Effect of inhaled oxygen level on dynamic glucose‐enhanced MRI in mouse brain
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30035 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Jianpan Huang und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic Radiology, The University of Hong Kong, Hong Kong, China. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Glucose, Oxygen, Brain. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 73 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIBIB NIH HHS, NIH HHS, University of Hong Kong.
Abstract (Originalsprache)
Purpose
To investigate the effect of inhaled oxygen level on dynamic glucose enhanced (DGE) MRI in mouse brain tissue and CSF at 3 T.
Methods
DGE data of brain tissue and CSF from mice under normoxia or hyperoxia were acquired in independent and interleaved experiments using on‐resonance variable delay multi‐pulse (onVDMP) MRI. A bolus of 0.15 mL filtered 50% D‐glucose was injected through the tail vein over 1 min during DGE acquisition. MRS was acquired before and after DGE experiments to confirm the presence of D‐glucose.
Results
A significantly higher DGE effect under normoxia than under hyperoxia was observed in brain tissue ( p = 0.0001 and p = 0.0002 for independent and interleaved experiments, respectively), but not in CSF ( p > 0.3). This difference is attributed to the increased baseline MR tissue signal under hyperoxia induced by a shortened T 1 and an increased BOLD effect. When switching from hyperoxia to normoxia without glucose injection, a signal change of ˜3.0% was found in brain tissue and a signal change of ˜1.5% was found in CSF.
Conclusions
DGE signal was significantly lower under hyperoxia than that under normoxia in brain tissue, but not in CSF. The reason is that DGE effect size of brain tissue is affected by the baseline signal, which could be influenced by T 1 change and BOLD effect. Therefore, DGE experiments in which the oxygenation level is changed from baseline need to be interpreted carefully.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Diagnostic Radiology, The University of Hong Kong, Hong Kong, China.
- Department of Biomedical Engineering, City University of Hong Kong, Hong Kong, China.
- F.M. Kirby Research Center for Functional Brain Imaging, Kennedy Krieger Research Institute, Baltimore, Maryland, USA.
- Russell H. Morgan Department of Radiology and Radiological Science, The Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
- Hong Kong Centre for Cerebro-Cardiovascular Health Engineering (COCHE), Hong Kong, China.
- City University of Hong Kong Shenzhen Research Institute, Shenzhen, China.
ORCID-Profile: Jianpan Huang, Kannie W Y Chan
Förderung
- NIBIB NIH HHS
- NIH HHS
- University of Hong Kong
- National Natural Science Foundation of China
- City University of Hong Kong
- National Institutes of Health
- Research Grants Council, University Grants Committee
Zitieren
APA
Huang, J., Chen, Z., van Zijl, P. C. M., Hin Law, L., Pemmasani Prabakaran, R. S., Park, S. W., Xu, J., & Chan, K. W. Y. (2024). Effect of inhaled oxygen level on dynamic glucose‐enhanced MRI in mouse brain. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30035
Vancouver
Huang J, Chen Z, van Zijl PCM, Hin Law L, Pemmasani Prabakaran RS, Park SW, et al. Effect of inhaled oxygen level on dynamic glucose‐enhanced MRI in mouse brain. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30035
BibTeX
@article{Huang2024,
author = {Huang, Jianpan and Chen, Zilin and van Zijl, Peter C M and Hin Law, Lok and Pemmasani Prabakaran, Rohith Saai and Park, Se Weon and Xu, Jiadi and Chan, Kannie W Y},
title = {Effect of inhaled oxygen level on dynamic glucose‐enhanced MRI in mouse brain},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30035}
}
EUVIMED