GPU‐based list‐mode TOF PET image reconstruction with complete correction techniques
Medical Physics · 2026
DOI: 10.1002/mp.70216 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2026. Verfasst von Ziquan Yuan und 6 weiteren Autor:innen. Beteiligte Einrichtung: School of Physics and Astronomy, Beijing Normal University, Beijing, China. Hauptthemen laut MEDLINE: Image Processing, Computer-Assisted, Positron-Emission Tomography, Computer Graphics. Die Arbeit nennt 53 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Beijing Nova Program.
Abstract (Originalsprache)
Background
Positron emission tomography (PET) is essential for the early diagnosis of cancer, neurological disorders, and cardiovascular diseases. However, achieving high‐quality PET images remains challenging due to the complex physical factors involved and the trade‐off between reconstruction accuracy and computational efficiency.
Purpose
This study aimed to develop QuanTOF, a GPU‐accelerated PET reconstruction framework integrating comprehensive physical corrections and advanced modeling to enhance image quality while maintaining clinical practicality.
Methods
QuanTOF employs a GPU‐accelerated Bayesian penalized‐likelihood reconstruction algorithm with time‐of‐flight (TOF) and point spread function (PSF) modeling. It incorporates full corrections for attenuation, normalization, random coincidences, and scatter coincidences. A memory‐efficient TOF single scatter simulation (SSS) algorithm enabled on‐the‐fly scatter correction without storing full TOF sinograms. Validation included Monte Carlo simulations, clinical phantom experiments, and blinded reader studies using a Siemens Biograph Vision PET/CT scanner.
Results
In phantom studies, QuanTOF achieved uniformity comparable to commercial systems and resolved 2.4 mm hot rods in Derenzo phantoms, with peak‐to‐valley ratios of 2.61 (3.2 mm rods) and 1.29 (2.4 mm rods). Scatter correction time was reduced to s, two orders of magnitude faster than existing methods. Clinicians rated QuanTOF images significantly higher (average score: 3.90 vs. 2.15 for clinical images) in reader studies. Reconstruction times remained clinically acceptable (e.g., 2.78 s for NEMA phantom scatter correction)
Conclusions
QuanTOF balances accuracy and efficiency through GPU‐optimized physics modeling and memory‐efficient algorithms. It delivers high‐resolution PET images with diagnostic confidence, demonstrating potential for clinical oncology, neurology, and cardiology applications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Physics and Astronomy, Beijing Normal University, Beijing, China.
- The First Affiliated Hospital of USTC, Division of Life Sciences and Medicine, University of Science and Technology of China, Hefei, China.
- Wuxi School of Medicine, Jiangnan University, Wuxi, China.
- School of Artificial Intelligence, Beijing Normal University, Beijing, China.
- Department of Biomedical Engineering, University of Melbourne, Melbourne, VIC, Australia.
Förderung
- National Natural Science Foundation of China
- Beijing Nova Program
Zitieren
APA
Yuan, Z., Zhan, F., Lu, H., Hou, Y., Liao, R., Li, C., & Jiang, J. (2026). GPU‐based list‐mode TOF PET image reconstruction with complete correction techniques. Medical Physics. https://doi.org/10.1002/mp.70216
Vancouver
Yuan Z, Zhan F, Lu H, Hou Y, Liao R, Li C, et al. GPU‐based list‐mode TOF PET image reconstruction with complete correction techniques. Medical Physics. 2026. doi:10.1002/mp.70216
BibTeX
@article{Yuan2026,
author = {Yuan, Ziquan and Zhan, Fenglin and Lu, Haoyu and Hou, Yucun and Liao, Runze and Li, Chenxi and Jiang, Jianyong},
title = {GPU‐based list‐mode TOF PET image reconstruction with complete correction techniques},
journal = {Medical Physics},
year = {2026},
doi = {10.1002/mp.70216}
}