4D Dynamic contrast‐enhanced breast CT: Phantom‐based reconstruction parameter optimization for iodine quantification

Mikhail Mikerov, Juan J Pautasso, Liselot Goris, Koen Michielsen, Ioannis Sechopoulos

Medical Physics · 2025

DOI: 10.1002/mp.17658 ↗

At a glance

Journal article in Medical Physics, 2025. Written by Mikhail Mikerov and 4 co-authors. Institution: Department of Medical Imaging, Radboudumc, Nijmegen, The Netherlands. Major MEDLINE topics: Phantoms, Imaging, Iodine, Contrast Media, Breast. Cited 4 times so far (Crossref); 32 references. The full text is freely available (open access, CC BY-NC 4.0). Funded by: European Union's Horizon 2020 research and innovation programme, European Research Council.

AbstractSubjectsDati bibliograficiCiteRelated studies

Abstract (original language)

Background

Four‐dimensional dynamic contrast‐enhanced breast CT (4D DCE‐bCT) offers promising high‐resolution spatial and temporal imaging capabilities for the characterization and monitoring of breast tumors. However, the optimal combination of parameters for iodine quantification in image space remains to be determined.

Purpose

This study aims to optimize a dedicated bCT system to perform long dynamic contrast‐enhanced scans with high spatio‐temporal resolution while maintaining a reasonable radiation dose.

Methods

Our protocol includes the acquisition of a high‐quality prior image that is reconstructed with a polychromatic iterative algorithm (IMPACT). The acquisition of the post‐contrast sequence is continuous but sparse and these images are reconstructed using prior image constrained compressed sensing (PICCS). A four‐step optimization process is performed using images of a physical phantom. First, the optimal tube current is selected by taking the noise level into account. Second, the optimal number of angles is selected based on the absence of streak artifacts. Third, the number of iterations in IMPACT is specified at the lowest value that achieves the highest spatial resolution. Finally, the number of iterations in PICCS is determined based on the quantitative accuracy of a range of iodine concentrations.

Results

When a high‐quality prior image is available, the imaging of post‐contrast images can be performed using just 40 projection angles with a tube current of 32 mA. The noise level in the post‐contrast images is inherited from the prior image and no streak artifacts are visible. Mean difference between the linear attenuation coefficients of samples containing iodine reconstructed with IMPACT using all 360 projections and PICCS using 40 projections is 0.0004 at most. The spatial resolution of images reconstructed with PICCS is lower than the one of IMPACT images and is concentration dependent. The cut‐off frequency at 10% modulation transfer function drops from 1.55 in the prior image to 0.9 when the target with the largest concentration is evaluated. The total mean glandular dose of the protocol does not exceed 22.5 mGy.

Conclusions

This study found the optimal acquisition and reconstruction parameters for a low‐dose dynamic contrast‐enhanced bCT protocol. The numerical accuracy of the proposed protocol was ensured by performing a physical phantom study.

Subjects

MeSH terms (MEDLINE)

Phantoms, ImagingIodineContrast MediaBreastFour-Dimensional Computed TomographyImage Processing, Computer-AssistedHumansAlgorithms

Author keywords

4D dynamic contrast‐enhanced breast CTPICCSiodine quantificationphantom studyradiation dose optimization

Substances

IodineContrast Media

Authors and institutions

  • Department of Medical Imaging, Radboudumc, Nijmegen, The Netherlands.
  • Technical Medicine Center, University of Twente, Enschede, The Netherlands.
  • Dutch Expert Centre for Screening (LRCB), Nijmegen, The Netherlands.

Funding

  • European Union's Horizon 2020 research and innovation programme
  • European Research Council

Cite

APA

Mikerov, M., Pautasso, J. J., Goris, L., Michielsen, K., & Sechopoulos, I. (2025). 4D Dynamic contrast‐enhanced breast CT: Phantom‐based reconstruction parameter optimization for iodine quantification. Medical Physics. https://doi.org/10.1002/mp.17658

Vancouver

Mikerov M, Pautasso JJ, Goris L, Michielsen K, Sechopoulos I. 4D Dynamic contrast‐enhanced breast CT: Phantom‐based reconstruction parameter optimization for iodine quantification. Medical Physics. 2025. doi:10.1002/mp.17658

BibTeX
@article{Mikerov2025,
  author = {Mikerov, Mikhail and Pautasso, Juan J and Goris, Liselot and Michielsen, Koen and Sechopoulos, Ioannis},
  title = {4D Dynamic contrast‐enhanced breast CT: Phantom‐based reconstruction parameter optimization for iodine quantification},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.17658}
}
Start for free →