Development and performance evaluation of an upright dedicated cone‐beam breast CT system

Thomas C Larsen iD, Hsin Wu Tseng iD, William Ross iD, Stephen Araujo iD, Pengwei Wu iD, Eri Haneda iD, Cynthia Davis iD, Andrew Karellas iD, Srinivasan Vedantham iD

Medical Physics · 2026

DOI: 10.1002/mp.70664 ↗

At a glance

Evaluationsstudie in Medical Physics, 2026. Written by Thomas C Larsen and 8 co-authors. Institution: Department of Biomedical Engineering, University of Arizona, Tucson, Arizona, USA. Major MEDLINE topics: Cone-Beam Computed Tomography, Breast, Mammography. 59 references. The full text is freely available in PubMed Central. Funded by: NCI NIH HHS, American Cancer Society, National Cancer Institute (NCI) of the National Institutes of Health.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Dedicated breast CT is an emerging breast X‐ray imaging modality. While current commercial breast CT systems use prone‐patient, pendant‐breast geometry, the system described here uses upright patient geometry with the uncompressed breast supported by a cup.

Purpose

The purpose of this work is to describe the development of a newly designed, upright geometry, dedicated cone‐beam breast CT system and to evaluate its imaging performance using objective metrics.

Methods

The prototype system uses a tungsten‐target, mammography‐format, X‐ray tube operating at 60 kV with 0.25 mm Cu and 1 mm Al added filtration, and a complementary metal‐oxide semiconductor (CMOS) detector with 0.152 mm pixel pitch coupled to 500 microns thick CsI:Tl scintillator. During short scan acquisition, the X‐ray source moves inferior to the breast, and 210 projections are acquired over an angular range of 210 degrees. The projections are reconstructed to an isotropic voxel pitch of 0.22 mm using Feldkamp‐Davis‐Kress (FDK) algorithm with Parker weights. Quantitative performance measures including linearity, modulation transfer function (MTF), and noise power spectrum (NPS) were evaluated. Phantom studies were conducted at various X‐ray tube current (mA) and pulse‐width (ms) combinations with the objective of determining the minimum detectable size of low‐contrast targets and calcium carbonate spheres representing soft tissue lesions and microcalcification clusters, respectively.

Results

The measured 1st HVL was 4.23 ± 0.01 mm of Al. The limiting resolution (10% MTF) was 2.18 mm −1 in the coronal (cross‐sectional) plane near the axis of rotation. In the coronal plane, the peak of the NPS occurred at 0.5 mm −1 . Phantom studies at a mean glandular dose of 3–5.7 mGy showed the ability to visualize 2–3 mm low‐contrast targets and 0.27–0.29 mm calcium carbonate spheres.

Conclusions

The developed upright breast CT system showed the ability to achieve high spatial resolution and low contrast resolution. The excellent technical performance of the breast CT system reported here suggests that further investigations using patient imaging are warranted.

Subjects

MeSH terms (MEDLINE)

Cone-Beam Computed TomographyBreastMammographyPhantoms, ImagingHumansEquipment DesignFemaleImage Processing, Computer-Assisted

Author keywords

breast CTbreast cancermodulation transfer functionnoise power spectrumupright geometry

Authors and institutions

  • Department of Biomedical Engineering, University of Arizona, Tucson, Arizona, USA.
  • Department of Radiology & Imaging Sciences, University of Arizona, Tucson, Arizona, USA.
  • GE Aerospace Research, Niskayuna, New York, USA.
  • GE HealthCare Technology & Innovation Center, Niskayuna, New York, USA.

ORCID profiles: Thomas C Larsen, Hsin Wu Tseng, William Ross, Stephen Araujo, Pengwei Wu, Eri Haneda, Cynthia Davis, Andrew Karellas, Srinivasan Vedantham

Funding

  • NCI NIH HHS
  • American Cancer Society
  • National Cancer Institute (NCI) of the National Institutes of Health

Cite

APA

Larsen, T. C., Tseng, H. W., Ross, W., Araujo, S., Wu, P., Haneda, E., Davis, C., Karellas, A., & Vedantham, S. (2026). Development and performance evaluation of an upright dedicated cone‐beam breast CT system. Medical Physics. https://doi.org/10.1002/mp.70664

Vancouver

Larsen TC, Tseng HW, Ross W, Araujo S, Wu P, Haneda E, et al. Development and performance evaluation of an upright dedicated cone‐beam breast CT system. Medical Physics. 2026. doi:10.1002/mp.70664

BibTeX
@article{Larsen2026,
  author = {Larsen, Thomas C and Tseng, Hsin Wu and Ross, William and Araujo, Stephen and Wu, Pengwei and Haneda, Eri and Davis, Cynthia and Karellas, Andrew and Vedantham, Srinivasan},
  title = {Development and performance evaluation of an upright dedicated cone‐beam breast CT system},
  journal = {Medical Physics},
  year = {2026},
  doi = {10.1002/mp.70664}
}
Start for free →