EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

HDR brachytherapy afterloader quality assurance optimization using monolithic silicon strip detectors

Broady Hunt, Dean Cutajar, Marco Petasecca, Anatoly Rosenfeld, Andrew Howie, Joseph Bucci, Joel Poder

Medical Physics · 2024

DOI: 10.1002/mp.17240 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Broady Hunt und 6 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, Australia. Hauptthemen laut MEDLINE: Brachytherapy, Silicon. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

There currently exists no widespread high dose‐rate (HDR) brachytherapy afterloader quality assurance (QA) tool for simultaneously assessing the afterloader's positional, temporal, transit velocity and air kerma strength accuracy.

Purpose

The purpose of this study was to develop a precise and rigorous technique for performing daily QA of HDR brachytherapy afterloaders, incorporating QA of: dwell position accuracy, dwell time accuracy, transit velocity consistency and relative air kerma strength (AKS) of an Ir‐192 source.

Method

A Sharp ProGuide 240 mm catheter (Elekta Brachytherapy, Veenendaal, The Netherlands) was fixed 5 mm above a 256 channel epitaxial diode array ‘dose magnifying glass’ (DMG256) (Centre for Medical and Radiation Physics, University of Wollongong). Three dwell positions, each of 5.0 s dwell times, were spaced 13.0 mm apart along the array with the Flexitron HDR afterloader (Elekta Brachytherapy, Veenendaal, The Netherlands). The DMG256 was connected to a data acquisition system (DAQ) and a computer via USB2.0 link for live readout and post‐processing. The outputted data files were analyzed using a Python script to provide positional and temporal localization of the Ir‐192 source by tracking the centroid of the detected response. Measurements were repeated on a weekly basis, for a period of 5 weeks to determine the consistency of the measured parameters over an extended period.

Results

Using the DMG256 for relative AKS measurements resulted in measured values within 0.6%–3.0% of the expected activity over a 7‐week period. The sub‐millisecond temporal accuracy of the device allowed for measurements of the transit velocity with an average of (10.88 ± 1.01) cm/s for 13 mm steps. The dwell position localization for 1, 2, 3, 5, and 10 mm steps had an accuracy between 0.1 and 0.3 mm (3σ), with a fixed temporal accuracy of 10 ms.

Conclusion

The DMG256 silicon strip detector allows for clinics to perform rigorous daily QA of HDR afterloader dwell position and dwell time accuracy with greater precision than the current standard methodology using closed circuit television and a stopwatch. Additionally, DMG256 unlocks the ability to perform measurements of transit velocity/time and relative AKS, which are not possible using current standard techniques.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BrachytherapySiliconQuality Assurance, Health CareRadiometryRadiotherapy DosageQuality Control

Stichwörter der Autor:innen

brachytherapydose magnifying glassquality assuranceremote afterloadersilicon strip detector

Substanzen

Silicon

Autor:innen und Institutionen

  • Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, Australia.
  • Department of Radiation Oncology, St George Cancer Care Centre, Kogarah, NSW, Australia.
  • School of Physics, University of Sydney, Camperdown, NSW, Australia.

Zitieren

APA

Hunt, B., Cutajar, D., Petasecca, M., Rosenfeld, A., Howie, A., Bucci, J., & Poder, J. (2024). HDR brachytherapy afterloader quality assurance optimization using monolithic silicon strip detectors. Medical Physics. https://doi.org/10.1002/mp.17240

Vancouver

Hunt B, Cutajar D, Petasecca M, Rosenfeld A, Howie A, Bucci J, et al. HDR brachytherapy afterloader quality assurance optimization using monolithic silicon strip detectors. Medical Physics. 2024. doi:10.1002/mp.17240

BibTeX
@article{Hunt2024,
  author = {Hunt, Broady and Cutajar, Dean and Petasecca, Marco and Rosenfeld, Anatoly and Howie, Andrew and Bucci, Joseph and Poder, Joel},
  title = {HDR brachytherapy afterloader quality assurance optimization using monolithic silicon strip detectors},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17240}
}
Kostenlos starten →