EUVIMEDUpptäck EUVIMED
Crossref

Impact of porous lung substitute on linear energy transfer (LET) assessed via Monte Carlo simulation and CR‐39 measurement with a carbon‐ion beam

Jingfang Zhao iD, Bo Chen, Michael Moyers, Zhiyuan Yang, Shiyan Yang, Weiwei Wang

Medical Physics · 2025

DOI: 10.1002/mp.17671 ↗

At a glance

Journal article in Medical Physics, 2025. Written by Jingfang Zhao and 5 co-authors. Institution: Department of Medical Physics, Shanghai Proton and Heavy Ion Center, Fudan University Cancer Hospital, Shanghai, People's Republic of China. Major MEDLINE topics: Monte Carlo Method, Linear Energy Transfer, Lung, Heavy Ion Radiotherapy. Cited 2 times so far (Crossref); 31 references. Funded by: Natural Science Foundation of Shanghai, China, Shanghai Pudong New Area Health Commission.

AbstractSubjectsBibliografiska uppgifterCiteRelated studies

Abstract (original language)

Background

Carbon‐ion beam radiotherapy offers substantial physical and biological advantages due to its distinct Bragg peak (BP) depth dose distribution and higher linear energy transfer (LET) in the peak region that enhances its efficacy in tumor eradication compared to x‐ray beams. Porous structures, such as those found in lung and lung‐equivalent tissues, unfortunately, introduce significant uncertainties in both dose and LET distributions, which current treatment planning systems (TPS) inadequately address.

Purpose

This study aims to investigate the effects of porous lung‐equivalent structures on LET distribution using Monte Carlo (MC) simulations and CR‐39 measurements. It seeks to understand how porous structures influence LET spectra and dose‐averaged LET (LET d ) in carbon‐ion beams.

Methods

A Gammex LN300 phantom and a binary voxel virtual phantom composed of water and air were used to represent lung‐equivalent tissues for measurements and MC simulations. LET spectra measured with CR‐39 at different depths within the LN300 slabs were compared with MC‐calculated LET d distributions. The impact of porous structures on dose and LET d distributions was evaluated using various beam configurations, including single‐beam and multi‐beam setups. Additionally, a convolution method with modulation power ( P mod ) was proposed to improve LET d prediction in porous media.

Results

The study demonstrated that porous structures broaden both the dose and LET d distributions, especially around the BP region. Multiple beam angles helped mitigate dose degradation but did not resolve discrepancies in the LET d distributions. Compared with calculation results based on CT images, intensity‐modulated particle therapy (IMPT) using a distal LET d patching method in porous structure increased the median LET d in the target from 67.2 to 69.6 keV/µm, and the minimum LET d from 51.5 to 58.0 keV/µm, respectively. Moreover, to improve the prediction of LET d in porous structures, analytical convolution‐based predictions showed good agreement with the MC simulations, with mean LET d deviations of −1.9% ± 1.6% in the plateau, −3.1% ± 4.9% in the BP, and −1.1% ± 7.7% in the tail region.

Conclusions

Porous lung‐equivalent structures significantly affect LET d distributions in carbon‐ion therapy, as confirmed by both CR‐39 measurements and MC simulations. IMPT with LET d optimization may be more impacted by porous structures in terms of median and minimum LET d values within the target. The Gaussian convolution function shows promise for enhancing LET d calculation accuracy, but further validation in anatomically complex models is needed to assess its clinical feasibility.

Subjects

MeSH terms (MEDLINE)

Monte Carlo MethodLinear Energy TransferLungHeavy Ion RadiotherapyPorosityPhantoms, ImagingHumansRadiotherapy Planning, Computer-AssistedRadiotherapy Dosage

Author keywords

CR‐39LETdMonte Carlocarbon‐ion beamlung‐equivalent tissueporous

Authors and institutions

  • Department of Medical Physics, Shanghai Proton and Heavy Ion Center, Fudan University Cancer Hospital, Shanghai, People's Republic of China.
  • Shanghai Key Laboratory of Radiation Oncology, Shanghai, People's Republic of China.
  • Shanghai Engineering Research Center of Proton and Heavy Ion Radiation Therapy, Shanghai, People's Republic of China.
  • Institute of Radiation Medicine, Fudan University, Shanghai, People's Republic of China.
  • Institute of Modern Physics, Fudan University, Shanghai, People's Republic of China.

ORCID profiles: Jingfang Zhao

Funding

  • Natural Science Foundation of Shanghai, China
  • Shanghai Pudong New Area Health Commission

Cite

APA

Zhao, J., Chen, B., Moyers, M., Yang, Z., Yang, S., & Wang, W. (2025). Impact of porous lung substitute on linear energy transfer (LET) assessed via Monte Carlo simulation and CR‐39 measurement with a carbon‐ion beam. Medical Physics. https://doi.org/10.1002/mp.17671

Vancouver

Zhao J, Chen B, Moyers M, Yang Z, Yang S, Wang W. Impact of porous lung substitute on linear energy transfer (LET) assessed via Monte Carlo simulation and CR‐39 measurement with a carbon‐ion beam. Medical Physics. 2025. doi:10.1002/mp.17671

BibTeX
@article{Zhao2025,
  author = {Zhao, Jingfang and Chen, Bo and Moyers, Michael and Yang, Zhiyuan and Yang, Shiyan and Wang, Weiwei},
  title = {Impact of porous lung substitute on linear energy transfer (LET) assessed via Monte Carlo simulation and CR‐39 measurement with a carbon‐ion beam},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.17671}
}
Start for free →