Reducing windmill artifacts in clinical spiral CT using a deep learning‐based projection raw data upsampling: Method and robustness evaluation
Medical Physics · 2024
DOI: 10.1002/mp.16938 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Jan Magonov und 6 weiteren Autor:innen. Beteiligte Einrichtung: Division of X-Ray Imaging and Computed Tomography, German Cancer Research Center (DKFZ) Heidelberg, Heidelberg, Germany. Hauptthemen laut MEDLINE: Artifacts, Deep Learning. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Siemens Healthineers AG, Siemens Healthineers.
Abstract (Originalsprache)
Background
Multislice spiral computed tomography (MSCT) requires an interpolation between adjacent detector rows during backprojection. Not satisfying the Nyquist sampling condition along the z‐axis results in aliasing effects, also known as windmill artifacts. These image distortions are characterized by bright streaks diverging from high contrast structures.
Purpose
The z‐flying focal spot (zFFS) is a well‐established hardware‐based solution that aims to double the sampling rate in longitudinal direction and therefore reduce aliasing artifacts. However, given the technical complexity of the zFFS, this work proposes a deep learning‐based approach as an alternative solution.
Methods
We propose a supervised learning approach to perform a mapping between input projections and the corresponding rows required for double sampling in the z‐direction. We present a comprehensive evaluation using both a clinical dataset obtained using raw data from 40 real patient scans acquired with zFFS and a synthetic dataset consisting of 100 simulated spiral scans using a phantom specifically designed for our problem. For the clinical dataset, we utilized 32 scans as training set and 8 scans as validation set, whereas for the synthetic dataset, we used 80 scans for training and 20 scans for validation purposes. Both qualitative and quantitative assessments are conducted on a test set consisting of nine real patient scans and six phantom measurements to validate the performance of our approach. A simulation study was performed to investigate the robustness against different scan configurations in terms of detector collimation and pitch value.
Results
In the quantitative comparison based on clinical patient scans from the test set, all network configurations show an improvement in the root mean square error (RMSE) of approximately 20% compared to neglecting the doubled longitudinal sampling by the zFFS. The results of the qualitative analysis indicate that both clinical and synthetic training data can reduce windmill artifacts through the application of a correspondingly trained network. Together with the qualitative results from the test set phantom measurements it is emphasized that a training of our method with synthetic data resulted in superior performance in windmill artifact reduction.
Conclusions
Deep learning‐based raw data interpolation has the potential to enhance the sampling in z‐direction and thus minimize aliasing effects, as it is the case with the zFFS. Especially a training with synthetic data showed promising results. While it may not outperform zFFS, our method represents a beneficial solution for CT scanners lacking the necessary hardware components for zFFS.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of X-Ray Imaging and Computed Tomography, German Cancer Research Center (DKFZ) Heidelberg, Heidelberg, Germany.
- Computed Tomography Division, Siemens Healthineers AG, Forchheim, Germany.
- Medical Faculty, Heidelberg University, Heidelberg, Germany.
Förderung
- Siemens Healthineers AG
- Siemens Healthineers
Zitieren
APA
Magonov, J., Maier, J., Erath, J., Sunnegårdh, J., Fournié, E., Stierstorfer, K., & Kachelrieß, M. (2024). Reducing windmill artifacts in clinical spiral CT using a deep learning‐based projection raw data upsampling: Method and robustness evaluation. Medical Physics. https://doi.org/10.1002/mp.16938
Vancouver
Magonov J, Maier J, Erath J, Sunnegårdh J, Fournié E, Stierstorfer K, et al. Reducing windmill artifacts in clinical spiral CT using a deep learning‐based projection raw data upsampling: Method and robustness evaluation. Medical Physics. 2024. doi:10.1002/mp.16938
BibTeX
@article{Magonov2024,
author = {Magonov, Jan and Maier, Joscha and Erath, Julien and Sunnegårdh, Johan and Fournié, Eric and Stierstorfer, Karl and Kachelrieß, Marc},
title = {Reducing windmill artifacts in clinical spiral CT using a deep learning‐based projection raw data upsampling: Method and robustness evaluation},
journal = {Medical Physics},
year = {2024},
doi = {10.1002/mp.16938}
}
EUVIMED