Evaluation of a CdTe semiconductor‐based gamma camera for real‐time dose dosimetry in boron neutron capture therapy

I-Huan Chiu, Takahito Osawa, Takehiro Sumita, Kazuhiko Ninomiya, Shin'ichiro Takeda, Tadayuki Takahashi

Medical Physics · 2025

DOI: 10.1002/mp.70062 ↗

At a glance

Evaluationsstudie in Medical Physics, 2025. Written by I-Huan Chiu and 5 co-authors. Institution: Materials Sciences Research Center, Japan Atomic Energy Agency, Ibaraki, Japan. Major MEDLINE topics: Boron Neutron Capture Therapy, Tellurium, Cadmium Compounds, Radiometry. Cited 1 times so far (Crossref); 41 references. The full text is freely available (open access, CC BY 4.0). Funded by: JSPS, Japan Atomic Energy Agency, Japan Society for the Promotion of Science.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Boron neutron capture therapy (BNCT) is a cancer treatment that leverages the nuclear reaction between boron‐10 () and thermal neutrons to generate high‐energy particles and nuclei that selectively destroy cancer cells while sparing healthy tissues. However, BNCT is limited by current dosimetry methods that are incapable of monitoring boron distribution during therapy. An imaging system capable of real‐time dosimetry is essential for optimizing treatment efficacy and minimizing collateral damage.

Purpose

This study aimed to develop a high‐resolution real‐time boron dosimetry system for BNCT by employing a cadmium telluride double‐sided strip detector (CdTe‐DSD). The CdTe‐DSD enables precise mapping of distribution by detecting the 478 keV prompt ‐rays emitted during the reaction.

Methods

The imaging system was constructed by integrating a CdTe‐DSD with a 2‐mm diameter pinhole collimator. The CdTe‐DSD, comprising a 2‐mm‐thick CdTe semiconductor, has a sensitive area of 32 32 . Neutron irradiation experiments were performed at Japan Research Reactor No. 3 using various boron‐containing samples, including boric acid solutions, powders, and granular boron, with boron masses ranging from 0.02 to 2.00 mg. We implemented the neutron shields using a 5‐mm‐thick plate and LiF tiles to reduce the background from scattered neutrons during the measurement.

Results

The imaging system successfully detected the 478 keV ‐ray signal with an energy resolution of 7.3 keV at 511 keV. The reconstructed two‐dimensional images demonstrate the capability of the CdTe‐DSD to accurately map the distribution in the samples. Notably, we evaluated the important background contributions from the scattering neutrons in the development of the CdTe‐DSD‐based imaging system. When the scattering neutrons in a neutron experiment hit the CdTe‐DSD, 558 keV ‐rays of were produced. Under three experimental conditions with different neutron shielding configurations, we found that enhanced neutron shielding considerably reduced the background contribution created by the background signal, thereby improving the contrast‐to‐noise ratio (CNR) in the image quality assessment. Upon comparing the condition with minimal neutron shielding to that with maximal shielding, a 14‐fold enhancement in the CNR was observed.

Conclusions

This study demonstrates that a CdTe‐DSD‐based gamma camera is a promising tool for real‐time boron dosimetry in BNCT. The detector's high energy resolution and excellent spatial resolution enable precise detection of 478 keV prompt ‐rays, facilitating accurate mapping of boron distribution during neutron irradiation. These findings support the developed system's potential to enhance BNCT treatment planning and patient‐specific dose monitoring. Future research will optimize neutron shielding and explore advanced collimator designs to improve sensitivity in clinical settings.

Subjects

MeSH terms (MEDLINE)

Boron Neutron Capture TherapyTelluriumCadmium CompoundsRadiometrySemiconductorsGamma CamerasRadiation DosageTime FactorsRadiotherapy Dosage

Author keywords

CdTe‐DSDboron neutron capture therapyreal‐time dosimetry

Substances

Telluriumcadmium tellurideCadmium Compounds

Authors and institutions

  • Materials Sciences Research Center, Japan Atomic Energy Agency, Ibaraki, Japan.
  • Department of Materials, Kyushu University, Fukuoka, Japan.
  • Natural Science Center for Basic Research and Development, Hiroshima University, Higashi-Hiroshima, Hiroshima, Japan.
  • Fukushima Institute for Research, Education and Innovation (F-REI), Fukushima, Japan.
  • Kavli Institute for the Physics and Mathematics of the Universe (WPI), The University of Tokyo, Kashiwa, Chiba, Japan.
  • International Center for Quantum-field Measurement Systems for Studies of the Universe and Particles (QUP, WPI), High Energy Accelerator Research Organization (KEK), Tsukuba, Ibaraki, Japan.

Funding

  • JSPS
  • Japan Atomic Energy Agency
  • Japan Society for the Promotion of Science

Cite

APA

Chiu, I. H., Osawa, T., Sumita, T., Ninomiya, K., Takeda, S., & Takahashi, T. (2025). Evaluation of a CdTe semiconductor‐based gamma camera for real‐time dose dosimetry in boron neutron capture therapy. Medical Physics. https://doi.org/10.1002/mp.70062

Vancouver

Chiu IH, Osawa T, Sumita T, Ninomiya K, Takeda S, Takahashi T. Evaluation of a CdTe semiconductor‐based gamma camera for real‐time dose dosimetry in boron neutron capture therapy. Medical Physics. 2025. doi:10.1002/mp.70062

BibTeX
@article{Chiu2025,
  author = {Chiu, I-Huan and Osawa, Takahito and Sumita, Takehiro and Ninomiya, Kazuhiko and Takeda, Shin'ichiro and Takahashi, Tadayuki},
  title = {Evaluation of a CdTe semiconductor‐based gamma camera for real‐time dose dosimetry in boron neutron capture therapy},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.70062}
}
Start for free →