Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2023. Verfasst von Inbal Beracha und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Chemical and Biological Physics, Weizmann Institute of Science, Rehovot, Israel. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 71 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Israel Science Foundation.
Abstract (Originalsprache)
Purpose
Conventional sequences are static in nature, fixing measurement parameters in advance in anticipation of a wide range of expected tissue parameter values. We set out to design and benchmark a new, personalized approach—termed adaptive MR —in which incoming subject data is used to update and fine‐tune the pulse sequence parameters in real time.
Methods
We implemented an adaptive, real‐time multi‐echo (MTE) experiment for estimating T 2 s. Our approach combined a Bayesian framework with model‐based reconstruction. It maintained and continuously updated a prior distribution of the desired tissue parameters, including T 2 , which was used to guide the selection of sequence parameters in real time.
Results
Computer simulations predicted accelerations between 1.7‐ and 3.3‐fold for adaptive multi‐echo sequences relative to static ones. These predictions were corroborated in phantom experiments. In healthy volunteers, our adaptive framework accelerated the measurement of T 2 for n‐acetyl‐aspartate by a factor of 2.5.
Conclusion
Adaptive pulse sequences that alter their excitations in real time could provide substantial reductions in acquisition times. Given the generality of our proposed framework, our results motivate further research into other adaptive model‐based approaches to MRI and MRS.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Chemical and Biological Physics, Weizmann Institute of Science, Rehovot, Israel.
- Siemens Healthcare Ltd., Rosh Ha'ayeen, Israel.
ORCID-Profile: Assaf Tal
Förderung
- Israel Science Foundation
Zitieren
APA
Beracha, I., Seginer, A., & Tal, A. (2023). Adaptive model‐based Magnetic Resonance. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.29688
Vancouver
Beracha I, Seginer A, Tal A. Adaptive model‐based Magnetic Resonance. Magnetic Resonance in Medicine. 2023. doi:10.1002/mrm.29688
BibTeX
@article{Beracha2023,
author = {Beracha, Inbal and Seginer, Amir and Tal, Assaf},
title = {Adaptive model‐based Magnetic Resonance},
journal = {Magnetic Resonance in Medicine},
year = {2023},
doi = {10.1002/mrm.29688}
}
EUVIMED