Development of an online neutron beam monitoring system for accelerator‐based boron neutron capture therapy in a hospital

Masashi Takada, Natsumi Yagi, Satoshi Nakamura, Kenzi Shimada, Jyun Itami, Hiroshi Igaki, Masaru Nakamura, Tomoya Nunomiya, Satoru Endo, Tsuyoshi Kajimoto, Kenichi Tanaka, Kei Aoyama und 2 weitere

Medical Physics · 2025-01 · 52 (1) · 605-618

DOI: 10.1002/mp.17480 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, online erschienen am 22.10.2024. Verfasst von Masashi Takada und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Applied Physics, National Defense Academy of Japan, Yokosuka, Kanagawa, Japan. Hauptthemen laut MEDLINE: Boron Neutron Capture Therapy, Neutrons. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 30 Literaturangaben. Gefördert von: National Defense Academy of Japan, Japan Society for the Promotion of Science.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Boron neutron capture therapy (BNCT) is a next‐generation radiotherapy, utilizing both an external neutron beam and a ‐containing pharmaceutical. A compact accelerator for a high intensity neutron source was installed to conduct BNCT in a hospital. The dose administered to a patient was evaluated by measuring the proton beam current.

Purpose

Neutron intensity should be monitored in real‐time by measuring the neutrons emitted from the target during BNCT irradiation. This is crucial due to potential neutron target degradation. Online neutron beam monitoring systems are required for reliable measurements of the administered neutron dose. An online neutron beam monitoring system was developed to monitor neutron intensity irradiating on the patient at the National Cancer Center Hospital (NCCH).

Methods

The neutron detector comprised a back‐illuminated thin Si diode of 40‐ thickness and an ultrathin natural LiF neutron converter of 0.05‐ thickness. The neutron detector was installed on the neutron target unit, regardless of whether a patient was present, without any additional modifications to the setup. The response functions for high photon dose rates of upto 100 Gy/h were measured. The pulse heights were measured using the neutron beam monitor during BNCT neutron irradiation. Neutron temporal response measured using the online beam monitor was acquired and compared with the proton beam current and the measurements at a patient position. From this measurement at the patient position, the neutron fluence rate irradiating on a patient was obtained.

Results

The neutron events were separated from the photon events. The neutron counting rates increased rapidly with the starting of proton beam irradiation and dropped to zero upon its termination. During intermittent drops and recoveries in the proton beam, the neutron beam monitor for counting rates responded quickly, synchronizing with the beam current. A scatter plot of the neutron counting rate and proton beam current indicated a good linear correlation. A direct relationship between the online neutron beam monitor's neutron counting rates and those of the patient neutron detector showed a good correlation coefficient of 0.84. A ratio of the both neutron counting rates showed a standard deviation of 6%. The correlation coefficient and standard deviation were improved to 0.94 and 1.5%, by re‐binning the neutron temporal response with longer acquisition period than 1 s. Using the online neutron beam monitor, the neutron fluence rate was obtained from the direct relationship within 1.5%. Therefore, real‐time monitoring of neutron intensity was achieved within the acceptable level as per the International Commission on Radiation Units and Measurements report.

Conclusions

The online neutron beam monitoring system was developed to monitor the BNCT neutron beam intensity at NCCH. The temporal response of the neutron beam monitor was synchronized with the neutron counting rate at the patient position. Using the online neutron beam monitor, the neutron fluence rate irradiating on the patient can be monitored from the direct relationship. Fluctuation of the neutron beam intensity through BNCT irradiation and the degradation of the lithium target through the lifespan of the neutron target could be monitored using the neutron beam monitor.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Boron Neutron Capture TherapyNeutronsParticle AcceleratorsHumansHospitalsRadiotherapy DosageOnline Systems

Stichwörter der Autor:innen

BNCTSi diodeneutrononline monitor

Autor:innen und Institutionen

  • Department of Applied Physics, National Defense Academy of Japan, Yokosuka, Kanagawa, Japan.
  • National Cancer Center Hospital, Chuo-ku, Tokyo, Japan.
  • Cancer Intelligence Care Systems, Inc., Koto-ku, Tokyo, Japan.
  • Shin-Matsudo Accuracy Radiation Therapy Center, Shin-Matsudo Central General Hospital, Matsudo, Chiba, Japan.
  • Fuji Electric Co., Ltd., Hino, Tokyo, Japan.
  • Quantum Energy Applications, Graduate School of Advanced Science and Engineering, Hiroshima University, Higashi-Hiroshima, Hiroshima, Japan.

Förderung

  • National Defense Academy of Japan
  • Japan Society for the Promotion of Science

Zitieren

APA

Takada, M., Yagi, N., Nakamura, S., Shimada, K., Itami, J., Igaki, H., Nakamura, M., Nunomiya, T., Endo, S., Kajimoto, T., Tanaka, K., Aoyama, K., Narita, M., & Nakamura, T. (2025). Development of an online neutron beam monitoring system for accelerator‐based boron neutron capture therapy in a hospital. Medical Physics, 52(1), 605-618. https://doi.org/10.1002/mp.17480

Vancouver

Takada M, Yagi N, Nakamura S, Shimada K, Itami J, Igaki H, et al. Development of an online neutron beam monitoring system for accelerator‐based boron neutron capture therapy in a hospital. Medical Physics. 2025;52(1):605-618. doi:10.1002/mp.17480

BibTeX
@article{Takada2025,
  author = {Takada, Masashi and Yagi, Natsumi and Nakamura, Satoshi and Shimada, Kenzi and Itami, Jyun and Igaki, Hiroshi and Nakamura, Masaru and Nunomiya, Tomoya and Endo, Satoru and Kajimoto, Tsuyoshi and Tanaka, Kenichi and Aoyama, Kei and Narita, Masakuni and Nakamura, Takashi},
  title = {Development of an online neutron beam monitoring system for accelerator‐based boron neutron capture therapy in a hospital},
  journal = {Medical Physics},
  year = {2025},
  volume = {52},
  number = {1},
  pages = {605--618},
  doi = {10.1002/mp.17480}
}
Kostenlos starten →