Magnetic resonance imaging safety‐driven, heat‐mitigating design for a magnetic resonance imaging‐guided, magnetically actuated robotic catheter
Medical Physics · 2026-08 · 53 (8)
DOI: 10.1002/mp.70630 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, online erschienen am 10.08.2026. Verfasst von E Erdem Tuna und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical, Computer, and Systems Engineering, Case Western Reserve University, Cleveland, Ohio, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Hot Temperature, Catheters, Robotic Surgical Procedures. Die Arbeit nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Science Foundation, NHLBI NIH HHS, NIH HHS.
Abstract (Originalsprache)
Background
Metallic components in magnetic resonance imaging (MRI)−guided robotic catheters, such as leadwires and microcoils, can interact with the scanner's radiofrequency (RF) fields, causing heating via the antenna effect and Joule heating, leading to localized temperature rises and an increased risk of thermal injury. Ensuring thermal safety will enable safe and reliable operation of magnetically actuated robotic catheters within MRI environments.
Purpose
This study aims to develop a novel MRI‐compatible robotic catheter design that mitigates RF‐induced and resistive heating, ensuring patient safety and proper functionality during MRI‐guided interventions, particularly for cardiac ablation procedures. The catheter is directly actuated by magnetic torques generated on current‐carrying microcoils at the tip by the MRI scanner's static magnetic field, allowing precise navigation.
Methods
The catheter design incorporates systematically tuned microcoil parameters and distributed capacitors along the leadwires to suppress resonance and minimize RF energy absorption. A dithering technique applying actuation currents at a 50 duty cycle (10 Hz, 50 ms on/off) promotes forced convection to reduce resistive heating. A resistor‐capacitor (RC) equivalent circuit model was used to characterize the catheter's thermal behavior. Thermal validation was conducted on a 3T MRI scanner using a saline‐filled phantom, fiber‐optic temperature probes, and a balanced steady‐state free precession (bSSFP) imaging sequence with actuation currents up to 1 A. Hot spot detection sweeps along the leadwires and spatial heating assessments at multiple bore positions were also performed.
Results
The prototype exhibited a maximum temperature increase of 0.6 during simultaneous imaging and 1 A dithered actuation, well within Food and Drug Administration (FDA) safety limits. No hot spots were detected along the leadwires, and thermal performance remained consistent across all bore positions tested.
Conclusions
The catheter design effectively mitigates both RF‐induced and resistive heating under the tested conditions. These results demonstrate the suitability of the design for MRI‐guided interventional procedures under the tested conditions, while remaining within FDA temperature limits.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Electrical, Computer, and Systems Engineering, Case Western Reserve University, Cleveland, Ohio, USA.
- Department of Physics, Case Western Reserve University, Cleveland, Ohio, USA.
- Department of Radiology, Case Western Reserve University, Cleveland, Ohio, USA.
Förderung
- National Science Foundation
- NHLBI NIH HHS
- NIH HHS
- National Institutes of Health
Zitieren
APA
Tuna, E. E., Foss, G. J., Kamath, S., Ali, R. Z., Chen, Y. C., Aydin, A. B., Bolding, R., Griswold, M., & Cavusoglu, M. C. (2026). Magnetic resonance imaging safety‐driven, heat‐mitigating design for a magnetic resonance imaging‐guided, magnetically actuated robotic catheter. Medical Physics, 53(8). https://doi.org/10.1002/mp.70630
Vancouver
Tuna EE, Foss GJ, Kamath S, Ali RZ, Chen YC, Aydin AB, et al. Magnetic resonance imaging safety‐driven, heat‐mitigating design for a magnetic resonance imaging‐guided, magnetically actuated robotic catheter. Medical Physics. 2026;53(8). doi:10.1002/mp.70630
BibTeX
@article{Tuna2026,
author = {Tuna, E Erdem and Foss, Gabriel J and Kamath, Sanjana and Ali, Ridaa Z and Chen, Yen-Chun and Aydin, Ahmed B and Bolding, Reid and Griswold, Mark and Cavusoglu, M Cenk},
title = {Magnetic resonance imaging safety‐driven, heat‐mitigating design for a magnetic resonance imaging‐guided, magnetically actuated robotic catheter},
journal = {Medical Physics},
year = {2026},
volume = {53},
number = {8},
doi = {10.1002/mp.70630}
}
EUVIMED