Accurate estimation of intravoxel incoherent motion parameters based on implicit neural representation
Medical Physics · 2026-08 · 53 (8)
DOI: 10.1002/mp.70599 ↗At a glance
Journal article in Medical Physics, published online on 2026-07-28. Written by Yunxiang Li and 4 co-authors. Institution: Department of Radiation Oncology, UT Southwestern Medical Center, Dallas, Texas, USA. Major MEDLINE topics: Diffusion Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Cited 1 times so far (Crossref); 52 references. The full text is freely available (open access, CC BY 4.0). Funded by: NIH HHS, NCI NIH HHS, NIBIB NIH HHS.
Abstract (original language)
Background
Intravoxel incoherent motion (IVIM) diffusion‐weighted imaging has important value in treatment response monitoring. However, traditional voxel‐wise independent fitting methods are highly sensitive to noise and do not utilize spatial correlation, resulting in unstable parameter estimation. Although existing deep learning methods have shown improvements, they are still limited by local receptive fields.
Purpose
To address this, we propose a two‐stage IVIM parameter estimation framework based on Implicit Neural Representation (IVIM‐INR).
Methods
Our IVIM‐INR method achieves global spatial perception through coordinate encoding and enhances spatial context modeling by leveraging local 3D patch information from multi‐ b ‐value images. The first stage INR performs signal denoising, and the second stage INR accurately fits IVIM parameters.
Results
Evaluation on brain digital phantoms, AAPM breast IVIM‐dMRI Challenge data, and clinical Glioblastoma (GBM) patient data demonstrates significant advantages of the proposed method over existing techniques. In brain simulation data, when SNR = 50, the normalized mean absolute errors (NMAEs) in tumor regions were 0.160.10 for , 0.020.01 for , and 0.070.05 for , all lower than comparison methods. In the tumor tissues of the 100 cases from the AAPM breast IVIM‐dMRI challenge dataset, error was reduced by 58% compared to ConvNet. Intraclass correlation coefficient (ICC) analysis of real clinical data indicates that our method achieves the best performance with an ICC of in normal tissues reaching 0.629.
Conclusions
By combining INR's continuous function modeling capability with spatial‐aware feature design, IVIM‐INR overcomes the inherent limitations of traditional methods under noisy conditions, providing a more reliable tool for clinical IVIM quantitative analysis.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Radiation Oncology, UT Southwestern Medical Center, Dallas, Texas, USA.
Funding
- NIH HHS
- NCI NIH HHS
- NIBIB NIH HHS
- National Institutes of Health
Cite
APA
Li, Y., Liao, Y. P., Dai, Y., Deng, J., & Zhang, Y. (2026). Accurate estimation of intravoxel incoherent motion parameters based on implicit neural representation. Medical Physics, 53(8). https://doi.org/10.1002/mp.70599
Vancouver
Li Y, Liao YP, Dai Y, Deng J, Zhang Y. Accurate estimation of intravoxel incoherent motion parameters based on implicit neural representation. Medical Physics. 2026;53(8). doi:10.1002/mp.70599
BibTeX
@article{Li2026,
author = {Li, Yunxiang and Liao, Yen-Peng and Dai, Yan and Deng, Jie and Zhang, You},
title = {Accurate estimation of intravoxel incoherent motion parameters based on implicit neural representation},
journal = {Medical Physics},
year = {2026},
volume = {53},
number = {8},
doi = {10.1002/mp.70599}
}
EUVIMED