EUVIMEDEUVIMED entdecken
Open Access

Experimental determination of PET modulation transfer functions incorporating positron range effects using an oversampled cylindrical source

Antonio González-López, Daniel Blasco-Avellaneda, Salvador González-Bernal

Medical Physics · 2026-08 · 53 (8)

DOI: 10.1002/mp.70603 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, online erschienen am 21.07.2026. Verfasst von Antonio González-López und 2 weiteren Autor:innen. Beteiligte Einrichtung: Hospital Clínico Universitario Virgen de la Arrixaca - Instituto Murciano de Investigación Biosanitaria Pascual Parrilla, Murcia, Spain. Hauptthemen laut MEDLINE: Positron-Emission Tomography, Image Processing, Computer-Assisted. Die Arbeit nennt 16 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Spatial resolution is a fundamental performance parameter of positron emission tomography (PET) systems and is commonly assessed using standardized point‐source‐in‐air procedures. While this configuration minimizes scattering and attenuation effects, it does not account for positron transport in tissue‐equivalent media, whose contribution may become increasingly relevant as PET system resolution improves.

Purpose

To develop and validate a direct experimental methodology for PET spatial resolution assessment that incorporates positron‐range effects without relying on Monte Carlo simulations or parametric modeling of the system response.

Methods

A modular polymethyl methacrylate (PMMA) phantom containing a thin linear radioactive source ( or ) was used to ensure full positron absorption. By slightly inclining the source relative to the scanner axis, an oversampling strategy was implemented. The reconstructed volume was projected onto a one‐dimensional profile to obtain a sub‐pixel sampled point spread function (PSF), from which the modulation transfer function (MTF) was derived by Fourier transformation. Finite source‐size correction and uncertainty propagation from PSF to MTF were performed. Radial and tangential MTFs were also determined at off‐center positions.

Results

The oversampling strategy yielded very small statistical uncertainties in both PSF and MTF estimates. Measurements performed on two PET/CT systems of different technological generations showed negligible differences between and in the older scanner, whereas measurable radionuclide‐dependent effects were observed in the newer system. In the latter, exhibited a broader PSF and lower MTF than , with values of 0.085 and 0.112 , respectively.

Conclusions

The proposed methodology provides a statistically efficient framework for experimental PET spatial‐resolution characterization under realistic conditions. It enables quantitative assessment of positron‐range effects and complements standardized resolution measurements, particularly in modern high‐resolution PET systems.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Positron-Emission TomographyImage Processing, Computer-AssistedPhantoms, ImagingMonte Carlo MethodPolymethyl Methacrylate

Stichwörter der Autor:innen

modulation transfer functionoversamplingpoint spread functionpositron emission tomographypositron rangespatial resolution

Substanzen

Polymethyl Methacrylate

Autor:innen und Institutionen

  • Hospital Clínico Universitario Virgen de la Arrixaca - Instituto Murciano de Investigación Biosanitaria Pascual Parrilla, Murcia, Spain.

Zitieren

APA

González-López, A., Blasco-Avellaneda, D., & González-Bernal, S. (2026). Experimental determination of PET modulation transfer functions incorporating positron range effects using an oversampled cylindrical source. Medical Physics, 53(8). https://doi.org/10.1002/mp.70603

Vancouver

González-López A, Blasco-Avellaneda D, González-Bernal S. Experimental determination of PET modulation transfer functions incorporating positron range effects using an oversampled cylindrical source. Medical Physics. 2026;53(8). doi:10.1002/mp.70603

BibTeX
@article{GonzlezLpez2026,
  author = {González-López, Antonio and Blasco-Avellaneda, Daniel and González-Bernal, Salvador},
  title = {Experimental determination of PET modulation transfer functions incorporating positron range effects using an oversampled cylindrical source},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {8},
  doi = {10.1002/mp.70603}
}
Kostenlos starten →