EUVIMEDEUVIMED entdecken
Crossref

Unsupervised motion artifact correction of turbo spin‐echo MRI using deep image prior

Jongyeon Lee iD, Hyunseok Seo iD, Wonil Lee iD, HyunWook Park iD

Magnetic Resonance in Medicine · 2024

DOI: 10.1002/mrm.30026 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Jongyeon Lee und 3 weiteren Autor:innen. Beteiligte Einrichtung: School of Electrical Engineering, Korea Advanced Institute of Science and Technology (KAIST), Daejeon, Republic of Korea. Hauptthemen laut MEDLINE: Artifacts, Magnetic Resonance Imaging, Image Processing, Computer-Assisted, Deep Learning. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 39 Literaturangaben. Gefördert von: Korea Health Industry Development Institute, Korea Institute of Science and Technology.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

In MRI, motion artifacts can significantly degrade image quality. Motion artifact correction methods using deep neural networks usually required extensive training on large datasets, making them time‐consuming and resource‐intensive. In this paper, an unsupervised deep learning‐based motion artifact correction method for turbo‐spin echo MRI is proposed using the deep image prior framework. Theory and Methods The proposed approach takes advantage of the high impedance to motion artifacts offered by the neural network parameterization to remove motion artifacts in MR images. The framework consists of parameterization of MR image, automatic spatial transformation, and motion simulation model. The proposed method synthesizes motion‐corrupted images from the motion‐corrected images generated by the convolutional neural network, where an optimization process minimizes the objective function between the synthesized images and the acquired images.

Results

In the simulation study of 280 slices from 14 subjects, the proposed method showed a significant increase in the averaged structural similarity index measure by 0.2737 in individual coil images and by 0.4550 in the root‐sum‐of‐square images. In addition, the ablation study demonstrated the effectiveness of each proposed component in correcting motion artifacts compared to the corrected images produced by the baseline method. The experiments on real motion dataset has shown its clinical potential.

Conclusion

The proposed method exhibited significant quantitative and qualitative improvements in correcting rigid and in‐plane motion artifacts in MR images acquired using turbo spin‐echo sequence.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ArtifactsMagnetic Resonance ImagingImage Processing, Computer-AssistedDeep LearningMotionAlgorithmsHumansBrainNeural Networks, ComputerComputer Simulation

Stichwörter der Autor:innen

deep image priordeep learningturbo spin‐echounsupervised motion correction

Autor:innen und Institutionen

  • School of Electrical Engineering, Korea Advanced Institute of Science and Technology (KAIST), Daejeon, Republic of Korea.
  • Bionics Research Center, Biomedical Research Division, Korea Institute of Science and Technology (KIST), Seoul, Republic of Korea.
  • Gordon Center for Medical Imaging, Department of Radiology, Massachusetts General Hospital, Boston, Massachusetts, USA.
  • Department of Radiology, Harvard Medical School, Boston, Massachusetts, USA.

ORCID-Profile: Jongyeon Lee, Hyunseok Seo, Wonil Lee, HyunWook Park

Förderung

  • Korea Health Industry Development Institute
  • Korea Institute of Science and Technology

Zitieren

APA

Lee, J., Seo, H., Lee, W., & Park, H. (2024). Unsupervised motion artifact correction of turbo spin‐echo MRI using deep image prior. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30026

Vancouver

Lee J, Seo H, Lee W, Park H. Unsupervised motion artifact correction of turbo spin‐echo MRI using deep image prior. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30026

BibTeX
@article{Lee2024,
  author = {Lee, Jongyeon and Seo, Hyunseok and Lee, Wonil and Park, HyunWook},
  title = {Unsupervised motion artifact correction of turbo spin‐echo MRI using deep image prior},
  journal = {Magnetic Resonance in Medicine},
  year = {2024},
  doi = {10.1002/mrm.30026}
}
Kostenlos starten →