High spatiotemporal resolution scintillation imaging of pulsed pencil beam scanning proton beams produced by a gantry‐mounted synchrocyclotron

S Murty Goddu, Yao Hao, Zhen Ji, Jufri Setianegara, Fengwei Liu, Winter Green, Lee G Sobotka, Tianyu Zhao, Stephanie Perkins, Arash Darafsheh iD

Medical Physics · 2024

DOI: 10.1002/mp.17116 ↗

At a glance

Journal article in Medical Physics, 2024. Written by S Murty Goddu and 9 co-authors. Institution: Department of Radiation Oncology, Washington University School of Medicine in St. Louis, St. Louis, Missouri, USA. Major MEDLINE topics: Scintillation Counting. Cited 5 times so far (Crossref); 49 references. The full text is freely available in PubMed Central. Funded by: Alvin J. Siteman Cancer Center, Department of Radiation Oncology, NCI NIH HHS.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

A dosimeter with high spatial and temporal resolution would be of significant interest for pencil beam scanning (PBS) proton beams’ characterization, especially when facing small fields and beams with high temporal dynamics. Optical imaging of scintillators has potential in providing sub‐millimeter spatial resolution with pulse‐by‐pulse basis temporal resolution when the imaging system is capable of operating in synchrony with the beam‐producing accelerator.

Purpose

We demonstrate the feasibility of imaging PBS proton beams as they pass through a plastic scintillator detector to simultaneously obtain multiple beam parameters, including proton range, pencil beam's widths at different depths, spot's size, and spot's position on a pulse‐by‐pulse basis with sub‐millimeter resolution. Materials and methods A PBS synchrocyclotron was used for proton irradiation. A BC‐408 plastic scintillator block with 30 × 30 × 5 cm 3 size, and another block with 30 × 30 × 0.5 cm 3 size, positioned in an optically sealed housing, were used sequentially to measure the proton range, and spot size/location, respectively. A high‐speed complementary metal‐oxide‐semiconductor (CMOS) camera system synchronized with the accelerator's pulses through a gating module was used for imaging. Scintillation images, captured with the camera directly facing the 5‐cm‐thick scintillator, were corrected for background (BG), and ionization quenching of the scintillator to obtain the proton range. Spots’ position and size were obtained from scintillation images of the 0.5‐cm‐thick scintillator when a 45° mirror was used to reflect the scintillation light toward the camera.

Results

Scintillation images with 0.16 mm/pixel resolution corresponding to all proton pulses were captured. Pulse‐by‐pulse analysis showed that variations of the range, spots’ position, and size were within ± 0.2% standard deviation of their average values. The absolute ranges were within ± 1 mm of their expected values. The average spot‐positions were mostly within ± 0.8 mm and spots’ sigma agreed within 0.2 mm of the expected values.

Conclusion

Scintillation‐imaging PBS beams with high‐spatiotemporal resolution is feasible and may help in efficient and cost‐effective acceptance testing and commissioning of existing and even emerging technologies such as FLASH, grid, mini‐beams, and so forth.

Subjects

MeSH terms (MEDLINE)

Scintillation CountingProtonsCyclotronsFeasibility StudiesProton Therapy

Author keywords

imagingproton therapyquality assurancescintillationsynchrocyclotron

Substances

Protons

Authors and institutions

  • Department of Radiation Oncology, Washington University School of Medicine in St. Louis, St. Louis, Missouri, USA.
  • Department of Chemistry, Washington University in St. Louis, St. Louis, Missouri, USA.

ORCID profiles: Arash Darafsheh

Funding

  • Alvin J. Siteman Cancer Center
  • Department of Radiation Oncology
  • NCI NIH HHS
  • Washington University School of Medicine in St. Louis
  • S. Lee Kling Proton Therapy Center
  • Foundation for Barnes-Jewish Hospital
  • National Cancer Institute

Cite

APA

Goddu, S. M., Hao, Y., Ji, Z., Setianegara, J., Liu, F., Green, W., Sobotka, L. G., Zhao, T., Perkins, S., & Darafsheh, A. (2024). High spatiotemporal resolution scintillation imaging of pulsed pencil beam scanning proton beams produced by a gantry‐mounted synchrocyclotron. Medical Physics. https://doi.org/10.1002/mp.17116

Vancouver

Goddu SM, Hao Y, Ji Z, Setianegara J, Liu F, Green W, et al. High spatiotemporal resolution scintillation imaging of pulsed pencil beam scanning proton beams produced by a gantry‐mounted synchrocyclotron. Medical Physics. 2024. doi:10.1002/mp.17116

BibTeX
@article{Goddu2024,
  author = {Goddu, S Murty and Hao, Yao and Ji, Zhen and Setianegara, Jufri and Liu, Fengwei and Green, Winter and Sobotka, Lee G and Zhao, Tianyu and Perkins, Stephanie and Darafsheh, Arash},
  title = {High spatiotemporal resolution scintillation imaging of pulsed pencil beam scanning proton beams produced by a gantry‐mounted synchrocyclotron},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17116}
}
Start for free →