Characterization of commercial detectors for absolute proton UHDR dosimetry on a compact clinical proton synchrocyclotron

Jufri Setianegara, Aoxiang Wang, Nicolas Gerard, Jarrick Nys, H Harold Li, Ronald C Chen, Hao Gao, Yuting Lin

Medical Physics · 2025-07 · 52 (7)

DOI: 10.1002/mp.17847 ↗

At a glance

Journal article in Medical Physics, published online on 2025-04-23. Written by Jufri Setianegara and 7 co-authors. Institution: Department of Radiation Oncology, University of Kansas Medical Center, Kansas City, Kansas, USA. Major MEDLINE topics: Radiometry, Protons, Radiation Dosage, Proton Therapy. Cited 9 times so far (Crossref); 79 references. The full text is freely available in PubMed Central. Funded by: NIH HHS, NCI NIH HHS, National Institutes of Health.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Background

Modern compact proton synchrocyclotrons can achieve ultra‐high dose rates (40 Gy/s) to support ultra‐high‐dose‐rate (UHDR) preclinical experiments utilizing pencil beam scanning (PBS) protons. Unique to synchrocyclotrons is a pulsed proton time structure as compared to the quasi‐continuous nature of other proton accelerators like isochronous cyclotrons. Thus, high instantaneous proton currents in the order of several µA must be generated to achieve UHDRs. This will lead to high doses‐per‐pulse (DPP), which may cause significant charge recombination for ionization chambers, which must be characterized for accurate UHDR dosimetry programs.

Purpose

In this work, we investigate the suitability of various commercial radiation detectors for accurate proton UHDR dosimetry using PBS proton beams from a compact proton synchrocyclotron (IBA ProteusONE). This is achieved by cross‐calibrating them with conventional dose rates, measuring UHDR recombination (P ion ) and polarity correction factors (P pol ) for ionization chambers, and determining the absorbed proton UHDR dose delivered for all detectors.

Methods

An IBA ProteusONE synchrocyclotron was initially tuned to achieve UHDRs with 228 MeV protons at 0° gantry angle. Various detectors, including Razor Chamber, Razor Nano Chamber, Razor Diode, and microDiamond, were cross‐calibrated against a PPC05 plane‐parallel ionization chamber (PPIC) that had an ADCL calibration coefficient of 59.23 cGy/nC. Then, all ionization chambers were exposed to UHDR protons with the P pol and P ion subsequently calculated. P ion was calculated using two methods: TRS‐398 methods and Niatel's model. Finally, the absolute UHDR proton doses delivered were determined for all detectors and cross‐compared.

Results

Faraday cup measurements were performed for a single spot proton UHDR beam, and the nozzle current at the isocenter was determined to be 129.5 nA during UHDR irradiations at 98.61% of the maximum theoretical dose rate. Repeated Faraday cup measurements of the UHDR beam yielded a percentage standard deviation of 0.8%, which was higher than 0.120% when similar repeated measurements were performed with conventional proton beams. P pol was found to be relatively dose‐rate independent for all ionization chambers investigated. P ion was found to be the lowest for the PPC05 ionization chamber (1.0097) compared to corresponding values of 1.0214 and 1.0294 for the Razor and Razor Nano detectors, respectively, for UHDRs. P ion values calculated using Niatel's model closely matched values from TRS‐398 if the V H /V L ratio were kept at 2.5 for the PPC05 and Razor detectors and 2.0 for the Razor Nano detector. Absolute proton UHDR doses determined using cross‐calibration factors were generally within ± 1% of PPC05 measurements. However, Razor Diode was found to over‐respond by up to 3.79% within UHDR proton beams, rendering them unsuitable for proton UHDR dosimetry.

Conclusion

In this work, we comprehensively evaluated the suitability of various commercial detectors for absolute dosimetry with a pulsed UHDR beam structure from a proton synchrocyclotron. PPC05 had the lowest ionic recombination correction compared to Razor and Razor Nano ion chambers. Other than the diode detector, all other investigated detectors (PPC05, Razor, Razor Nano, microDiamond) were within ± 1% of one another and can be used for accurate absolute proton UHDR dosimetry.

Subjects

MeSH terms (MEDLINE)

RadiometryProtonsRadiation DosageProton Therapy

Author keywords

Boag theoryIAEA TRS‐398Niatel modelcompact ionization chambersdiode detectorsion recombinationionization chambersmicroDiamond detectorspencil beam scanning (PBS)plane‐parallel ionization chambers (PPIC)proton FLASHpulsed beamultra‐high‐dose‐rate

Substances

Protons

Authors and institutions

  • Department of Radiation Oncology, University of Kansas Medical Center, Kansas City, Kansas, USA.
  • Department of Biomedical Engineering, Huazhong University of Science and Technology, Wuhan, China.
  • Ion Beam Applications (IBA), Louvain-la-Neuve, Belgium.

Funding

  • NIH HHS
  • NCI NIH HHS
  • National Institutes of Health

Cite

APA

Setianegara, J., Wang, A., Gerard, N., Nys, J., Harold Li, H., Chen, R. C., Gao, H., & Lin, Y. (2025). Characterization of commercial detectors for absolute proton UHDR dosimetry on a compact clinical proton synchrocyclotron. Medical Physics, 52(7). https://doi.org/10.1002/mp.17847

Vancouver

Setianegara J, Wang A, Gerard N, Nys J, Harold Li H, Chen RC, et al. Characterization of commercial detectors for absolute proton UHDR dosimetry on a compact clinical proton synchrocyclotron. Medical Physics. 2025;52(7). doi:10.1002/mp.17847

BibTeX
@article{Setianegara2025,
  author = {Setianegara, Jufri and Wang, Aoxiang and Gerard, Nicolas and Nys, Jarrick and Harold Li, H and Chen, Ronald C and Gao, Hao and Lin, Yuting},
  title = {Characterization of commercial detectors for absolute proton UHDR dosimetry on a compact clinical proton synchrocyclotron},
  journal = {Medical Physics},
  year = {2025},
  volume = {52},
  number = {7},
  doi = {10.1002/mp.17847}
}
Start for free →