EUVIMEDEUVIMED entdecken
EvaluationsstudieOpen Access · CC BY 4.0

First evaluation of a heteroepitaxial diamond ionization chamber operating at low voltage for diagnostic X‐ray dosimetry

Kiyomitsu Shinsho, Koji Koyama, Keitaro Hitomi, Mitsuhiro Nogami, Osamu Maida, Toshiyuki Onodera, Kanata Kikkawa, Shimma Hashimoto, Yuta Hirai, Yusuke Koba, Ako Haga, Daiki Maruyama und 1 weitere

Medical Physics · 2026-03 · 53 (3)

DOI: 10.1002/mp.70363 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Medical Physics, online erschienen am 27.02.2026. Verfasst von Kiyomitsu Shinsho und 12 weiteren Autor:innen. Beteiligte Einrichtung: Graduate School of Human Health Sciences, Tokyo Metropolitan University, Arakawa-ku, Tokyo, Japan. Hauptthemen laut MEDLINE: Diamond, Radiometry, Electricity. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: joint research fund, Tokyo Metropolitan University, Tohoku University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Heteroepitaxial diamond has recently gained attention as a radiation detector material due to its wide bandgap, radiation hardness, and near‐tissue equivalence. Despite these advantages, its use as a solid‐state ionization chamber for diagnostic X‐ray dosimetry has not yet been established. Demonstrating stable, high‐efficiency operation at low voltage would enable compact dosimeters with a very small sensitive volume, which is difficult to achieve with conventional air ionization chambers.

Purpose

To perform the first characterization of a heteroepitaxial diamond ionization chamber (HED‐IC) operated at low bias voltage under diagnostic X‐ray conditions and to evaluate its feasibility as a compact, high‐efficiency dosimeter.

Methods

A heteroepitaxial diamond detector (4 × 4 × 0.5 mm 3 ) with Ti/Au electrodes was fabricated and evaluated using diagnostic X‐ray beams at tube voltages from 50 to 120 kV. Charge‐collection characteristics, dose linearity, energy dependence, and temporal response were assessed at negative bias voltages with magnitudes between −1 and −100 V. Monte Carlo simulations were performed using PHITS to compute the expected diamond‐to‐air sensitivity ratio under the same beam qualities for comparison with the experimental measurements.

Results

The HED‐IC exhibited excellent dose linearity ( R 2 > 0.997) and weak energy dependence (< 10%) across effective energies from 28.4 to 40.1 keV. The detector enables dose measurements within a very small sensitive volume, only 1/1250 of that of a typical air ionization chamber. The volume‐normalized sensitivity exceeded theoretical expectations, suggesting enhanced effective ionization efficiency. An increased response with higher bias voltage further indicated potential for high‐sensitivity operation.

Conclusions

The results demonstrate that the HED‐IC can operate as a low‐voltage, high‐efficiency solid‐state ionization chamber under diagnostic X‐ray conditions. Owing to the scalability of heteroepitaxial diamond growth, this detector concept provides a promising basis for compact, tissue‐equivalent dosimeters capable of real‐time dose monitoring across a wide range of radiological applications.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

DiamondRadiometryElectricityX-RaysMonte Carlo MethodElectrodes

Stichwörter der Autor:innen

diagnostic X‐ray dosimetryheteroepitaxial diamond detectorhigh‐sensitivity dosimetry

Substanzen

Diamond

Autor:innen und Institutionen

  • Graduate School of Human Health Sciences, Tokyo Metropolitan University, Arakawa-ku, Tokyo, Japan.
  • Faculty of Health Sciences, Tokyo Metropolitan University, Arakawa-ku, Tokyo, Japan.
  • Orbray Co., Ltd., Adachi-ku, Tokyo, Japan.
  • Graduate School of Engineering, Tohoku University, Sendai, Miyagi, Japan.
  • Graduate School of Engineering, Osaka University, Suita, Osaka, Japan.
  • Faculty of Engineering, Tohoku Institute of Technology, Sendai, Miyagi, Japan.

Förderung

  • joint research fund
  • Tokyo Metropolitan University
  • Tohoku University
  • Orbray Co., Ltd

Zitieren

APA

Shinsho, K., Koyama, K., Hitomi, K., Nogami, M., Maida, O., Onodera, T., Kikkawa, K., Hashimoto, S., Hirai, Y., Koba, Y., Haga, A., Maruyama, D., & Kim, S. W. (2026). First evaluation of a heteroepitaxial diamond ionization chamber operating at low voltage for diagnostic X‐ray dosimetry. Medical Physics, 53(3). https://doi.org/10.1002/mp.70363

Vancouver

Shinsho K, Koyama K, Hitomi K, Nogami M, Maida O, Onodera T, et al. First evaluation of a heteroepitaxial diamond ionization chamber operating at low voltage for diagnostic X‐ray dosimetry. Medical Physics. 2026;53(3). doi:10.1002/mp.70363

BibTeX
@article{Shinsho2026,
  author = {Shinsho, Kiyomitsu and Koyama, Koji and Hitomi, Keitaro and Nogami, Mitsuhiro and Maida, Osamu and Onodera, Toshiyuki and Kikkawa, Kanata and Hashimoto, Shimma and Hirai, Yuta and Koba, Yusuke and Haga, Ako and Maruyama, Daiki and Kim, Seong-Woo},
  title = {First evaluation of a heteroepitaxial diamond ionization chamber operating at low voltage for diagnostic X‐ray dosimetry},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {3},
  doi = {10.1002/mp.70363}
}
Kostenlos starten →