HDR brachytherapy afterloader quality assurance optimization using monolithic silicon strip detectors
Medical Physics · 2024
DOI: 10.1002/mp.17240 ↗At a glance
Journal article in Medical Physics, 2024. Written by Broady Hunt and 6 co-authors. Institution: Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, Australia. Major MEDLINE topics: Brachytherapy, Silicon. Cited 4 times so far (Crossref); 26 references. The full text is freely available (open access, CC BY-NC-ND 4.0).
Abstract (original language)
Background
There currently exists no widespread high dose‐rate (HDR) brachytherapy afterloader quality assurance (QA) tool for simultaneously assessing the afterloader's positional, temporal, transit velocity and air kerma strength accuracy.
Purpose
The purpose of this study was to develop a precise and rigorous technique for performing daily QA of HDR brachytherapy afterloaders, incorporating QA of: dwell position accuracy, dwell time accuracy, transit velocity consistency and relative air kerma strength (AKS) of an Ir‐192 source.
Method
A Sharp ProGuide 240 mm catheter (Elekta Brachytherapy, Veenendaal, The Netherlands) was fixed 5 mm above a 256 channel epitaxial diode array ‘dose magnifying glass’ (DMG256) (Centre for Medical and Radiation Physics, University of Wollongong). Three dwell positions, each of 5.0 s dwell times, were spaced 13.0 mm apart along the array with the Flexitron HDR afterloader (Elekta Brachytherapy, Veenendaal, The Netherlands). The DMG256 was connected to a data acquisition system (DAQ) and a computer via USB2.0 link for live readout and post‐processing. The outputted data files were analyzed using a Python script to provide positional and temporal localization of the Ir‐192 source by tracking the centroid of the detected response. Measurements were repeated on a weekly basis, for a period of 5 weeks to determine the consistency of the measured parameters over an extended period.
Results
Using the DMG256 for relative AKS measurements resulted in measured values within 0.6%–3.0% of the expected activity over a 7‐week period. The sub‐millisecond temporal accuracy of the device allowed for measurements of the transit velocity with an average of (10.88 ± 1.01) cm/s for 13 mm steps. The dwell position localization for 1, 2, 3, 5, and 10 mm steps had an accuracy between 0.1 and 0.3 mm (3σ), with a fixed temporal accuracy of 10 ms.
Conclusion
The DMG256 silicon strip detector allows for clinics to perform rigorous daily QA of HDR afterloader dwell position and dwell time accuracy with greater precision than the current standard methodology using closed circuit television and a stopwatch. Additionally, DMG256 unlocks the ability to perform measurements of transit velocity/time and relative AKS, which are not possible using current standard techniques.
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, Australia.
- Department of Radiation Oncology, St George Cancer Care Centre, Kogarah, NSW, Australia.
- School of Physics, University of Sydney, Camperdown, NSW, Australia.
Cite
APA
Hunt, B., Cutajar, D., Petasecca, M., Rosenfeld, A., Howie, A., Bucci, J., & Poder, J. (2024). HDR brachytherapy afterloader quality assurance optimization using monolithic silicon strip detectors. Medical Physics. https://doi.org/10.1002/mp.17240
Vancouver
Hunt B, Cutajar D, Petasecca M, Rosenfeld A, Howie A, Bucci J, et al. HDR brachytherapy afterloader quality assurance optimization using monolithic silicon strip detectors. Medical Physics. 2024. doi:10.1002/mp.17240
BibTeX
@article{Hunt2024,
author = {Hunt, Broady and Cutajar, Dean and Petasecca, Marco and Rosenfeld, Anatoly and Howie, Andrew and Bucci, Joseph and Poder, Joel},
title = {HDR brachytherapy afterloader quality assurance optimization using monolithic silicon strip detectors},
journal = {Medical Physics},
year = {2024},
doi = {10.1002/mp.17240}
}
EUVIMED