EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Neutron flux evaluation algorithm with a combination of Monte Carlo and removal‐diffusion calculation methods for boron neutron capture therapy

Mai Nojiri, Takushi Takata, Naonori Hu, Yoshinori Sakurai, Minoru Suzuki, Hiroki Tanaka

Medical Physics · 2024

DOI: 10.1002/mp.16931 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Mai Nojiri und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nuclear Engineering, Graduate School of Engineering, Kyoto University, Kyoto, Japan. Hauptthemen laut MEDLINE: Boron Neutron Capture Therapy, Monte Carlo Method, Algorithms, Neutrons. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: JSPS KAKENHI, JST SPRING.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

In Japan, the clinical treatment of boron neutron capture therapy (BNCT) has been applied to unresectable, locally advanced, and recurrent head and neck carcinomas using an accelerator‐based neutron source since June of 2020. Considering the increase in the number of patients receiving BNCT, efficiency of the treatment planning procedure is becoming increasingly important. Therefore, novel and rapid dose calculation algorithms must be developed. We developed a novel algorithm for calculating neutron flux, which comprises of a combination of a Monte Carlo (MC) method and a method based on the removal‐diffusion (RD) theory (RD calculation method) for the purpose of dose calculation of BNCT.

Purpose

We present the details of our novel algorithm and the verification results of the calculation accuracy based on the MC calculation result.

Methods

In this study, the “MC‐RD” calculation method was developed, wherein the RD calculation method was used to calculate the thermalization process of neutrons and the MC method was used to calculate the moderation process. The RD parameters were determined by MC calculations in advance. The MC‐RD calculation accuracy was verified by comparing the results of the MC‐RD and MC calculations with respect to the neutron flux distributions in each of the cubic and head phantoms filled with water.

Results

Comparing the MC‐RD calculation results with those of MC calculations, it was found that the MC‐RD calculation accurately reproduced the thermal neutron flux distribution inside the phantom, with the exception of the region near the surface of the phantom.

Conclusions

The MC‐RD calculation method is useful for the evaluation of the neutron flux distribution for the purpose of BNCT dose calculation, except for the region near the surface.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Boron Neutron Capture TherapyMonte Carlo MethodAlgorithmsNeutronsRadiotherapy Planning, Computer-AssistedDiffusionRadiotherapy DosagePhantoms, ImagingHumans

Stichwörter der Autor:innen

Monte Carlo methodboron neutron capture therapydose calculation algorithmremoval‐diffusiontreatment planning

Autor:innen und Institutionen

  • Department of Nuclear Engineering, Graduate School of Engineering, Kyoto University, Kyoto, Japan.
  • Institute for Integrated Radiation and Nuclear Science, Kyoto University, Osaka, Japan.
  • Kansai BNCT Medical Center, Osaka Medical and Pharmaceutical University, Takatsuki, Osaka, Japan.

Förderung

  • JSPS KAKENHI
  • JST SPRING

Zitieren

APA

Nojiri, M., Takata, T., Hu, N., Sakurai, Y., Suzuki, M., & Tanaka, H. (2024). Neutron flux evaluation algorithm with a combination of Monte Carlo and removal‐diffusion calculation methods for boron neutron capture therapy. Medical Physics. https://doi.org/10.1002/mp.16931

Vancouver

Nojiri M, Takata T, Hu N, Sakurai Y, Suzuki M, Tanaka H. Neutron flux evaluation algorithm with a combination of Monte Carlo and removal‐diffusion calculation methods for boron neutron capture therapy. Medical Physics. 2024. doi:10.1002/mp.16931

BibTeX
@article{Nojiri2024,
  author = {Nojiri, Mai and Takata, Takushi and Hu, Naonori and Sakurai, Yoshinori and Suzuki, Minoru and Tanaka, Hiroki},
  title = {Neutron flux evaluation algorithm with a combination of Monte Carlo and removal‐diffusion calculation methods for boron neutron capture therapy},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.16931}
}
Infokarte | EUVIMED