EUVIMEDSkopri EUVIMED
Open access · CC BY-NC-ND 4.0

A new in silico model to precisely design focused ultrasound brain therapies

Allegra Conti, Marta Grossi, Patricia Cardoso De Andrade, Andrea Duggento, Nicola Toschi

Medical Physics · 2026-07 · 53 (7)

DOI: 10.1002/mp.70520 ↗

At a glance

Journal article in Medical Physics, published online on 2026-06-24. Written by Allegra Conti and 4 co-authors. Institution: Medical Physics Section, Department of Biomedicine and Prevention, University of Rome Tor Vergata, Rome, Italy. Major MEDLINE topics: Computer Simulation, Ultrasonic Therapy, Brain. 69 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: AIRC- italian association for cancer research, European Innovation Council, Italian Ministry of University and Research.

AstrattSubjectsDettalji biblijografiċiCiteRelated studies

Abstract (original language)

Background

Focused ultrasound (FUS) combined with microbubbles enables transient and noninvasive blood–brain barrier (BBB) opening, facilitating targeted drug delivery. However, accurate treatment planning remains difficult due to inter‐patient anatomical variability and the common assumption in simulations that brain tissues behave like water.

Purpose

To develop and evaluate MODFUS ( ), an in silico acoustic simulation framework that integrates a high‐resolution anatomical head model to quantify the impact of intracranial tissue heterogeneity and probe alignment on transmitted acoustic pressure, supporting treatment planning for BBB opening.

Methods

MODFUS integrates an anatomical head model comprising 115 tissue types and Computed Tomography (CT)‐derived skull properties. FUS simulations were conducted using a 250 kHz single‐element transducer at 13 distinct skull entry locations. Two modeling approaches were evaluated: a classical model, in which the skull was embedded in water and brain tissues were homogenized as water, and an heterogeneous anatomical model. To further assess robustness, an additional set of 50 simulations introduced controlled perturbations in probe positioning, consisting of angular deviations () and translational offsets (7 mm) relative to the reference configuration. Model‐ and configuration‐dependent differences were quantified using peak positive pressure (PPP), peak negative pressure (PNP), and potential therapeutic volume. Statistical significance was assessed using Wilcoxon rank‐sum or Wilcoxon signed‐rank tests ( = 0.05). For Classical vs Realistic models, Wilcoxon rank‐sum tests were applied to PPP, PNP, and BBB exposure volume, and Levene's test assessed variance differences across 13 positions. Multiple testing was controlled using the Holm–Bonferroni procedure ( = 0.05) across all tests. Effect sizes were quantified using Cohen's d with 95% confidence intervals.

Results

Compared to the classical “skull+water” benchmark model, the heterogeneous model predicted up to 11% lower PPP, slightly lower PNP (–5%), and 35% smaller potential BBB exposure volumes across the 13 paired sonication positions. After Bonferroni correction, Paired statistical testing (Wilcoxon signed‐rank, two‐sided) showed significant differences for PPP ( p = 0.0270) and potential BBB exposure volume ( p = 0.0015), while differences in PNP were not statistically significant ( p = 0.8286). Levene's test for variance confirmed significant heteroscedasticity for PPP ( p = 0.027) and PNP ( p = 3.210), but not for BBB exposure volume ( p = 0.2680). Cohen's d effect sizes indicated a large positive effect for PPP, a small negative effect for PNP, and a very large positive effect for BBB exposure volume.

Conclusions

MODFUS demonstrates the influence of incorporating detailed tissue heterogeneity on simulation outcomes, including pressure distribution and potential BBB exposure volume. These results highlight the importance of realistic soft tissue modeling and stereotaxic probe alignment for safe and effective FUS treatment planning. The study serves as a preliminary proof‐of‐concept. Future studies incorporating in vivo experiments will be required to quantify the accuracy of this approach.

Subjects

MeSH terms (MEDLINE)

Computer SimulationUltrasonic TherapyBrainHumansBlood-Brain Barrier

Author keywords

blood–brain barrier openingfocused ultrasoundpatient‐specific acoustic simulation

Authors and institutions

  • Medical Physics Section, Department of Biomedicine and Prevention, University of Rome Tor Vergata, Rome, Italy.
  • Athinoula A. Martinos Center, Harvard Medical School, Boston, Massachusetts, USA.

Funding

  • AIRC- italian association for cancer research
  • European Innovation Council
  • Italian Ministry of University and Research

Cite

APA

Conti, A., Grossi, M., De Andrade, P. C., Duggento, A., & Toschi, N. (2026). A new in silico model to precisely design focused ultrasound brain therapies. Medical Physics, 53(7). https://doi.org/10.1002/mp.70520

Vancouver

Conti A, Grossi M, De Andrade PC, Duggento A, Toschi N. A new in silico model to precisely design focused ultrasound brain therapies. Medical Physics. 2026;53(7). doi:10.1002/mp.70520

BibTeX
@article{Conti2026,
  author = {Conti, Allegra and Grossi, Marta and De Andrade, Patricia Cardoso and Duggento, Andrea and Toschi, Nicola},
  title = {A new in silico model to precisely design focused ultrasound brain therapies},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {7},
  doi = {10.1002/mp.70520}
}
Start for free →