Theoretical assessment of DECT noise on physical and biological dose accuracy in carbon‐ion radiation therapy

Weiguang Li, Yan Li, Shutong Yu, Xueying Yang, Chao Yang, Cheng Chang, Mingqing Wang, Chong Xu, Kai-Wen Li, Li-Sheng Geng, Yibao Zhang

Medical Physics · 2026-08 · 53 (8)

DOI: 10.1002/mp.70590 ↗

At a glance

Journal article in Medical Physics, published online on 2026-08-06. Written by Weiguang Li and 10 co-authors. Institution: School of Physics, Beihang University, Beijing, China. Major MEDLINE topics: Heavy Ion Radiotherapy, Tomography, X-Ray Computed, Radiation Dosage, Signal-To-Noise Ratio. 37 references. The full text is freely available in PubMed Central. Funded by: Beijing Natural Science Foundation, National Natural Science Foundation of China, Clinical Research Fund For Distinguished Young Scholars of Peking University Cancer Hospital.

TiivistelmäSubjectsBibliografiset tiedotCiteRelated studies

Abstract (original language)

Background

Accurate characterization of tissue parameters is essential for precise dose calculation in Carbon Ion Radiation Therapy (CIRT). Recent advances in Dual‐Energy CT (DECT) have improved the estimation of tissue parameters, yet DECT‐based methods are susceptible to image noise. The influence on physical and biological dose accuracy has not been thoroughly investigated, undermining the evidence‐based clinical application and protocol optimization.

Objective

To systematically examine how image noise in DECT influences the estimated tissue parameters and its consequent impact on the accuracy of physical and biological dose distributions in CIRT. Materials and methods Four DECT‐based elemental decomposition methods were evaluated. A machine‐learning (ML) approach was compared with three parameterization (PA) methods, that is, the Hünemohr model using input parameters (, ) obtained from the Saito‐Hünemohr, Saito‐Landry, and Bourque schemes, respectively. Clinically relevant noise levels of 0%, 2%, and 5% were chosen to assess the four methods for calculating the carbon‐ion range deviations in 85 reference tissues and estimate the elemental composition of the ICRP110 human phantom. Physical and biological dose distributions in CIRT were calculated using Monte Carlo simulations. The biological doses were modeled using the Linear Quadratic Model (LQM), Microdosimetric Kinetic Model (MKM), and Local Effect Model (LEM). Gamma analysis was applied to evaluate the dose deviations.

Results

Across the investigated noise levels, the ML approach consistently outperformed the three PA methods. Compared with PA‐based methods, the ML approach reduced the average water‐equivalent range deviations by 0.4 mm–1.4 mm and improved gamma passing rates by 0.8%–3.9% (physical dose) and 6.4%–24.1% (biological dose) under the 1 mm/1% criteria.

Conclusion

The ML method provides superior robustness and accuracy in physical and biological dose calculation based on DECT of various noise levels. The LEM demonstrated superior noise robustness compared to the LQM and the MKM.

Subjects

MeSH terms (MEDLINE)

Heavy Ion RadiotherapyTomography, X-Ray ComputedRadiation DosageSignal-To-Noise RatioRadiotherapy DosageHumansPhantoms, Imaging

Author keywords

biological dosecarbon ion radiation therapydose calculationdual energy CTmaterial decomposition

Authors and institutions

  • School of Physics, Beihang University, Beijing, China.
  • Department of Technology, CAS Ion Medical Technology Co., Ltd, Beijing, China.
  • Key Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Radiation Oncology, Peking University Cancer Hospital & Institute, Beijing, China.
  • Hangzhou International Innovation Institute, Beihang University, Hangzhou, Zhejiang, China.
  • Department of Radiation Oncology, Cancer Center, Peking University Third Hospital, Beijing, China.
  • Zhejiang Cancer Hospital, Department of Radiation Physics, Hangzhou Institute of Medicine (HIM), Chinese Academy of Sciences, Hangzhou, Zhejiang, China.

Funding

  • Beijing Natural Science Foundation
  • National Natural Science Foundation of China
  • Clinical Research Fund For Distinguished Young Scholars of Peking University Cancer Hospital
  • Fundamental Research Funds for the Central Universities
  • Science Foundation of Peking University Cancer Hospital
  • Heavy Ion Research Facility in Lanzhou
  • Natural Science Foundation of Beijing Municipality

Cite

APA

Li, W., Li, Y., Yu, S., Yang, X., Yang, C., Chang, C., Wang, M., Xu, C., Li, K. W., Geng, L. S., & Zhang, Y. (2026). Theoretical assessment of DECT noise on physical and biological dose accuracy in carbon‐ion radiation therapy. Medical Physics, 53(8). https://doi.org/10.1002/mp.70590

Vancouver

Li W, Li Y, Yu S, Yang X, Yang C, Chang C, et al. Theoretical assessment of DECT noise on physical and biological dose accuracy in carbon‐ion radiation therapy. Medical Physics. 2026;53(8). doi:10.1002/mp.70590

BibTeX
@article{Li2026,
  author = {Li, Weiguang and Li, Yan and Yu, Shutong and Yang, Xueying and Yang, Chao and Chang, Cheng and Wang, Mingqing and Xu, Chong and Li, Kai-Wen and Geng, Li-Sheng and Zhang, Yibao},
  title = {Theoretical assessment of DECT noise on physical and biological dose accuracy in carbon‐ion radiation therapy},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {8},
  doi = {10.1002/mp.70590}
}
Start for free →