Ultrahigh‐resolution K‐edge imaging for guidance of microrobots in minimally invasive interventions: A feasibility study

Arnaud R Brian-Choux, Thomas Wesley Holmes, Zhye Yin, Kimberly B Hoang, Aria Salyapongse, Azadeh Ansari, Amir Pourmorteza

Medical Physics · 2025

DOI: 10.1002/mp.70105 ↗

At a glance

Journal article in Medical Physics, 2025. Written by Arnaud R Brian-Choux and 6 co-authors. Institution: Department of Biomedical Engineering, Georgia Institute of Technology, Department of Radiology and Imaging Sciences, Winship Cancer Institute, Emory …. Major MEDLINE topics: Tomography, X-Ray Computed, Minimally Invasive Surgical Procedures, Robotics, Robotic Surgical Procedures. Cited 1 times so far (Crossref); 17 references.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Photon‐counting detector (PCD) CT offers enhanced spatial resolution, improved image contrast, reduced radiation dose, and material differentiation through K‐edge imaging. These features may be of value for guiding microrobots and surgical instruments during minimally invasive brain surgery.

Purpose

This study evaluates the feasibility of using K‐edge imaging with deep silicon (dSi) PCD CT to estimate the pose and location of microrobots within the human head, utilizing materials like neodymium (Nd) and tungsten (W) with distinct K‐edge energies.

Methods

A micro‐driller robot with a cubic Nd magnet (K‐edge energy = 43.5 keV) and a 3D‐printed drill bit was used. W (K‐edge energy = 69.5 keV) was added to aid in orientation detection. These materials were positioned at distances of 0–1 mm within a 3D‐printed setup and placed in a human skull filled with ballistic gel, alongside other metallic parts. The setup was scanned using a prototype dSi CT scanner with eight energy bins. Scanning parameters included 120 kVp, 300 mAs, 0.5 s rotation time, and 4‐cm z‐collimation.

Results

Nd and W components were distinguishable, with Nd showing higher contrast in low‐energy virtual monoenergetic images (VMIs) and W being more radiopaque in higher VMIs. The signal‐to‐noise ratios (SNR) at 70 keV and 1‐mm distance were 31.0 for Nd and 132.7 for W. The error in estimating the distance between components (Δ L ) ranged from 82 to 263 µm comparable to one image voxel.

Conclusions

Deep silicon PCD‐CT with K‐edge imaging successfully differentiated microrobot components in a human head phantom, indicating its potential for precision image‐guided robotic surgery. Further studies are needed to optimize radiation doses for clinical use.

Subjects

MeSH terms (MEDLINE)

Tomography, X-Ray ComputedMinimally Invasive Surgical ProceduresRoboticsRobotic Surgical ProceduresSurgery, Computer-AssistedFeasibility StudiesHumansSignal-To-Noise RatioPhantoms, ImagingPrinting, Three-Dimensional

Author keywords

K‐edge imagingphoton‐counting CTrobotic intervention

Authors and institutions

  • Department of Biomedical Engineering, Georgia Institute of Technology, Department of Radiology and Imaging Sciences, Winship Cancer Institute, Emory University, Atlanta, Georgia, USA.
  • CT Research, GE HealthCare, Niskayuna, New York, USA.
  • Department of Medical Physics, University of Wisconsin, Madison, Wisconsin, USA.

Cite

APA

Brian-Choux, A. R., Holmes, T. W., Yin, Z., Hoang, K. B., Salyapongse, A., Ansari, A., & Pourmorteza, A. (2025). Ultrahigh‐resolution K‐edge imaging for guidance of microrobots in minimally invasive interventions: A feasibility study. Medical Physics. https://doi.org/10.1002/mp.70105

Vancouver

Brian-Choux AR, Holmes TW, Yin Z, Hoang KB, Salyapongse A, Ansari A, et al. Ultrahigh‐resolution K‐edge imaging for guidance of microrobots in minimally invasive interventions: A feasibility study. Medical Physics. 2025. doi:10.1002/mp.70105

BibTeX
@article{BrianChoux2025,
  author = {Brian-Choux, Arnaud R and Holmes, Thomas Wesley and Yin, Zhye and Hoang, Kimberly B and Salyapongse, Aria and Ansari, Azadeh and Pourmorteza, Amir},
  title = {Ultrahigh‐resolution K‐edge imaging for guidance of microrobots in minimally invasive interventions: A feasibility study},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.70105}
}
Start for free →