EUVIMEDOpdag EUVIMED
Open access · CC BY 4.0

Characterization of a Commercial Ionization Chamber Array With Scanned Proton Beams for Applications in MRI‐Guided Proton Therapy

Benjamin Gebauer, Sebastian Gantz, Daniela Kunath, Aswin Hoffmann, Jörg Pawelke, Felix Horst

Medical Physics · 2025

DOI: 10.1002/mp.17875 ↗

At a glance

Journal article in Medical Physics, 2025. Written by Benjamin Gebauer and 5 co-authors. Institution: OncoRay - National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität D…. Major MEDLINE topics: Proton Therapy, Magnetic Resonance Imaging, Radiotherapy, Image-Guided. Cited 4 times so far (Crossref); 37 references. The full text is freely available (open access, CC BY 4.0).

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

The integration of MRI‐guidance and proton therapy is a current research topic. Proton therapy with the patient being placed inside an in‐beam MR scanner would require the presence of its static magnetic () field to be taken into account in dose calculation and treatment planning. Therefore, dosimetric tools are needed to characterize dose distributions in presence of the field of the MR scanner. Furthermore, patient‐specific quality assurance (QA) and treatment plan verification measurements should also be performed within the magnetic field.

Purpose

In this work, the PTW Octavius 1500 ionization chamber array was characterized experimentally and tested for its suitability as a dosimetric tool for beam characterization and QA in MRI‐guided proton therapy.

Methods

The dose rate response, response homogeneity and effective measurement depth of the detector were determined in experiments with scanned proton beams delivered by a horizontal beamline at OncoRay, Dresden. A patient‐specific QA test including gamma analysis was performed for a realistic proton patient treatment plan at two different distances from the beam nozzle. In addition, experiments were performed in a in‐beam MR scanner. These included measurements of square reference scanning patterns at different proton energies as well as measurements of a two‐field patient treatment plan at different water equivalent depths.

Results

The dose rate response was found to be linear up to . The effective measurement depth was determined to be . The response homogeneity was found to be suitable for the verification of proton treatment plans. The patient‐specific QA test without magnetic field was satisfactory and also the measurements inside the in‐beam MR scanner provided reasonable results. Their comparison allowed an assessment of the magnetic field effects on the dose distributions.

Conclusions

Concluding from these tests, the Octavius 1500 was found to be suitable for use as a dosimetric tool in MRI‐guided proton therapy.

Subjects

MeSH terms (MEDLINE)

Proton TherapyMagnetic Resonance ImagingRadiotherapy, Image-GuidedRadiometryHumansRadiotherapy DosageRadiotherapy Planning, Computer-Assisted

Author keywords

MRI‐guidanceionization chamber arrayproton therapy

Authors and institutions

  • OncoRay - National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden and Helmholtz-Zentrum Dresden-Rossendorf, Dresden, Germany.
  • Helmholtz-Zentrum Dresden-Rossendorf, Institute of Radiooncology - OncoRay, Dresden, Germany.
  • Department of Radiotherapy and Radiation Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Dresden, Germany.

Cite

APA

Gebauer, B., Gantz, S., Kunath, D., Hoffmann, A., Pawelke, J., & Horst, F. (2025). Characterization of a Commercial Ionization Chamber Array With Scanned Proton Beams for Applications in MRI‐Guided Proton Therapy. Medical Physics. https://doi.org/10.1002/mp.17875

Vancouver

Gebauer B, Gantz S, Kunath D, Hoffmann A, Pawelke J, Horst F. Characterization of a Commercial Ionization Chamber Array With Scanned Proton Beams for Applications in MRI‐Guided Proton Therapy. Medical Physics. 2025. doi:10.1002/mp.17875

BibTeX
@article{Gebauer2025,
  author = {Gebauer, Benjamin and Gantz, Sebastian and Kunath, Daniela and Hoffmann, Aswin and Pawelke, Jörg and Horst, Felix},
  title = {Characterization of a Commercial Ionization Chamber Array With Scanned Proton Beams for Applications in MRI‐Guided Proton Therapy},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.17875}
}
Start for free →