Impact of partial volume correction on radiomics reproducibility in theranostic SPECT/CT imaging

Mohammad Saber Azimi, Maryam Cheraghi, Mohammad Ali Ghodsi Rad, Mohadeseh Bayat, Maryam Alhashim, Habibollah Dadgar, Mahmoud Alabedi, Nahid Ibrahim, Enas Alwuhaib, Hossein Arabi, Arman Rahmim, Habib Zaidi

Medical Physics · 2026-04 · 53 (4)

DOI: 10.1002/mp.70427 ↗

At a glance

Journal article in Medical Physics, published online on 2026-04-03. Written by Mohammad Saber Azimi and 11 co-authors. Institution: Doctoral School of Applied Informatics and Applied Mathematics, Óbuda University, Budapest, Hungary. Major MEDLINE topics: Radiomics, Single Photon Emission Computed Tomography Computed Tomography, Image Processing, Computer-Assisted. Cited 3 times so far (Crossref); 57 references. The full text is freely available (open access, CC BY 4.0). Funded by: Swiss National Science Foundation, Geneva League Against Cancer, Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung.

AstrattSubjectsDettalji biblijografiċiCiteRelated studies

Abstract (original language)

Background

Radiomics has shown potential for quantitative characterization of tumors in molecular imaging; however, its clinical translation in theranostic 1 7 7 Lu SPECT/CT remains limited due to poor robustness of extracted features to reconstruction variability and partial volume effects. Establishing reproducible radiomics biomarkers across correction strategies is therefore a prerequisite for reliable clinical modeling and treatment monitoring.

Purpose

This study aimed to evaluate radiomics feature reproducibility, defined as the stability of feature values across different partial volume correction (PVC) strategies and reconstruction settings, in clinical 1 7 7 Lu SPECT/CT imaging. In addition, we explored two volumetric shape‐based indices, the metastasis‐to‐liver ratio (MLR) and metastasis‐to‐spare liver ratio (MSLR), as surrogate markers of hepatic metastatic burden in the theranostic treatment setting.

Methods

In 13 patients (40 scans) treated with 177 Lu, 837 radiomics features were extracted from 11 abdominal regions and metastases on SPECT/CT using original and wavelet‐decomposed images across four bin widths (50–200). Two post‐reconstruction PVC methods, namely Richardson‐Lucy (RL) and Reblurred Van Cittert (RVC), were applied. Feature reproducibility was quantified using two complementary metrics: the intraclass correlation coefficient (ICC) to assess feature‐level stability across PVC strategies, and the concordance correlation coefficient (CCC) to evaluate pairwise agreement and systematic bias among reconstruction methods. Visual image quality assessments were independently performed by two experienced nuclear medicine specialists in a blinded setting. Exploratory metastatic tumor burden was assessed descriptively using 3D shape‐based MLR and MSLR indices.

Results

Low‐frequency wavelet decomposition (LLL‐wavelet) and original features showed the highest reproducibility (ICC ≥ 0.90 in >95% of liver and metastasis features at BW50), whereas high‐frequency features and larger bin widths demonstrated reduced stability. CCC analysis revealed excellent agreement between RL and RVC (≥0.95 in major organs at BW50–100), while agreement with uncorrected SPECT (no PVC) was consistently lower, especially for high‐frequency features. RL achieved higher visual scores in sharpness and contrast ( p < 0.01), with good inter‐reader agreement supporting the consistency of these assessments. MLR/MSLR demonstrated inter‐patient variability and were explored descriptively as indices of metastatic liver burden.

Conclusions

Reproducibility in theranostic SPECT radiomics is highly feature‐ and organ‐dependent and is further influenced by scanner‐specific factors and reconstruction protocols, which remain critical for real‐world clinical translation. RL and RVC showed stronger mutual agreements than each with uncorrected SPECT. Importantly, only RVC translated visual improvements into enhanced feature‐level reproducibility, while RL provided the most consistent overall balance of reproducibility and image quality, supporting its role as the preferred PVC strategy for clinical and modeling applications. Robust radiomics feature selection as well as standardized reproducible PVC strategies are essential to generate methodological harmonization for future clinical translation and to support integration of radiomics analyses into personalized SPECT theranostics.

Subjects

MeSH terms (MEDLINE)

RadiomicsSingle Photon Emission Computed Tomography Computed TomographyImage Processing, Computer-AssistedHumansReproducibility of ResultsLiver NeoplasmsFemaleMale

Author keywords

177Lu SPECT/CTliver metastasispartial volume correctionquantitative SPECTradiomics reproducibilitytheranostics

Authors and institutions

  • Doctoral School of Applied Informatics and Applied Mathematics, Óbuda University, Budapest, Hungary.
  • University Research and Innovation Center, Óbuda University, Budapest, Hungary.
  • PET-CT Unit, Department of Nuclear Medicine, Masih Daneshvari Hospital, Shahid Beheshti University of Medical Sciences, Tehran, Iran.
  • Department of Nuclear Medicine, Shohada-e Tajrish Medical Center, School of Medicine, Shahid Beheshti University of Medical Sciences, Tehran, Iran.
  • Department of Nuclear Medicine, Taleghani Hospital, School of Medicine, Shahid Beheshti University of Medical Sciences, Tehran, Iran.
  • Department of Radiology, College of Medicine, Imam Abdulrahman Bin Faisal University, Dammam, Saudi Arabia.

Funding

  • Swiss National Science Foundation
  • Geneva League Against Cancer
  • Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung

Cite

APA

Azimi, M. S., Cheraghi, M., Ghodsi Rad, M. A., Bayat, M., Alhashim, M., Dadgar, H., Alabedi, M., Ibrahim, N., Alwuhaib, E., Arabi, H., Rahmim, A., & Zaidi, H. (2026). Impact of partial volume correction on radiomics reproducibility in theranostic SPECT/CT imaging. Medical Physics, 53(4). https://doi.org/10.1002/mp.70427

Vancouver

Azimi MS, Cheraghi M, Ghodsi Rad MA, Bayat M, Alhashim M, Dadgar H, et al. Impact of partial volume correction on radiomics reproducibility in theranostic SPECT/CT imaging. Medical Physics. 2026;53(4). doi:10.1002/mp.70427

BibTeX
@article{Azimi2026,
  author = {Azimi, Mohammad Saber and Cheraghi, Maryam and Ghodsi Rad, Mohammad Ali and Bayat, Mohadeseh and Alhashim, Maryam and Dadgar, Habibollah and Alabedi, Mahmoud and Ibrahim, Nahid and Alwuhaib, Enas and Arabi, Hossein and Rahmim, Arman and Zaidi, Habib},
  title = {Impact of partial volume correction on radiomics reproducibility in theranostic SPECT/CT imaging},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {4},
  doi = {10.1002/mp.70427}
}
Start for free →