EUVIMEDOpdag EUVIMED
Open access · CC BY-NC-ND 4.0

Commissioning of a 142.4 MeV ultra‐high dose rate (UHDR) proton beamline in a synchrotron‐based proton therapy system

Lingshu Yin, Daniel Sforza, Devin Miles, Umezawa Masumi, Kan Ota, Xun Jia, Heng Li

Medical Physics · 2025

DOI: 10.1002/mp.18008 ↗

At a glance

Journal article in Medical Physics, 2025. Written by Lingshu Yin and 6 co-authors. Institution: Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, Maryland, USA. Major MEDLINE topics: Proton Therapy, Synchrotrons, Radiation Dosage. Cited 4 times so far (Crossref); 39 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: NCI NIH HHS, NIH HHS, National Institutes of Health.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Recent studies suggest that radiotherapy at ultrahigh dose rates (>40 Gy/s, FLASH) offers normal tissue sparing effects while maintaining tumor control. There is significant interest in preclinical studies investigating the mechanism of FLASH sparing effects.

Purpose

This study aims to commission a fixed proton beamline within a synchrotron‐based proton therapy system for preclinical proton FLASH research.

Methods

Modifications were made to the Hitachi PROBEAT‐CR synchrotron system to enhance RF extraction power and increase proton beam current at 142.4 MeV. A high‐speed electrometer and an optimized transmission ion chamber (IC) were implemented for ultra‐high dose rate (UHDR) beam monitoring and delivery, replacing the conventional beam monitoring IC. Beam output was measured using a Faraday cup in both UHDR and clinical modes. Gafchromic film measurements and Monte Carlo simulations were employed to validate dose delivery in a solid water phantom with various spot scanning patterns.

Results

The calibration of transmission IC against Faraday cup shows sufficient charge collection efficiency at both clinical dose rates and UHDR. The UHDR PBS beamline demonstrates better than 1% reproducibility and linearity in the absolute beam output. Due to the limited charge per spill, the delivered dose per spill is inversely proportional to the field size. However, the system can deliver up to 41.4 Gy (268.1 Gy/sec) at 2 cm depth with a field size (FWHM) of 8.2 mm, demonstrating suitability for small animal proton FLASH irradiation studies.

Conclusion

We successfully commissioned a fixed beam proton UHDR PBS beamline in a synchrotron‐based proton therapy system. Despite synchrotron‐specific system constraints, our system enables controlled UHDR delivery for preclinical proton FLASH research.

Subjects

MeSH terms (MEDLINE)

Proton TherapySynchrotronsRadiation DosageRadiotherapy DosageMonte Carlo MethodPhantoms, ImagingCalibration

Author keywords

commissioningproton dosimetryultra‐high dose rate

Authors and institutions

  • Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
  • Department of Radiation Oncology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Center for Technology Innovation-Energy, Research and Development Group, Hitachi, Ltd., Hitachi-shi, Ibaraki-ken, Japan.
  • Pyramid Technical Consultants, Inc., Boston, Massachusetts, USA.

Funding

  • NCI NIH HHS
  • NIH HHS
  • National Institutes of Health

Cite

APA

Yin, L., Sforza, D., Miles, D., Masumi, U., Ota, K., Jia, X., & Li, H. (2025). Commissioning of a 142.4 MeV ultra‐high dose rate (UHDR) proton beamline in a synchrotron‐based proton therapy system. Medical Physics. https://doi.org/10.1002/mp.18008

Vancouver

Yin L, Sforza D, Miles D, Masumi U, Ota K, Jia X, et al. Commissioning of a 142.4 MeV ultra‐high dose rate (UHDR) proton beamline in a synchrotron‐based proton therapy system. Medical Physics. 2025. doi:10.1002/mp.18008

BibTeX
@article{Yin2025,
  author = {Yin, Lingshu and Sforza, Daniel and Miles, Devin and Masumi, Umezawa and Ota, Kan and Jia, Xun and Li, Heng},
  title = {Commissioning of a 142.4 MeV ultra‐high dose rate (UHDR) proton beamline in a synchrotron‐based proton therapy system},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.18008}
}
Start for free →