EUVIMEDDescoperă EUVIMED
Crossref

Feasibility of prior‐informed domain‐based electrical impedance tomography for deep tissue thermometry: Phantom experiments and computed tomography‐derived numerical simulations

Jinyoung Hong, Jonghyun Kim, Bumsoo Lee, Yongha Gi, Yunhui Jo, Byung‐Jou Lee, Sung Hwan Ahn, Myonggeun Yoon

Medical Physics · 2026-12 · 53 (12)

DOI: 10.1002/mp.70774 ↗

At a glance

Journal article in Medical Physics, published online on 2026-11-18. Written by Jinyoung Hong and 7 co-authors. 29 references. Funded by: National Research Foundation of Korea, Korea Medical Device Development Fund, Ministry of Trade, Industry and Energy.

RezumatDate bibliograficeCiteRelated studies

Abstract (original language)

Background

Accurate monitoring of deep tissue temperature is important during hyperthermia treatment. Because the electrical conductivity of biological tissue generally increases with temperature, Electrical Impedance Tomography (EIT) may provide a non‐invasive approach for estimating temperature changes in deep tissue.

Purpose

This study aimed to evaluate a domain‐based EIT thermometry framework for quantitative temperature estimation in electrically heterogeneous settings composed of piecewise homogeneous domains, where conventional voxel‐based EIT reconstruction is limited by the ill‐posed inverse problem and inaccurate separation of domain‐level conductivity changes.

Methods

Simulation and phantom studies were performed. In the simulation study, reconstruction performance was evaluated using reconstruction error and stability. In the phantom experiments, performance was evaluated using the mean and maximum absolute temperature errors over reconstructed frames by comparing sensor‐measured temperatures with EIT‐based estimates. An anatomically realistic model was additionally constructed from a single respiratory phase of one patient computed tomography (CT) dataset, and nine anatomical subdomains were reconstructed under target‐centered and external‐entry heating scenarios.

Results

In the simulation study, the proposed domain‐based reconstruction yielded the lowest reconstruction error across the evaluated noise conditions. In phantom experiments, it showed close agreement with direct measurements under all tested conditions. The average conductivity reconstruction error in the homogeneous phantom was 0.29%, while the mean absolute temperature errors in the multi‐domain phantoms ranged from 0.22°C to 0.71°C. In pooled phantom comparisons, the proposed method showed mean and maximum absolute temperature errors of 0.44 ± 0.31°C and 0.91 ± 0.77°C, respectively. In the CT‐derived model, the mean absolute temperature errors were 0.50°C and 1.43°C for the target‐centered and external‐entry scenarios, compared with 1.21–1.25°C and 2.60–2.67°C for the benchmark methods, respectively, and the location of the hottest region was preserved in both scenarios.

Conclusion

The proposed domain‐based EIT thermometry demonstrated feasibility in phantom experiments and a CT‐derived anatomical model, establishing a foundation for its potential application as a non‐invasive temperature monitoring approach.

Funding

  • National Research Foundation of Korea
  • Korea Medical Device Development Fund
  • Ministry of Trade, Industry and Energy
  • Ministry of Food and Drug Safety

Cite

APA

Hong, J., Kim, J., Lee, B., Gi, Y., Jo, Y., Lee, B., Ahn, S. H., & Yoon, M. (2026). Feasibility of prior‐informed domain‐based electrical impedance tomography for deep tissue thermometry: Phantom experiments and computed tomography‐derived numerical simulations. Medical Physics, 53(12). https://doi.org/10.1002/mp.70774

Vancouver

Hong J, Kim J, Lee B, Gi Y, Jo Y, Lee B, et al. Feasibility of prior‐informed domain‐based electrical impedance tomography for deep tissue thermometry: Phantom experiments and computed tomography‐derived numerical simulations. Medical Physics. 2026;53(12). doi:10.1002/mp.70774

BibTeX
@article{Hong2026,
  author = {Hong, Jinyoung and Kim, Jonghyun and Lee, Bumsoo and Gi, Yongha and Jo, Yunhui and Lee, Byung‐Jou and Ahn, Sung Hwan and Yoon, Myonggeun},
  title = {Feasibility of prior‐informed domain‐based electrical impedance tomography for deep tissue thermometry: Phantom experiments and computed tomography‐derived numerical simulations},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {12},
  doi = {10.1002/mp.70774}
}
Start for free →