AAPM Task Group Report 299: Quality control in multi‐energy computed tomography
Medical Physics · 2024-10 · 51 (10) · 7012-7037
DOI: 10.1002/mp.17322 ↗At a glance
Journal article in Medical Physics, published online on 2024-07-29. Written by Rick R Layman and 19 co-authors. Institution: Department of Imaging Physics, University of Texas MD Anderson Cancer Center, Houston, Texas, USA. Major MEDLINE topics: Quality Control, Tomography, X-Ray Computed. Cited 21 times so far (Crossref); 53 references.
Abstract (original language)
Multi‐energy computed tomography (MECT) offers the opportunity for advanced visualization, detection, and quantification of select elements (e.g., iodine) or materials (e.g., fat) beyond the capability of standard single‐energy computed tomography (CT). However, the use of MECT requires careful consideration as substantially different hardware and software approaches have been used by manufacturers, including different sets of user‐selected or hidden parameters that affect the performance and radiation dose of MECT. Another important consideration when designing MECT protocols is appreciation of the specific tasks being performed; for instance, differentiating between two different materials or quantifying a specific element. For a given task, it is imperative to consider both the radiation dose and task‐specific image quality requirements. Development of a quality control (QC) program is essential to ensure the accuracy and reproducibility of these MECT applications. Although standard QC procedures have been well established for conventional single‐energy CT, the substantial differences between single‐energy CT and MECT in terms of system implementations, imaging protocols, and clinical tasks warrant QC tests specific to MECT. This task group was therefore charged with developing a systematic QC program designed to meet the needs of MECT applications. In this report, we review the various MECT approaches that are commercially available, including information about hardware implementation, MECT image types, image reconstruction, and postprocessing techniques that are unique to MECT. We address the requirements for MECT phantoms, review representative commercial MECT phantoms, and offer guidance regarding homemade MECT phantoms. We discuss the development of MECT protocols, which must be designed carefully with proper consideration of MECT technology, imaging task, and radiation dose. We then outline specific recommended QC tests in terms of general image quality, radiation dose, differentiation and quantification tasks, and diagnostic and therapeutic applications.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Imaging Physics, University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
- Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
- Canon Medical Systems, Otawara, Japan.
- U.S. Food and Drug Administration, Silver Spring, Maryland, USA.
- Philips, Cleveland, Ohio, USA.
- Department of Radiology, University of Texas Southwestern Medical Center, Dallas, Texas, USA.
Cite
APA
Layman, R. R., Leng, S., Boedeker, K. L., Burk, L. M., Dang, H., Duan, X., Jacobsen, M. C., Li, B., Li, K., Little, K., Madhav, P., Miller, J., Nute, J. L., Giraldo, J. C. R., Ruchala, K. J., Tao, S., Varchena, V., Vedantham, S., Zeng, R., & Zhang, D. (2024). AAPM Task Group Report 299: Quality control in multi‐energy computed tomography. Medical Physics, 51(10), 7012-7037. https://doi.org/10.1002/mp.17322
Vancouver
Layman RR, Leng S, Boedeker KL, Burk LM, Dang H, Duan X, et al. AAPM Task Group Report 299: Quality control in multi‐energy computed tomography. Medical Physics. 2024;51(10):7012-7037. doi:10.1002/mp.17322
BibTeX
@article{Layman2024,
author = {Layman, Rick R and Leng, Shuai and Boedeker, Kirsten L and Burk, Laurel M and Dang, Hao and Duan, Xinhui and Jacobsen, Megan C and Li, Baojun and Li, Ke and Little, Kevin and Madhav, Priti and Miller, Jessica and Nute, Jessica L and Giraldo, Juan Carlos Ramirez and Ruchala, Kenneth J and Tao, Shengzhen and Varchena, Vladimir and Vedantham, Srinivasan and Zeng, Rongping and Zhang, Da},
title = {AAPM Task Group Report 299: Quality control in multi‐energy computed tomography},
journal = {Medical Physics},
year = {2024},
volume = {51},
number = {10},
pages = {7012--7037},
doi = {10.1002/mp.17322}
}
EUVIMED