Spatiotemporal encoding MRI in a portable low‐field system
Magnetic Resonance in Medicine · 2024-09 · 92 (3) · 1011-1021
DOI: 10.1002/mrm.30104 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 16.04.2024. Verfasst von Yueqi Qiu und 7 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, People's Republic of China. Hauptthemen laut MEDLINE: Brain, Magnetic Resonance Imaging, Imaging, Three-Dimensional. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 51 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Israel Science Foundation, Shanghai Science and Technology Development Funds.
Abstract (Originalsprache)
Purpose
Demonstrate the potential of spatiotemporal encoding (SPEN) MRI to deliver largely undistorted 2D, 3D, and diffusion weighted images on a 110 mT portable system.
Methods
SPEN's quadratic phase modulation was used to subsample the low‐bandwidth dimension of echo planar acquisitions, delivering alias‐free images with an enhanced immunity to image distortions in a laboratory‐built, low‐field, portable MRI system lacking multiple receivers.
Results
Healthy brain images with different SPEN time‐bandwidth products and subsampling factors were collected. These compared favorably to EPI acquisitions including topup corrections. Robust 3D and diffusion weighted SPEN images of diagnostic value were demonstrated, with 2.5 mm isotropic resolutions achieved in 3 min scans. This performance took advantage of the low specific absorption rate and relative long TEs associated with low‐field MRI.
Conclusion
SPEN MRI provides a robust and advantageous fast acquisition approach to obtain faithful 3D images and DWI data in low‐cost, portable, low‐field systems without parallel acceleration.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, People's Republic of China.
- National Engineering Research Center of Advanced Magnetic Resonance Technologies for Diagnosis and Therapy (NERC-AMRT), Shanghai Jiao Tong University, Shanghai, People's Republic of China.
- Institute of Medical Robotics, Shanghai Jiao Tong University, Shanghai, People's Republic of China.
- Department of Chemical and Biological Physics, Weizmann Institute of Science, Rehovot, Israel.
ORCID-Profile: Yueqi Qiu, Ke Dai, Sijie Zhong, Suen Chen, Changyue Wang, Hao Chen, Lucio Frydman, Zhiyong Zhang
Förderung
- National Natural Science Foundation of China
- Israel Science Foundation
- Shanghai Science and Technology Development Funds
- Startup Fund for Young Faculty at SJTU (SFYF at SJTU)
- Shanghai Science and Technology Development Fund
Zitieren
APA
Qiu, Y., Dai, K., Zhong, S., Chen, S., Wang, C., Chen, H., Frydman, L., & Zhang, Z. (2024). Spatiotemporal encoding MRI in a portable low‐field system. Magnetic Resonance in Medicine, 92(3), 1011-1021. https://doi.org/10.1002/mrm.30104
Vancouver
Qiu Y, Dai K, Zhong S, Chen S, Wang C, Chen H, et al. Spatiotemporal encoding MRI in a portable low‐field system. Magnetic Resonance in Medicine. 2024;92(3):1011-1021. doi:10.1002/mrm.30104
BibTeX
@article{Qiu2024,
author = {Qiu, Yueqi and Dai, Ke and Zhong, Sijie and Chen, Suen and Wang, Changyue and Chen, Hao and Frydman, Lucio and Zhang, Zhiyong},
title = {Spatiotemporal encoding MRI in a portable low‐field system},
journal = {Magnetic Resonance in Medicine},
year = {2024},
volume = {92},
number = {3},
pages = {1011--1021},
doi = {10.1002/mrm.30104}
}