Evaluation of the impact of cardiac implantable electronic devices on cine MRI for real‐time adaptive cardiac radioablation on a 1.5 T MR‐linac

Osman Akdag, Stefano Mandija, Pim T S Borman, Paris Tzitzimpasis, Astrid L H M W van Lier, Rick Keesman, Bas W Raaymakers, Martin F Fast

Medical Physics · 2025

DOI: 10.1002/mp.17438 ↗

At a glance

Evaluationsstudie in Medical Physics, 2025. Written by Osman Akdag and 7 co-authors. Institution: Department of Radiotherapy, University Medical Center Utrecht, Utrecht, The Netherlands. Major MEDLINE topics: Magnetic Resonance Imaging, Cine. 40 references. The full text is freely available (open access, CC BY 4.0). Funded by: Nederlandse Organisatie voor Wetenschappelijk Onderzoek.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Stereotactic arrhythmia radioablation (STAR) is a novel treatment approach for refractory ventricular tachycardia (VT). The risk of treatment‐induced toxicity and geographic miss can be reduced with online MRI‐guidance on an MR‐linac. However, most VT patients carry cardiac implantable electronic devices (CIED), which compromise MR images.

Purpose

Robust MR‐linac imaging sequences are required for cardiac visualization and accurate motion monitoring in presence of a CIED during MRI‐guided STAR. We optimized two clinically available cine sequences for cardiorespiratory motion estimation in presence of a CIED on a 1.5 T MR‐linac. The image quality, motion estimation accuracy, and geometric fidelity using these cine sequences were evaluated.

Methods

Clinically available 2D balanced steady‐state free precession (bSSFP, voxel size = 3.0 3.0 10 mm 3 , T scan = 96 ms, bandwidth (BW) = 1884 Hz/px) and ‐spoiled gradient echo (‐GRE, voxel size = 4.0 4.0 10 mm 3 , T scan = 97 ms, BW = 500 Hz/px) sequences were adjusted for real‐time cardiac visualization and cardiorespiratory motion estimation on a 1.5 T Unity MR‐linac (Elekta AB, Stockholm, Sweden), while complying with safety guidelines for MRI in presence of CIEDs (specific absorption rate 2 W/kg and 80 mT/s). Cine acquisitions were performed in five healthy volunteers, with and without an implantable cardioverter– defibrillator (ICD) placed on the clavicle, and a VT patient. Generalized divergence‐curl (GDC) deformable image registration (DIR) was used for automated landmark motion estimation in the left ventricle (LV). Gaussian processes (GP), a machine‐learning technique, was trained using GDC landmarks and deployed for real‐time cardiorespiratory motion prediction. ‐mapping was performed to assess geometric image fidelity in the presence of CIEDs.

Results

CIEDs introduced banding artifacts partially obscuring cardiac structures in bSSFP acquisitions. In contrast, the ‐GRE was more robust to CIED‐induced artifacts at the expense of a lower signal‐to‐noise ratio. In presence of an ICD, image‐based cardiorespiratory motion estimation was possible for 85% (100%) of the volunteers using the bSSFP (‐GRE) sequence. The in‐plane 2D root‐mean‐squared deviation (RMSD) range between GDC‐derived landmarks and manual annotations using the bSSFP (T 1 ‐GRE) sequence was 3.1–3.3 (3.3–4.1) mm without ICD and 4.6–4.6 (3.2–3.3) mm with ICD. Without ICD, the RMSD between the GP‐predictions and GDC‐derived landmarks ranged between 0.9 and 2.2 mm (1.3–3.0 mm) for the bSSFP (T 1 ‐GRE) sequence. With ICD, the RMSD between the GP‐predictions and GDC‐derived landmarks ranged between 1.3 and 2.2 mm (1.2–3.2 mm) using the bSSFP (T 1 ‐GRE) sequence resulting in an RMSD‐increase of 42%–143% (bSSFP) and −61%–142% (T 1 ‐GRE). Lead‐induced spatial distortions ranged between −0.2 and 0.2 mm (−0.7–1.2 mm) using the bSSFP (‐GRE) sequence. The 98 th percentile range of the spatial distortions in the gross target volume of the patient was between 0.0 and 0.4 mm (0.0–1.8 mm) when using bSSFP (‐GRE).

Conclusions

Tailored bSSFP and ‐GRE sequences can facilitate real‐time cardiorespiratory estimation using GP trained with GDC‐derived landmarks in the majority of landmark locations in the LV despite the presence of CIEDs. The need for high temporal resolution noticeably reduced achievable spatial resolution of the cine MRIs. However, the effect of the CIED‐induced artifacts is device, patient and sequence dependent and requires specific assessment per case.

Subjects

MeSH terms (MEDLINE)

Magnetic Resonance Imaging, CineHumansHeartRadiosurgeryDefibrillators, ImplantableImage Processing, Computer-AssistedTime FactorsTachycardia, VentricularMovement

Author keywords

MRI‐guided radiotherapycardiac radiotherapyreal‐time

Authors and institutions

  • Department of Radiotherapy, University Medical Center Utrecht, Utrecht, The Netherlands.
  • Computational Imaging Group for MR Diagnostics and Therapy, Center for Image Sciences, University Medical Center Utrecht, Utrecht, The Netherlands.
  • Department of Radiotherapy, Radboud University Medical Center, Nijmegen, The Netherlands.

Funding

  • Nederlandse Organisatie voor Wetenschappelijk Onderzoek

Cite

APA

Akdag, O., Mandija, S., Borman, P. T. S., Tzitzimpasis, P., van Lier, A. L. H. M. W., Keesman, R., Raaymakers, B. W., & Fast, M. F. (2025). Evaluation of the impact of cardiac implantable electronic devices on cine MRI for real‐time adaptive cardiac radioablation on a 1.5 T MR‐linac. Medical Physics. https://doi.org/10.1002/mp.17438

Vancouver

Akdag O, Mandija S, Borman PTS, Tzitzimpasis P, van Lier ALHMW, Keesman R, et al. Evaluation of the impact of cardiac implantable electronic devices on cine MRI for real‐time adaptive cardiac radioablation on a 1.5 T MR‐linac. Medical Physics. 2025. doi:10.1002/mp.17438

BibTeX
@article{Akdag2025,
  author = {Akdag, Osman and Mandija, Stefano and Borman, Pim T S and Tzitzimpasis, Paris and van Lier, Astrid L H M W and Keesman, Rick and Raaymakers, Bas W and Fast, Martin F},
  title = {Evaluation of the impact of cardiac implantable electronic devices on cine MRI for real‐time adaptive cardiac radioablation on a 1.5 T MR‐linac},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.17438}
}
Start for free →