Considerations for hyperpolarized 13 C MR at reduced field: Comparing 1.5T versus 3T

Julia Traechtler iD, Maximilian Fuetterer iD, Mohammed M Albannay iD, Tobias Hoh iD, Sebastian Kozerke iD

Magnetic Resonance in Medicine · 2023

DOI: 10.1002/mrm.29579 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2023. Verfasst von Julia Traechtler und 4 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Enhancement. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

In contrast to conventional MR, signal‐to‐noise ratio (SNR) is not linearly dependent on field strength in hyperpolarized MR, as polarization is generated outside the MR system. Moreover, field inhomogeneity‐induced artifacts and other practical limitations associated with field strengths 3T are alleviated at lower fields. The potential of hyperpolarized C spectroscopy and imaging at 1.5T versus 3T is demonstrated in silico, in vitro, and in vivo for applications on clinical MR systems. Theory and Methods Theoretical noise and SNR behavior at different field strengths are investigated based on simulations. A thorough field comparison between 1.5T and 3T is performed using thermal and hyperpolarized C spectroscopy and imaging. Cardiac in vivo data is obtained in pigs using hyperpolarized [1‐C]pyruvate spectroscopy and imaging at 1.5T and 3T.

Results

Based on theoretical considerations and simulations, the SNR of hyperpolarized MR at identical acquisition bandwidths is independent of the field strength for typical coil setups, while adaptively changing the acquisition bandwidth proportional to the static magnetic field allows for net SNR gains of up to 40% at 1.5T compared to 3T. In vitro C data verified these considerations with less than 7% deviation. In vivo feasibility of hyperpolarized [1‐C]pyruvate dynamic metabolic spectroscopy and imaging at 1.5T is demonstrated in the pig heart with comparable SNR between 1.5T and 3T while B artifacts are noticeably reduced at 1.5T.

Conclusion

Hyperpolarized C MR at lower field strengths is favorable in terms of SNR and off‐resonance effects, which makes 1.5T a promising alternative to 3T, especially for clinical cardiac metabolic imaging.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingImage EnhancementAnimalsSwineMagnetic Resonance SpectroscopySignal-To-Noise RatioMagnetic Fields

Stichwörter der Autor:innen

1.5T versus 3Tfield comparisonfield inhomogeneitieshyperpolarized 13 $$ {}^{13} $$ Csignal-to-noise ratio

Autor:innen und Institutionen

  • Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland.
  • Laboratory of Physical Chemistry, ETH Zurich, Zurich, Switzerland.

ORCID-Profile: Julia Traechtler, Maximilian Fuetterer, Mohammed M Albannay, Tobias Hoh, Sebastian Kozerke

Zitieren

APA

Traechtler, J., Fuetterer, M., Albannay, M. M., Hoh, T., & Kozerke, S. (2023). Considerations for hyperpolarized 13 C MR at reduced field: Comparing 1.5T versus 3T. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.29579

Vancouver

Traechtler J, Fuetterer M, Albannay MM, Hoh T, Kozerke S. Considerations for hyperpolarized 13 C MR at reduced field: Comparing 1.5T versus 3T. Magnetic Resonance in Medicine. 2023. doi:10.1002/mrm.29579

BibTeX
@article{Traechtler2023,
  author = {Traechtler, Julia and Fuetterer, Maximilian and Albannay, Mohammed M and Hoh, Tobias and Kozerke, Sebastian},
  title = {Considerations for hyperpolarized 13 C MR at reduced field: Comparing 1.5T versus 3T},
  journal = {Magnetic Resonance in Medicine},
  year = {2023},
  doi = {10.1002/mrm.29579}
}
Kostenlos starten →