Fast high‐resolution prospective motion correction for single‐voxel spectroscopy
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.29950 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Isaac M Adanyeguh und 3 weiteren Autor:innen. Beteiligte Einrichtung: Center for Magnetic Resonance Research and Department of Radiology, University of Minnesota Medical School, Minneapolis, Minnesota, USA. Hauptthemen laut MEDLINE: Brain, Artifacts. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIBIB NIH HHS, NIH HHS, NINDS NIH HHS.
Abstract (Originalsprache)
Purpose
To develop a fast high‐resolution image‐based motion correction method using spiral navigators with multislice‐to‐volume registration.
Methods
A semi‐LASER sequence was modified to include a multislice spiral navigator for prospective motion correction (∼305 ms including acquisition, processing, and feedback) as well as shim and frequency navigators for prospective shim and frequency correction (∼100 ms for each). MR spectra were obtained in the prefrontal cortex in five healthy subjects at 3 T with and without prospective motion and shim correction. The effect of key navigator parameters (number of slices, image resolution, and excitation flip angle) on registration accuracy was assessed using simulations.
Results
Without prospective motion and shim correction, spectral quality degraded significantly in the presence of voluntary motion. In contrast, with prospective motion and shim correction, spectral quality was improved (metabolite linewidth = 6.7 ± 0.6 Hz, SNR= 67 ± 9) and in good agreement with baseline data without motion (metabolite linewidth = 6.9 ± 0.9 Hz, SNR = 73 ± 9). In addition, there was no significant difference in metabolites concentrations measured without motion and with prospective motion and shim correction in the presence of motion. Simulations showed that the registration precision was comparable when using three navigator slices with 3 mm resolution and when using the entire volume (all slices) with 8 mm resolution.
Conclusion
The proposed motion correction scheme allows fast and precise prospective motion and shim correction for single‐voxel spectroscopy at 3 T. With 3 mm resolution, only a few navigator slices are necessary to achieve excellent motion correction performance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Center for Magnetic Resonance Research and Department of Radiology, University of Minnesota Medical School, Minneapolis, Minnesota, USA.
ORCID-Profile: Isaac M Adanyeguh, Dinesh K Deelchand
Förderung
- NIBIB NIH HHS
- NIH HHS
- NINDS NIH HHS
- NCI NIH HHS
- National Institutes of Health
Zitieren
APA
Adanyeguh, I. M., Bikkamane Jayadev, N., Henry, P. G., & Deelchand, D. K. (2024). Fast high‐resolution prospective motion correction for single‐voxel spectroscopy. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.29950
Vancouver
Adanyeguh IM, Bikkamane Jayadev N, Henry PG, Deelchand DK. Fast high‐resolution prospective motion correction for single‐voxel spectroscopy. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.29950
BibTeX
@article{Adanyeguh2024,
author = {Adanyeguh, Isaac M and Bikkamane Jayadev, Nutandev and Henry, Pierre-Gilles and Deelchand, Dinesh K},
title = {Fast high‐resolution prospective motion correction for single‐voxel spectroscopy},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.29950}
}
EUVIMED