Initial validation of a diaphragm tracking system for multiple breath‐hold volumetric modulated arc therapy of abdominal tumors: A phantom study
Medical Physics · 2024
DOI: 10.1002/mp.17008 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Yuki Nozawa und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, The University of Tokyo Hospital, Bunkyo-ku, Tokyo, Japan. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated, Lung Neoplasms, Abdominal Neoplasms. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The breath‐hold radiotherapy has been increasingly used to mitigate interfractional and intrafractional breathing impact on treatment planning and beam delivery. Previous techniques include body surface measurements or radiopaque metal markers, each having known disadvantages.
Purpose
We recently proposed a new markerless technique without the disadvantages, where diaphragm was registered between DRR and fluoroscopic x‐ray projection images every 180 ms during VMAT delivery. An initial validation of the proposed diaphragm tracking system (DiaTrak) was performed using a chest phantom to evaluate its characteristics.
Methods
Diaphragm registration was performed between DRR and projection streaming kV x‐ray images of a chest phantom during VMAT delivery. Streaming data including the projection images and the beam angles were transferred from a linac system to an external PC, where the diaphragm registration accuracy and beam‐off latency were measured based on image cross correlation between the DRR and the projection images every 180 ms.
Results
It was shown that the average of the beam‐off latency was 249.5 ms and the average of the diaphragm registration error was 0.84 mm Conclusions Initial validation of the proposed DiaTrak system for multiple breath‐hold VMAT of abdominal tumors has been successfully completed with a chest phantom. The resulting beam‐off latency and the diaphragm registration error were regarded clinically acceptable.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, The University of Tokyo Hospital, Bunkyo-ku, Tokyo, Japan.
- Department of Comprehensive Radiation Oncology, The University of Tokyo, Bunkyo-ku, Tokyo, Japan.
Zitieren
APA
Nozawa, Y., Ohta, T., Katano, A., Nawa, K., Ohira, S., Yamashita, H., & Nakagawa, K. (2024). Initial validation of a diaphragm tracking system for multiple breath‐hold volumetric modulated arc therapy of abdominal tumors: A phantom study. Medical Physics. https://doi.org/10.1002/mp.17008
Vancouver
Nozawa Y, Ohta T, Katano A, Nawa K, Ohira S, Yamashita H, et al. Initial validation of a diaphragm tracking system for multiple breath‐hold volumetric modulated arc therapy of abdominal tumors: A phantom study. Medical Physics. 2024. doi:10.1002/mp.17008
BibTeX
@article{Nozawa2024,
author = {Nozawa, Yuki and Ohta, Takeshi and Katano, Atsuto and Nawa, Kanabu and Ohira, Shingo and Yamashita, Hideomi and Nakagawa, Keiichi},
title = {Initial validation of a diaphragm tracking system for multiple breath‐hold volumetric modulated arc therapy of abdominal tumors: A phantom study},
journal = {Medical Physics},
year = {2024},
doi = {10.1002/mp.17008}
}
EUVIMED