EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Improving textural realism in breast phantom images

Luana de M Omena, Guilherme M de Oliveira, Thaís G do Rêgo, Yuri de A M Barbosa, Joao P V Teixeira, Telmo M Silva Filho, Bruno Barufaldi

Medical Physics · 2026

DOI: 10.1002/mp.70501 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2026. Verfasst von Luana de M Omena und 6 weiteren Autor:innen. Beteiligte Einrichtung: Center of Informatics, Federal University of Paraíba, Rua dos Escoteiros s/n, João Pessoa, Paraíba, Brazil. Hauptthemen laut MEDLINE: Phantoms, Imaging, Mammography, Breast, Image Processing, Computer-Assisted. Die Arbeit nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Terri Brodeur Breast Cancer Foundation, Susan G. Komen.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Breast phantom images can lack textural realism, limiting their utility for imaging research. Achieving realism across all BI‐RADS breast densities is essential for representative, unbiased datasets.

Purpose

This study validated simplex noise for improving realism and density fidelity in simulated mammary parenchyma across BI‐RADS categories.

Methods

Two reader trials were conducted with eight students from the Artificial Intelligence Applications Laboratory (ARIA) at the Federal University of Paraíba, Brazil. In the first trial, participants compared paired images with and without simplex noise for overall realism in BI‐RADS A (fatty) and BI‐RADS D (dense) patterns using 2‐AFC. During the second trial, optimized parameters were tested across all four BI‐RADS categories. Readers selected the most realistic image in each pair and rated confidence levels.

Results

In Trial 1, the simplex enhanced reconstruction was judged as more realistic in 96% of BI‐RADS D evaluations, compared to 58.6% for BI‐RADS A. Trial 2 achieved a high realism consensus for simplex enhanced reconstructions (1,288 vs. 312 selections), with category improvements: BI‐RADS A (92.5%), B (92.25%), C (81%), and D (57.75%). ROC analysis confirmed high discriminative performance, with AUC values of 0.95 (A), 0.74 (B), 0.85 (C), and 0.93 (D). Results indicate density dependent effects, with more pronounced improvements in lower density categories.

Conclusions

The introduction of simplex noise enhances the textural realism of simulated mammographic images, particularly representing variability across BI‐RADS categories. These simulations provide a resource for testing and validation without patient data.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingMammographyBreastImage Processing, Computer-AssistedHumansFemaleSignal-To-Noise Ratio

Stichwörter der Autor:innen

breast phantomsimplex noisetexture realism

Autor:innen und Institutionen

  • Center of Informatics, Federal University of Paraíba, Rua dos Escoteiros s/n, João Pessoa, Paraíba, Brazil.
  • School of Engineering Maths and Technology, University of Bristol, Bristol, UK.
  • Department of Radiology, University of Pennsylvania, Hamilton Walk, Pennsylvania, USA.

Förderung

  • Terri Brodeur Breast Cancer Foundation
  • Susan G. Komen

Zitieren

APA

Omena, L. M., de Oliveira, G. M., do Rêgo, T. G., Barbosa, Y. A. M., Teixeira, J. P. V., Filho, T. M. S., & Barufaldi, B. (2026). Improving textural realism in breast phantom images. Medical Physics. https://doi.org/10.1002/mp.70501

Vancouver

Omena LM, de Oliveira GM, do Rêgo TG, Barbosa YAM, Teixeira JPV, Filho TMS, et al. Improving textural realism in breast phantom images. Medical Physics. 2026. doi:10.1002/mp.70501

BibTeX
@article{Omena2026,
  author = {Omena, Luana de M and de Oliveira, Guilherme M and do Rêgo, Thaís G and Barbosa, Yuri de A M and Teixeira, Joao P V and Filho, Telmo M Silva and Barufaldi, Bruno},
  title = {Improving textural realism in breast phantom images},
  journal = {Medical Physics},
  year = {2026},
  doi = {10.1002/mp.70501}
}
Kostenlos starten →