MR signature matching ( MRSIGMA ) implementation for true real‐time free‐breathing volumetric imaging with sub‐200 ms latency on an MR‐Linac

Saad Siddiq iD, Victor Murray iD, Neelam Tyagi, Pim Borman, Chengcheng Gui, Christopher Crane, Can Wu iD, Ricardo Otazo iD

Magnetic Resonance in Medicine · 2024

DOI: 10.1002/mrm.30097 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Saad Siddiq und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Magnetic Resonance Imaging, Imaging, Three-Dimensional, Respiration. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS, National Institutes of Health/National Cancer Institute.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Develop a true real‐time implementation of MR signature matching (MRSIGMA) for free‐breathing 3D MRI with sub‐200 ms latency on the Elekta Unity 1.5T MR‐Linac.

Methods

MRSIGMA was implemented on an external computer with a network connection to the MR‐Linac. Stack‐of‐stars with partial k z sampling was used to accelerate data acquisition and ReconSocket was employed for simultaneous data transmission. Movienet network computed the 4D MRI motion dictionary and correlation analysis was used for signature matching. A programmable 4D MRI phantom was utilized to evaluate MRSIGMA with respect to a ground‐truth translational motion reference. In vivo validation was performed on patients with pancreatic cancer, where 15 patients were employed to train Movienet and 7 patients to test the real‐time implementation of MRSIGMA. Dice coefficients between real‐time MRSIGMA and a retrospectively computed 4D reference were used to evaluate motion tracking performance.

Results

Motion dictionary was computed in under 5 s. Signature acquisition and matching presented 173 ms latency on the phantom and 193 ms on patients. MRSIGMA presented a mean error of 1.3–1.6 mm for all phantom experiments, which was below the 2 mm acquisition resolution along the motion direction. The Dice coefficient over time between MRSIGMA and reference contours was 0.88 ± 0.02 (GTV), 0.87 ± 0.02(duodenum‐stomach), and 0.78 ± 0.02(small bowel), demonstrating high motion tracking performance for both tumor and organs at risk.

Conclusion

The real‐time implementation of MRSIGMA enabled true real‐time free‐breathing 3D MRI with sub‐200 ms imaging latency on a clinical MR‐Linac system, which can be used for treatment monitoring, adaptive radiotherapy and dose accumulation mapping in tumors affected by respiratory motion.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingMagnetic Resonance ImagingImaging, Three-DimensionalRespirationAlgorithmsPancreatic NeoplasmsHumansMotionImage Processing, Computer-AssistedRetrospective StudiesImage Interpretation, Computer-Assisted

Stichwörter der Autor:innen

4D MRIMRI‐guided radiotherapyMR‐Linaclow‐latencymotion trackingreal‐time

Autor:innen und Institutionen

  • Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York, USA.
  • Department of Radiotherapy, University Medical Center Utrecht, Utrecht, The Netherlands.
  • Department of Radiation Oncology, Memorial Sloan Kettering Cancer Center, New York, New York, USA.
  • Department of Radiology, Memorial Sloan Kettering Cancer Center, New York, New York, USA.

ORCID-Profile: Saad Siddiq, Victor Murray, Can Wu, Ricardo Otazo

Förderung

  • NCI NIH HHS
  • National Institutes of Health/National Cancer Institute

Zitieren

APA

Siddiq, S., Murray, V., Tyagi, N., Borman, P., Gui, C., Crane, C., Wu, C., & Otazo, R. (2024). MR signature matching ( MRSIGMA ) implementation for true real‐time free‐breathing volumetric imaging with sub‐200 ms latency on an MR‐Linac. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30097

Vancouver

Siddiq S, Murray V, Tyagi N, Borman P, Gui C, Crane C, et al. MR signature matching ( MRSIGMA ) implementation for true real‐time free‐breathing volumetric imaging with sub‐200 ms latency on an MR‐Linac. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30097

BibTeX
@article{Siddiq2024,
  author = {Siddiq, Saad and Murray, Victor and Tyagi, Neelam and Borman, Pim and Gui, Chengcheng and Crane, Christopher and Wu, Can and Otazo, Ricardo},
  title = {MR signature matching ( MRSIGMA ) implementation for true real‐time free‐breathing volumetric imaging with sub‐200 ms latency on an MR‐Linac},
  journal = {Magnetic Resonance in Medicine},
  year = {2024},
  doi = {10.1002/mrm.30097}
}
Kostenlos starten →