EUVIMEDEUVIMED entdecken
Crossref

Anti‐jamming thermoacoustic imaging based on fiber Bragg grating ultrasonic detection and photoelectric conversion triggering

Yulong Guo, Lin Huang, Tiantao Hu, Wanting Peng, Kaili Jin, Liuge Cui, Ziqi Zhou, Yan Luo, Wenwu Ling

Medical Physics · 2025

DOI: 10.1002/mp.17944 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Yulong Guo und 8 weiteren Autor:innen. Beteiligte Einrichtung: School of Electronic Science and Engineering, University of Electronic Science and Technology of China, Chengdu, Sichuan, China. Hauptthemen laut MEDLINE: Optical Fibers, Temperature. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 58 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Key Research and Development Program of Science and Technology Department of Sichuan Province, Sichuan Provincial Health Commission Innovation Team Project.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Thermoacoustic Imaging (TAI) combines the high contrast of microwave imaging with the high resolution of ultrasound imaging, establishing itself as a novel, non‐invasive medical diagnostic technique. However, the high‐power (peak power: 5∼100 kW; pulse width: 10∼1000 ns) used in TAI often cause remarkable interference with thermoacoustic signal acquisition systems and traditional piezoelectric ultrasound transducers. This interference leads to degraded thermoacoustic image quality and a pronounced “near‐field dead zone” (NFDZ), which severely restricts the clinical application and development of TAI.

Purpose

To address these challenges, this paper proposes using fiber Bragg grating (FBG) for thermoacoustic signal detection. The high sensitivity and strong resistance to electromagnetic interference offered by FBG are leveraged to achieve interference‐resistant signal acquisition. Additionally, a photoelectric conversion synchronous triggering method is adopted to prevent strong coupling of spatially distributed pulse microwave signals to the data acquisition system due to excessively long coaxial cables. This approach ensures more reliable and interference‐resistant data acquisition for TAI.

Methods

Utilizing FBG with their high sensitivity and electromagnetic interference resistance, the system replaces traditional piezoelectric transducers for thermoacoustic signal detection. Additionally, a photoelectric conversion triggering method was implemented to avoid microwave pulse coupling interference caused by long coaxial cables. Experimental validation included comparisons between FBGs and traditional transducers in terms of signal‐to‐noise ratio (SNR), as well as noise measurements using both coaxial bayonet nut connector (BNC) cable and photoelectric trigger configurations to contrast test noise levels. Furthermore, the distance between the microwave antenna and data acquisition system was adjusted to validate the attenuation pattern of electromagnetic interference. TAI experiments used soy sauce tubes for their controllable composition/morphology. Systems imaging the tube also apply to human arms and ex vivo porcine liver (similar microwave absorption to tumors). This standardized, cost‐efficient, and ethical approach provides a reliable basis for biological tissue imaging research.

Results

Experimental results show FBG with photoelectric triggering achieves SNR values of 28.32 (70.2% improvement) and 30.74 dB (66.3% improvement) compared to traditional transducers at 16.64 and 18.48 dB, respectively, representing an overall SNR enhancement of 64.0%∼88.0%. Furthermore, the photoelectric conversion triggering method reduces noise levels, which means shortening the NFDZ, leading to improved imaging quality and an extended effective detection range. Additionally, it was found that increasing the distance between the microwave antenna and the data acquisition system effectively reduced the noise amplitude, with the attenuation of direct spatial coupling interference conforming to the electric field intensity attenuation law (1/r attenuation).

Conclusion

This study establishes that the FBG‐based thermoacoustic imaging system, leveraging FBG's high sensitivity and electromagnetic interference resistance combined with photoelectric triggering, enhances signal quality and effectively suppresses near‐field artifacts, enabling reliable clinical imaging. Validation through noise, signal, and spatial analyses confirms FBG's utility in advancing TAI. In summary, the method proposed in this article can be used for thermoacoustic endoscopy (especially for intravascular thermoacoustic endoscopy), thyroid and breast ablation TAI monitoring, as well as multimodal imaging combined with MRI and TAI.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Optical FibersTemperatureUltrasonographyPhantoms, ImagingMicrowavesSignal-To-Noise Ratio

Stichwörter der Autor:innen

anti‐jammingfiber Bragg gratingnear‐field dead zonephotoelectric conversion triggerthermoacoustic imaging

Autor:innen und Institutionen

  • School of Electronic Science and Engineering, University of Electronic Science and Technology of China, Chengdu, Sichuan, China.
  • School of Information Engineering, Southwest University of Science and Technology, Mianyang, Sichuan, China.
  • Guizhou Aerospace Institute of Measuring and Testing Technology, Guiyang, Guizhou, China.
  • Department of Ultrasound, West China Hospital, Sichuan University, Chengdu, Sichuan, China.
  • Department of Biomedical Engineering, Case Western Reserve University, Cleveland, USA.

Förderung

  • National Natural Science Foundation of China
  • Key Research and Development Program of Science and Technology Department of Sichuan Province
  • Sichuan Provincial Health Commission Innovation Team Project

Zitieren

APA

Guo, Y., Huang, L., Hu, T., Peng, W., Jin, K., Cui, L., Zhou, Z., Luo, Y., & Ling, W. (2025). Anti‐jamming thermoacoustic imaging based on fiber Bragg grating ultrasonic detection and photoelectric conversion triggering. Medical Physics. https://doi.org/10.1002/mp.17944

Vancouver

Guo Y, Huang L, Hu T, Peng W, Jin K, Cui L, et al. Anti‐jamming thermoacoustic imaging based on fiber Bragg grating ultrasonic detection and photoelectric conversion triggering. Medical Physics. 2025. doi:10.1002/mp.17944

BibTeX
@article{Guo2025,
  author = {Guo, Yulong and Huang, Lin and Hu, Tiantao and Peng, Wanting and Jin, Kaili and Cui, Liuge and Zhou, Ziqi and Luo, Yan and Ling, Wenwu},
  title = {Anti‐jamming thermoacoustic imaging based on fiber Bragg grating ultrasonic detection and photoelectric conversion triggering},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.17944}
}
Kostenlos starten →