Interventional device tracking under MRI via alternating current controlled inhomogeneities
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30031 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Dogangun Uzun und 9 weiteren Autor:innen. Beteiligte Einrichtung: Cardiovascular Branch, Division of Intramural Research, National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, Maryland,…. Hauptthemen laut MEDLINE: Phantoms, Imaging, Magnetic Resonance Imaging, Equipment Design. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Intramural NIH HHS.
Abstract (Originalsprache)
Purpose
To introduce alternating current‐controlled, conductive ink‐printed marker that could be implemented with both custom and commercial interventional devices for device tracking under MRI using gradient echo, balanced SSFP, and turbo spin‐echo sequences.
Methods
Tracking markers were designed as solenoid coils and printed on heat shrink tubes using conductive ink. These markers were then placed on three MR‐compatible test samples that are typically challenging to visualize during MRI scans. MRI visibility of markers was tested by applying alternating and direct current to the markers, and the effects of applied current parameters (amplitude, frequency) on marker artifacts were tested for three sequences (gradient echo, turbo spin echo, and balanced SSFP) in a gel phantom, using 0.55T and 1.5T MRI scanners. Furthermore, an MR‐compatible current supply circuit was designed, and the performance of the current‐controlled markers was tested in one postmortem animal experiment using the current supply circuit.
Results
Direction and parameters of the applied current were determined to provide the highest conspicuity for all three sequences. Marker artifact size was controlled by adjusting the current amplitude, successfully. Visibility of a custom‐designed, 20‐gauge nitinol needle was increased in both in vitro and postmortem animal experiments using the current supply circuit.
Conclusion
Current‐controlled conductive ink‐printed markers can be placed on custom or commercial MR‐compatible interventional tools and can provide an easy and effective solution to device tracking under MRI for three sequences by adjusting the applied current parameters with respect to pulse sequence parameters using the current supply circuit.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiovascular Branch, Division of Intramural Research, National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, Maryland, USA.
- Institute of Biomedical Engineering, Bogazici University, Istanbul, Turkey.
ORCID-Profile: Dogangun Uzun, Dursun Korel Yildirim, Christopher G Bruce, Adrienne E Campbell-Washburn
Förderung
- Intramural NIH HHS
Zitieren
APA
Uzun, D., Yildirim, D. K., Bruce, C. G., Halaby, R. N., Jaimes, A. E., Potersnak, A., Ramasawmy, R., Campbell-Washburn, A. E., Lederman, R. J., & Kocaturk, O. (2024). Interventional device tracking under MRI via alternating current controlled inhomogeneities. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30031
Vancouver
Uzun D, Yildirim DK, Bruce CG, Halaby RN, Jaimes AE, Potersnak A, et al. Interventional device tracking under MRI via alternating current controlled inhomogeneities. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30031
BibTeX
@article{Uzun2024,
author = {Uzun, Dogangun and Yildirim, Dursun Korel and Bruce, Christopher G and Halaby, Rim N and Jaimes, Andrea E and Potersnak, Amanda and Ramasawmy, Rajiv and Campbell-Washburn, Adrienne E and Lederman, Robert J and Kocaturk, Ozgur},
title = {Interventional device tracking under MRI via alternating current controlled inhomogeneities},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30031}
}
EUVIMED