EUVIMEDSkopri EUVIMED
Open access · CC BY 4.0

Magnetic resonance imaging safety‐driven, heat‐mitigating design for a magnetic resonance imaging‐guided, magnetically actuated robotic catheter

E Erdem Tuna, Gabriel J Foss, Sanjana Kamath, Ridaa Z Ali, Yen-Chun Chen, Ahmed B Aydin, Reid Bolding, Mark Griswold, M Cenk Cavusoglu

Medical Physics · 2026-08 · 53 (8)

DOI: 10.1002/mp.70630 ↗

At a glance

Journal article in Medical Physics, published online on 2026-08-10. Written by E Erdem Tuna and 8 co-authors. Institution: Department of Electrical, Computer, and Systems Engineering, Case Western Reserve University, Cleveland, Ohio, USA. Major MEDLINE topics: Magnetic Resonance Imaging, Hot Temperature, Catheters, Robotic Surgical Procedures. 50 references. The full text is freely available (open access, CC BY 4.0). Funded by: National Science Foundation, NHLBI NIH HHS, NIH HHS.

AstrattSubjectsDettalji biblijografiċiCiteRelated studies

Abstract (original language)

Background

Metallic components in magnetic resonance imaging (MRI)−guided robotic catheters, such as leadwires and microcoils, can interact with the scanner's radiofrequency (RF) fields, causing heating via the antenna effect and Joule heating, leading to localized temperature rises and an increased risk of thermal injury. Ensuring thermal safety will enable safe and reliable operation of magnetically actuated robotic catheters within MRI environments.

Purpose

This study aims to develop a novel MRI‐compatible robotic catheter design that mitigates RF‐induced and resistive heating, ensuring patient safety and proper functionality during MRI‐guided interventions, particularly for cardiac ablation procedures. The catheter is directly actuated by magnetic torques generated on current‐carrying microcoils at the tip by the MRI scanner's static magnetic field, allowing precise navigation.

Methods

The catheter design incorporates systematically tuned microcoil parameters and distributed capacitors along the leadwires to suppress resonance and minimize RF energy absorption. A dithering technique applying actuation currents at a 50 duty cycle (10 Hz, 50 ms on/off) promotes forced convection to reduce resistive heating. A resistor‐capacitor (RC) equivalent circuit model was used to characterize the catheter's thermal behavior. Thermal validation was conducted on a 3T MRI scanner using a saline‐filled phantom, fiber‐optic temperature probes, and a balanced steady‐state free precession (bSSFP) imaging sequence with actuation currents up to 1 A. Hot spot detection sweeps along the leadwires and spatial heating assessments at multiple bore positions were also performed.

Results

The prototype exhibited a maximum temperature increase of 0.6 during simultaneous imaging and 1 A dithered actuation, well within Food and Drug Administration (FDA) safety limits. No hot spots were detected along the leadwires, and thermal performance remained consistent across all bore positions tested.

Conclusions

The catheter design effectively mitigates both RF‐induced and resistive heating under the tested conditions. These results demonstrate the suitability of the design for MRI‐guided interventional procedures under the tested conditions, while remaining within FDA temperature limits.

Subjects

MeSH terms (MEDLINE)

Magnetic Resonance ImagingHot TemperatureCathetersRobotic Surgical ProceduresRoboticsEquipment DesignEquipment Safety

Author keywords

MRI RF heatingMRI‐guided robotic catheterinterventional MRI device safety

Authors and institutions

  • Department of Electrical, Computer, and Systems Engineering, Case Western Reserve University, Cleveland, Ohio, USA.
  • Department of Physics, Case Western Reserve University, Cleveland, Ohio, USA.
  • Department of Radiology, Case Western Reserve University, Cleveland, Ohio, USA.

Funding

  • National Science Foundation
  • NHLBI NIH HHS
  • NIH HHS
  • National Institutes of Health

Cite

APA

Tuna, E. E., Foss, G. J., Kamath, S., Ali, R. Z., Chen, Y. C., Aydin, A. B., Bolding, R., Griswold, M., & Cavusoglu, M. C. (2026). Magnetic resonance imaging safety‐driven, heat‐mitigating design for a magnetic resonance imaging‐guided, magnetically actuated robotic catheter. Medical Physics, 53(8). https://doi.org/10.1002/mp.70630

Vancouver

Tuna EE, Foss GJ, Kamath S, Ali RZ, Chen YC, Aydin AB, et al. Magnetic resonance imaging safety‐driven, heat‐mitigating design for a magnetic resonance imaging‐guided, magnetically actuated robotic catheter. Medical Physics. 2026;53(8). doi:10.1002/mp.70630

BibTeX
@article{Tuna2026,
  author = {Tuna, E Erdem and Foss, Gabriel J and Kamath, Sanjana and Ali, Ridaa Z and Chen, Yen-Chun and Aydin, Ahmed B and Bolding, Reid and Griswold, Mark and Cavusoglu, M Cenk},
  title = {Magnetic resonance imaging safety‐driven, heat‐mitigating design for a magnetic resonance imaging‐guided, magnetically actuated robotic catheter},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {8},
  doi = {10.1002/mp.70630}
}