Diamond detectors for dose and instantaneous dose‐rate measurements for ultra‐high dose‐rate scanned helium ion beams
Medical Physics · 2024-02 · 51 (2) · 1450-1459
DOI: 10.1002/mp.16757 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, online erschienen am 24.09.2023. Verfasst von Thomas Tessonnier und 5 weiteren Autor:innen. Beteiligte Einrichtung: Heidelberg Ion Beam Therapy Center (HIT), Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany. Hauptthemen laut MEDLINE: Diamond, Helium. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS.
Abstract (Originalsprache)
Background
The possible emergence of the FLASH effect—the sparing of normal tissue while maintaining tumor control—after irradiations at dose‐rates exceeding several tens of Gy per second, has recently spurred a surge of studies attempting to characterize and rationalize the phenomenon. Investigating and reporting the dose and instantaneous dose‐rate of ultra‐high dose‐rate (UHDR) particle radiotherapy beams is crucial for understanding and assessing the FLASH effect, towards pre‐clinical application and quality assurance programs.
Purpose
The purpose of the present work is to investigate a novel diamond‐based detector system for dose and instantaneous dose‐rate measurements in UHDR particle beams.
Methods
Two types of diamond detectors, a microDiamond (PTW 60019) and a diamond detector prototype specifically designed for operation in UHDR beams (flashDiamond), and two different readout electronic chains, were investigated for absorbed dose and instantaneous dose‐rate measurements. The detectors were irradiated with a helium beam of 145.7 MeV/u under conventional and UHDR delivery. Dose‐rate delivery records by the monitoring ionization chamber and diamond detectors were studied for single spot irradiations. Dose linearity at 5 cm depth and in‐depth dose response from 2 to 16 cm were investigated for both measurement chains and both detectors in a water tank. Measurements with cylindrical and plane‐parallel ionization chambers as well as Monte‐Carlo simulations were performed for comparisons.
Results
Diamond detectors allowed for recording the temporal structure of the beam, in good agreement with the one obtained by the monitoring ionization chamber. A better time resolution of the order of few μs was observed as compared to the approximately 50 μs of the monitoring ionization chamber. Both diamonds detectors show an excellent linearity response in both delivery modalities. Dose values derived by integrating the measured instantaneous dose‐rates are in very good agreement with the ones obtained by the standard electrometer readings. Bragg peak curves confirmed the consistency of the charge measurements by the two systems.
Conclusions
The proposed novel dosimetric system allows for a detailed investigation of the temporal evolution of UHDR beams. As a result, reliable and accurate determinations of dose and instantaneous dose‐rate are possible, both required for a comprehensive characterization of UHDR beams and relevant for FLASH effect assessment in clinical treatments.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Heidelberg Ion Beam Therapy Center (HIT), Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany.
- Clinical Cooperation Unit Translational Radiation Oncology, German Cancer Consortium (DKTK) Core-Center Heidelberg, National Center for Tumor Diseases (NCT), Heidelberg University Hospital (UKHD) and German Cancer Resea…
- Industrial Engineering Department, University of Rome Tor Vergata, Rome, Italy.
- Faculty of Physics, Karlsruhe Institute of Technology (KIT), Karlsruhe, Germany.
- PTW-Freiburg, Freiburg, Germany.
- University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.
Förderung
- NCI NIH HHS
Zitieren
APA
Tessonnier, T., Verona-Rinati, G., Rank, L., Kranzer, R., Mairani, A., & Marinelli, M. (2024). Diamond detectors for dose and instantaneous dose‐rate measurements for ultra‐high dose‐rate scanned helium ion beams. Medical Physics, 51(2), 1450-1459. https://doi.org/10.1002/mp.16757
Vancouver
Tessonnier T, Verona-Rinati G, Rank L, Kranzer R, Mairani A, Marinelli M. Diamond detectors for dose and instantaneous dose‐rate measurements for ultra‐high dose‐rate scanned helium ion beams. Medical Physics. 2024;51(2):1450-1459. doi:10.1002/mp.16757
BibTeX
@article{Tessonnier2024,
author = {Tessonnier, Thomas and Verona-Rinati, Gianluca and Rank, Luisa and Kranzer, Rafael and Mairani, Andrea and Marinelli, Marco},
title = {Diamond detectors for dose and instantaneous dose‐rate measurements for ultra‐high dose‐rate scanned helium ion beams},
journal = {Medical Physics},
year = {2024},
volume = {51},
number = {2},
pages = {1450--1459},
doi = {10.1002/mp.16757}
}
EUVIMED