Real‐time monitoring and appropriate margin selection in linac‐based prostate SBRT: Analysis of two different intrafraction motion monitoring systems
Medical Physics · 2025-08 · 52 (8)
DOI: 10.1002/mp.18060 ↗At a glance
Journal article in Medical Physics, published online on 2025-08-13. Written by Denis Panizza and 7 co-authors. Institution: Medical Physics, Fondazione IRCCS San Gerardo dei Tintori, Monza, Italy. Major MEDLINE topics: Radiosurgery, Prostatic Neoplasms, Dose Fractionation, Radiation, Particle Accelerators. Cited 5 times so far (Crossref); 35 references.
Abstract (original language)
Background
Intrafraction motion poses a major challenge in ultrahypofractionated prostate SBRT, where high doses and fewer sessions amplify the impact of positional shifts. Its management varies across centers due to differences in technology, protocols, and resources, and standardized clinical guidelines are still lacking despite widespread recognition of its importance.
Purpose
To provide insights into real‐time monitoring and margin selection by quantifying intrafraction prostate motion using two distinct monitoring systems—system A (electromagnetic transmitter‐based) and system B (4D transperineal ultrasound‐based)—and evaluate their impact on motion patterns in Linac‐based SBRT.
Methods
Intrafraction motion data were collected from 92 treatment fractions using system A (47 patients) and 100 fractions using system B (30 patients). All patients underwent SBRT (1–5 fractions) with strict rectum and bladder preparation. Motion was recorded at 1 Hz (system A) and 2 Hz (system B) from CBCT acquisition to treatment completion, independent of any patient repositioning. Motion trend was analyzed using a mixed‐effects regression model, and Kaplan–Meier failure analysis estimated the time to exceed 2, 3, 4, and 5 mm motion thresholds.
Results
Median monitoring time was 13 min for system A (range: 5.9–35.6) and 6.4 min for system B (range: 4.9–29.3). Motion increased over time in all directions ( p < 0.001), predominantly in the posterior direction (time coefficient = −0.20, p < 0.001). A significant longitudinal (LNG) motion divergence was observed between systems ( p < 0.001). Kaplan–Meier failure curves were comparable, except for a significant difference in the LNG axis at the 2 mm threshold (log‐rank test, p = 0.006). The failure rates at 3–10 min for this threshold ranged from 7.3% to 52.4% (system A) and 4.2% to 22.4% (system B). A substantial percentage of fractions (> 5%) exceeded the 2 mm motion threshold at 1.8 min, the 3 mm threshold at 3.8 min, the 4 mm threshold at 6.6 min, and the 5 mm threshold at 8.1 min.
Conclusions
Both systems exhibited predominant posterior motion, with system B reducing small LNG prostate fluctuations, likely due to probe pressure against the perineum. For fast Linac‐based SBRT, intrafraction motion can be managed with 4–5 mm margins without continuous monitoring. However, if 2–3 mm margins are used, real‐time monitoring becomes essential.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Medical Physics, Fondazione IRCCS San Gerardo dei Tintori, Monza, Italy.
- School of Medicine and Surgery, University of Milan Bicocca, Milan, Italy.
- Department of Physics, University of Milan, Milan, Italy.
- Radiation Oncology, Fondazione IRCCS San Gerardo dei Tintori, Monza, Italy.
Cite
APA
Panizza, D., Faccenda, V., Ingraito, C., Chissotti, C., Ferrario, F., Colciago, R. R., De Ponti, E., & Arcangeli, S. (2025). Real‐time monitoring and appropriate margin selection in linac‐based prostate SBRT: Analysis of two different intrafraction motion monitoring systems. Medical Physics, 52(8). https://doi.org/10.1002/mp.18060
Vancouver
Panizza D, Faccenda V, Ingraito C, Chissotti C, Ferrario F, Colciago RR, et al. Real‐time monitoring and appropriate margin selection in linac‐based prostate SBRT: Analysis of two different intrafraction motion monitoring systems. Medical Physics. 2025;52(8). doi:10.1002/mp.18060
BibTeX
@article{Panizza2025,
author = {Panizza, Denis and Faccenda, Valeria and Ingraito, Chiara and Chissotti, Chiara and Ferrario, Federica and Colciago, Riccardo Ray and De Ponti, Elena and Arcangeli, Stefano},
title = {Real‐time monitoring and appropriate margin selection in linac‐based prostate SBRT: Analysis of two different intrafraction motion monitoring systems},
journal = {Medical Physics},