Automatic vessel attenuation measurement for quality control of contrast‐enhanced CT: Validation on the portal vein

Kevin McCoy, Sujay Marisetty, Dominique Tan, Corey T Jensen, Jeffrey H Siewerdsen, Christine B Peterson, Moiz Ahmad

Medical Physics · 2024

DOI: 10.1002/mp.17267 ↗

At a glance

Validierungsstudie in Medical Physics, 2024. Written by Kevin McCoy and 6 co-authors. Institution: Department of Statistics, Rice University, Houston, Texas, USA. Major MEDLINE topics: Portal Vein, Tomography, X-Ray Computed, Quality Control, Contrast Media. 37 references. The full text is freely available in PubMed Central. Funded by: NCI NIH HHS, National Science Foundation, Cancer Prevention and Research Institute of Texas.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Adequate image enhancement of organs and blood vessels of interest is an important aspect of image quality in contrast‐enhanced computed tomography (CT). There is a need for an objective method for evaluation of vessel contrast that can be automatically and systematically applied to large sets of CT exams.

Purpose

The purpose of this work was to develop a method to automatically segment and measure attenuation Hounsfield Unit (HU) in the portal vein (PV) in contrast‐enhanced abdomen CT examinations.

Methods

Input CT images were processed by a vessel enhancing filter to determine candidate PV segmentations. Multiple machine learning (ML) classifiers were evaluated for classifying a segmentation as corresponding to the PV based on segmentation shape, location, and intensity features. A public data set of 82 contrast‐enhanced abdomen CT examinations was used to train the method. An optimal ML classifier was selected by training and tuning on 66 out of the 82 exams (80% training split) in the public data set. The method was evaluated in terms of segmentation classification accuracy and PV attenuation measurement accuracy, compared to manually determined ground truth, on a test set of the remaining 16 exams (20% test split) held out from public data set. The method was further evaluated on a separate, independently collected test set of 21 examinations.

Results

The best classifier was found to be a random forest, with a precision of 0.892 in the held‐out test set to correctly identify the PV from among the input candidate segmentations. The mean absolute error of the measured PV attenuation relative to ground truth manual measurement was 13.4 HU. On the independent test set, the overall precision decreased to 0.684. However, the PV attenuation measurement remained relatively accurate with a mean absolute error of 15.2 HU.

Conclusions

The method was shown to accurately measure PV attenuation over a large range of attenuation values, and was validated in an independently collected dataset. The method did not require time‐consuming manual contouring to supervise training. The method may be applied to systematic quality control of contrast‐enhanced CT examinations.

Subjects

MeSH terms (MEDLINE)

Portal VeinTomography, X-Ray ComputedQuality ControlContrast MediaImage Processing, Computer-AssistedHumansMachine LearningAutomation

Author keywords

computed tomographycontrast enhancementimage qualitymachine learningquality controlrandom forest

Substances

Contrast Media

Authors and institutions

  • Department of Statistics, Rice University, Houston, Texas, USA.
  • Department of Biostatistics, Division of Basic Science Research, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • Rice University, Houston, Texas, USA.
  • The University of Texas, Austin, Texas, USA.
  • Department of Abdominal Imaging, Division of Diagnostic Imaging, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • Department of Imaging Physics, Division of Diagnostic Imaging, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.

Funding

  • NCI NIH HHS
  • National Science Foundation
  • Cancer Prevention and Research Institute of Texas
  • National Cancer Institute

Cite

APA

McCoy, K., Marisetty, S., Tan, D., Jensen, C. T., Siewerdsen, J. H., Peterson, C. B., & Ahmad, M. (2024). Automatic vessel attenuation measurement for quality control of contrast‐enhanced CT: Validation on the portal vein. Medical Physics. https://doi.org/10.1002/mp.17267

Vancouver

McCoy K, Marisetty S, Tan D, Jensen CT, Siewerdsen JH, Peterson CB, et al. Automatic vessel attenuation measurement for quality control of contrast‐enhanced CT: Validation on the portal vein. Medical Physics. 2024. doi:10.1002/mp.17267

BibTeX
@article{McCoy2024,
  author = {McCoy, Kevin and Marisetty, Sujay and Tan, Dominique and Jensen, Corey T and Siewerdsen, Jeffrey H and Peterson, Christine B and Ahmad, Moiz},
  title = {Automatic vessel attenuation measurement for quality control of contrast‐enhanced CT: Validation on the portal vein},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17267}
}
Start for free →