EUVIMEDDescoperă EUVIMED
Crossref

Analysis on radiation protection of FLASH treatment in conventional proton therapy facilities

Yuanyuan Wang, Chan Deng, Nan Gao, Yangfan Zhang, Yuxiang Wang, Ming Liu, Tao Zhou, Yi-Ming Lu, Hsiao-Ming Lu

Medical Physics · 2026-11 · 53 (11)

DOI: 10.1002/mp.70747 ↗

At a glance

Journal article in Medical Physics, published online on 2026-10-30. Written by Yuanyuan Wang and 8 co-authors. Institution: Hefei Ion Medical Center, the First Affiliated Hospital of USTC, Division of Life Sciences and Medicine, University of Science and Technology of Chin…. Major MEDLINE topics: Proton Therapy, Radiation Protection. 30 references. Funded by: Nuclear Technology R&D Project of the State Administration of Science, Technology and Industry for National Defense, Anhui Province Special Project on Clinical Medical Research, Anhui Provincial Ecological and Environmental Science and Technology Project.

RezumatSubjectsDate bibliograficeCiteRelated studies

Abstract (original language)

Background

FLASH radiotherapy is an advanced radiotherapy technique featuring ultra‐high dose‐rate irradiation. Previous preclinical studies have shown that it can significantly reduce damage to normal tissues while preserving the ability to kill tumor cells. Due to the very high current intensity of FLASH radiotherapy, the effectiveness of radiation shielding in current proton therapy facilities needs to be evaluated prior to clinical studies.

Purpose

Based on the existing shielding conditions of a ProBeam proton therapy system for conventional proton therapy treatment, we studied the shielding effectiveness of FLASH treatment in both the transmission and Bragg‐peak modes.

Methods

Under the conditions of single‐point irradiation using a proton beam with the highest clinical energy of 244MeV, and the beam intensity of 5nA, 10nA, and 20nA at the nozzle, the ambient equivalent dose rate outside the shielding door was measured with neutron and γ detector, in order to determine the linear relationship between the current intensity and the ambient dose rate. In addition, a gantry room was used as the study object, and the layout design of the points of interest was carried out according to the layout of the proton treatment facility and the transport path and direction of the proton beam, and the most unfavorable working conditions were set for each point of interest. The beam energy was set to 244 MeV, with a cylindrical water phantom (20 cm in diameter) and a solid water block (40 × 40 × 40 cm 3 ) placed at the isocenter. The current at the nozzle was set to 20 nA, and comprehensive environmental dose rate measurements were carried out for the proton radiation workplace.

Results

A linear relationship was observed between nozzle current and ambient dose equivalent rate. Based on the linear relationship, when the current was extrapolated to a FLASH current of 170 nA, the measured instantaneous dose rates at some locations were relatively high, with the maximum value of 74.04 µSv/h recorded outside the shielding door. Under the conservative assumption that the total annual beam delivery time is 5 h (treating 6 patients per hour), the maximum annual occupational exposure doses for radiation workers under the transmission and Bragg Peak modes are 0.370 mSv and 0.236 mSv, respectively, which are well below the annual occupational dose limit.

Conclusions

The results indicate that, under existing conventional shielding conditions, the radiation risk posed to the environment outside the treatment room by FLASH irradiation using the Varian ProBeam proton therapy system is safe and controllable, and the annual occupational radiation dose for radiation workers is well below the annual dose equivalent limit. This provides important radiation safety justification for the clinical implementation of FLASH technology in existing proton therapy facilities.

Subjects

MeSH terms (MEDLINE)

Proton TherapyRadiation ProtectionRadiotherapy DosageHumans

Author keywords

FLASH radiotherapyambient dose equivalentradiation safetyscanning proton therapyshielding evaluation

Authors and institutions

  • Hefei Ion Medical Center, the First Affiliated Hospital of USTC, Division of Life Sciences and Medicine, University of Science and Technology of China, Hefei, China.
  • University of Science and Technology of China, Hefei, China.
  • Cancer Hospital Affiliated to Shandong First Medical University, Jinan, China.
  • Anhui Environmental Radiation Supervision Station, Hefei, China.

Funding

  • Nuclear Technology R&D Project of the State Administration of Science, Technology and Industry for National Defense
  • Anhui Province Special Project on Clinical Medical Research
  • Anhui Provincial Ecological and Environmental Science and Technology Project

Cite

APA

Wang, Y., Deng, C., Gao, N., Zhang, Y., Wang, Y., Liu, M., Zhou, T., Lu, Y. M., & Lu, H. M. (2026). Analysis on radiation protection of FLASH treatment in conventional proton therapy facilities. Medical Physics, 53(11). https://doi.org/10.1002/mp.70747

Vancouver

Wang Y, Deng C, Gao N, Zhang Y, Wang Y, Liu M, et al. Analysis on radiation protection of FLASH treatment in conventional proton therapy facilities. Medical Physics. 2026;53(11). doi:10.1002/mp.70747

BibTeX
@article{Wang2026,
  author = {Wang, Yuanyuan and Deng, Chan and Gao, Nan and Zhang, Yangfan and Wang, Yuxiang and Liu, Ming and Zhou, Tao and Lu, Yi-Ming and Lu, Hsiao-Ming},
  title = {Analysis on radiation protection of FLASH treatment in conventional proton therapy facilities},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {11},
  doi = {10.1002/mp.70747}
}
Start for free →