EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Open‐source myocardial T 1 mapping with simultaneous multi‐slice acceleration: Combining an auto‐calibrated blipped‐bSSFP readout with VERSE‐MB pulses

Andreia S Gaspar iD, Nuno A Silva, Anthony N Price iD, António M Ferreira, Rita G Nunes iD

Magnetic Resonance in Medicine · 2023

DOI: 10.1002/mrm.29661 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2023. Verfasst von Andreia S Gaspar und 4 weiteren Autor:innen. Beteiligte Einrichtung: Instituto de Sistemas e Robótica - Lisboa and Departamento de Bioengenharia, Instituto Superior Técnico, Universidade de Lisboa, Lisbon, Portugal. Hauptthemen laut MEDLINE: Image Interpretation, Computer-Assisted, Myocardium. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Fundação para a Ciência e a Tecnologia.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Enabling fast and accessible myocardial T 1 mapping is crucial for extending its clinical application. We introduce Open‐MOLLI‐SMS combining simultaneous multi‐slice (SMS) with auto‐calibration and variable‐rate selective excitation (VERSE)‐multiband pulses to obtain all slices in a fast single‐shot T 1 mapping sequence.

Methods

Open‐MOLLI‐SMS was developed by integrating SMS with the open‐source method Open‐MOLLI previously implemented in Pulseq. Three methods were integrated for Open‐MOLLI‐SMS: (1) auto‐calibration blip patterns to ensure consistency between the data and coil information; (2) a blipped‐balanced SSFP (bSSFP) readout to induce controlled aliasing in parallel imaging shifts without disturbing the bSSFP frequency response; and (3) a VERSE‐multiband pulse for minimizing the achievable TR and the specific absortion rate (SAR) impact of SMS. Two (SMS2) or three (SMS3) slices were excited simultaneously and encoded with an in‐plane acceleration factor of 2. Experiments were performed in the International Society for Magnetic Resonance in Medicine/National Institute of Standards and Technology phantom and five healthy volunteers.

Results

Phantom results show accurate T 1 estimates for reference values between 400 to 2200 ms. Artifacts were visible for Open‐MOLLI‐SMS3 but not replicated in vivo. In vivo Open‐MOLLI‐SMS (T 1 SMS2 = 993 ± 10 ms; T 1 SMS3 = 1031 ± 17 ms) provided similar values to mean T 1 single‐band Open‐MOLLI estimates (T 1 Open‐MOLLI = 1005 ± 47 ms). Open‐MOLLI‐SMS2 provided the closest estimates to the reference.

Conclusion

This proof‐of‐principle implementation study demonstrates the feasibility of speeding up T 1 ‐mapping acquisitions and increasing coverage by combining auto‐calibration strategies with a blipped‐bSFFP readout and VERSE multiband RF excitation pulses. The proposed methodology was built on the Open‐MOLLI mapping sequence, which provides a fast means for prototyping and enables open‐source sharing of the method.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Image Interpretation, Computer-AssistedMyocardiumHumansMagnetic Resonance ImagingMagnetic Resonance SpectroscopyPhantoms, ImagingAccelerationReproducibility of ResultsHeart

Stichwörter der Autor:innen

Open-MOLLIT1 mappingauto-calibrationmyocardiumopen-sourcesimultaneous multi-slice

Autor:innen und Institutionen

  • Instituto de Sistemas e Robótica - Lisboa and Departamento de Bioengenharia, Instituto Superior Técnico, Universidade de Lisboa, Lisbon, Portugal.
  • Hospital da Luz Learning Health, Luz Saúde, Lisbon, Portugal.
  • Medical Physics, Guy's & St Thomas' NHS Foundation Trust, London, UK.
  • Serviço de Cardiologia, Hospital de Santa Cruz, Centro Hospitalar Lisboa Ocidental, Lisbon, Portugal.
  • Unidade de Imagiologia Cardíaca Avançada, Hospital da Luz, Lisbon, Portugal.

ORCID-Profile: Andreia S Gaspar, Anthony N Price, Rita G Nunes

Förderung

  • Fundação para a Ciência e a Tecnologia

Zitieren

APA

Gaspar, A. S., Silva, N. A., Price, A. N., Ferreira, A. M., & Nunes, R. G. (2023). Open‐source myocardial T 1 mapping with simultaneous multi‐slice acceleration: Combining an auto‐calibrated blipped‐bSSFP readout with VERSE‐MB pulses. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.29661

Vancouver

Gaspar AS, Silva NA, Price AN, Ferreira AM, Nunes RG. Open‐source myocardial T 1 mapping with simultaneous multi‐slice acceleration: Combining an auto‐calibrated blipped‐bSSFP readout with VERSE‐MB pulses. Magnetic Resonance in Medicine. 2023. doi:10.1002/mrm.29661

BibTeX
@article{Gaspar2023,
  author = {Gaspar, Andreia S and Silva, Nuno A and Price, Anthony N and Ferreira, António M and Nunes, Rita G},
  title = {Open‐source myocardial T 1 mapping with simultaneous multi‐slice acceleration: Combining an auto‐calibrated blipped‐bSSFP readout with VERSE‐MB pulses},
  journal = {Magnetic Resonance in Medicine},
  year = {2023},
  doi = {10.1002/mrm.29661}
}
Kostenlos starten →