Efficient estimation of gadolinium‐based contrast agent concentration using transient‐state keyhole MR‐STAT

Fei Xu, Edwin Versteeg, Hongyan Liu, Miha Fuderer, Oscar van den Heide, Wybe J M van der Kemp, Cornelis A T van den Berg, Alessandro Sbrizzi

Medical Physics · 2025

DOI: 10.1002/mp.70155 ↗

At a glance

Journal article in Medical Physics, 2025. Written by Fei Xu and 7 co-authors. Institution: Computational Imaging Group for MR Diagnostics & Therapy, Center for Image Sciences, UMC Utrecht, Utrecht, the Netherlands. Major MEDLINE topics: Contrast Media, Magnetic Resonance Imaging, Gadolinium, Organometallic Compounds. 45 references. The full text is freely available (open access, CC BY 4.0). Funded by: Nederlandse Organisatie voor Wetenschappelijk Onderzoek (NWO, Netherlands Organisation for Scientific Research).

PovzetekSubjectsBibliografski podatkiCiteRelated studies

Abstract (original language)

Background

Fast quantitative MRI (qMRI) techniques have emerged to quantify multiple tissue parameters from a single acquisition in clinically feasible scan times. To integrate fast qMRI techniques into a clinical MRI protocol, one needs to consider imaging after injection of a contrast agent (e.g., dynamic contrast‐enhanced [DCE] imaging). However, fully repeating the multi‐parametric scan after the injection requires lengthy scan times, which are impractical for tracking the contrast agent concentration.

Purpose

This work aimed to develop and validate an accelerated multi‐parametric method capable of estimating the concentration of Gadolinium‐based contrast agents (GBCAs) in a more time‐efficient way.

Methods

We designed an accelerated 2D Magnetic Resonance Spin TomogrAphy in Time‐domain (MR‐STAT) protocol and tested it on a 3T MRI scanner. This is achieved by combining transient‐state acquisitions with Cartesian keyhole‐based acceleration. Based on the reconstructed T 1 values, GBCA concentrations were quantified in phantoms containing water, manganese chloride, and five different concentrations of gadobutrol (Gadovist), ranging from 0.05 to 0.9 mM. Linear regression analysis between the estimated GBCA concentration and the reference values was performed to assess the accuracy. The GBCA concentration quantification method was then applied for synthetic patient data and a hybrid human/phantom study to demonstrate the feasibility of this method.

Results

The accelerated MR‐STAT protocol, with an undersampling factor of 4, provided accurate T 1 , T 2, and GBCA concentration measurements. Estimated GBCA concentrations were in a strong linear relationship with reference values, with a slope and intercept on simple linear regression analysis of 1.034 and 0.009, respectively.

Conclusion

We present a multi‐parametric approach for quantitatively assessing the concentration of gadolinium in vitro and in a hybrid in vivo/in vitro controlled setup using the accelerated MR‐STAT protocol. The concentration of gadobutrol in the range of 0.05–0.9 mM could be measured in a clinically applicable scan time using the proposed method, but needs further in vivo validation.

Subjects

MeSH terms (MEDLINE)

Contrast MediaMagnetic Resonance ImagingGadoliniumOrganometallic CompoundsHumansPhantoms, ImagingTime FactorsImage Processing, Computer-Assisted

Author keywords

MR‐STATconcentration estimationcontrast enhancementgadolinium‐based contrast agentquantitative imaging

Substances

Contrast MediaGadoliniumgadobutrolOrganometallic Compounds

Authors and institutions

  • Computational Imaging Group for MR Diagnostics & Therapy, Center for Image Sciences, UMC Utrecht, Utrecht, the Netherlands.
  • Department of Radiology, University Medical Center Utrecht, Utrecht, the Netherlands.

Funding

  • Nederlandse Organisatie voor Wetenschappelijk Onderzoek (NWO, Netherlands Organisation for Scientific Research)

Cite

APA

Xu, F., Versteeg, E., Liu, H., Fuderer, M., van den Heide, O., van der Kemp, W. J. M., van den Berg, C. A. T., & Sbrizzi, A. (2025). Efficient estimation of gadolinium‐based contrast agent concentration using transient‐state keyhole MR‐STAT. Medical Physics. https://doi.org/10.1002/mp.70155

Vancouver

Xu F, Versteeg E, Liu H, Fuderer M, van den Heide O, van der Kemp WJM, et al. Efficient estimation of gadolinium‐based contrast agent concentration using transient‐state keyhole MR‐STAT. Medical Physics. 2025. doi:10.1002/mp.70155

BibTeX
@article{Xu2025,
  author = {Xu, Fei and Versteeg, Edwin and Liu, Hongyan and Fuderer, Miha and van den Heide, Oscar and van der Kemp, Wybe J M and van den Berg, Cornelis A T and Sbrizzi, Alessandro},
  title = {Efficient estimation of gadolinium‐based contrast agent concentration using transient‐state keyhole MR‐STAT},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.70155}
}
Start for free →