSE (CG‐SENSE) and L1‐wavelet compressed sensing (CS) through quantitative metrics and visual assessment as coil overlap was increased.
Results
Results were generally consistent between experiments. As coil overlap increased, there was a significant ( p < 0.001) decrease in performance in most cases for all methods. The decrease was most pronounced for CG‐SENSE, and the DL models significantly outperformed ( p < 0.001) their non‐DL counterparts in all scenarios. CS showed improved robustness to coil overlap and signal‐to‐noise ratio (SNR) versus CG‐SENSE, but had quantitatively and visually poorer reconstructions characterized by blurriness as compared to DL. DL showed virtually no change in performance across SNR and very small changes across coil overlap.
Conclusion
The DL image reconstruction method produced images that were robust to coil overlap and of higher quality than CG‐SENSE and CS. This suggests that geometric coil design constraints can be relaxed when using DL reconstruction methods.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Biomedical Engineering, University of Calgary, Calgary, Alberta, Canada.
- Seaman Family MR Research Centre, Foothills Medical Centre, Calgary, Alberta, Canada.
- Hotchkiss Brain Institute, University of Calgary, Calgary, Alberta, Canada.
- O'Brien Centre for the Health Sciences, Cumming School of Medicine, Calgary, Alberta, Canada.
- Radiology and Clinical Neuroscience, University of Calgary, Calgary, Alberta, Canada.
- Department of Electrical and Software Engineering, University of Calgary, Calgary, Alberta, Canada.
ORCID profiles: Michel Louis Lauzon, Roberto Souza
Funding
- Natural Sciences and Engineering Research Council of Canada
- Alberta Innovates
Cite
APA
Dubljevic, N., Moore, S., Lauzon, M. L., Souza, R., & Frayne, R. (2024). Effect of MR head coil geometry on deep‐learning‐based MR image reconstruction. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30130
Vancouver
Dubljevic N, Moore S, Lauzon ML, Souza R, Frayne R. Effect of MR head coil geometry on deep‐learning‐based MR image reconstruction. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30130
BibTeX
@article{Dubljevic2024,
author = {Dubljevic, Natalia and Moore, Stephen and Lauzon, Michel Louis and Souza, Roberto and Frayne, Richard},
title = {Effect of MR head coil geometry on deep‐learning‐based MR image reconstruction},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30130}
}
EUVIMED