EUVIMEDDescobrir EUVIMED
Open access · CC BY 4.0

An anthropomorphic thyroid phantom for ultrasound‐guided radiofrequency ablation of nodules

Tim Boers, Wyger Brink, Leonardo Bianchi, Paola Saccomandi, Johan van Hespen, Germen Wennemars, Sicco Braak, Michel Versluis, Srirang Manohar

Medical Physics · 2024-02 · 51 (2) · 826-838

DOI: 10.1002/mp.16906 ↗

At a glance

Journal article in Medical Physics, published online on 2023-12-23. Written by Tim Boers and 8 co-authors. Institution: Multi-Modality Medical Imaging group, TechMed Centre, University of Twente, Enschede, the Netherlands. Major MEDLINE topics: Thyroid Nodule, Catheter Ablation. Cited 15 times so far (Crossref); 51 references. The full text is freely available (open access, CC BY 4.0). Funded by: ZGT Wetenschapsfonds and the Cooperation of Medical Specialists U.A. of the ZGT Hospital, Netherlands Organisation for Scientific Research (NWO)/the Netherlands Organisation for Health Research and Development (ZonMw), European Research Council (ERC), Horizon 2020 research and innovation program.

ResumoSubjectsDados bibliográficosCiteRelated studies

Abstract (original language)

Background

Needle‐based procedures, such as fine needle aspiration and thermal ablation, are often applied for thyroid nodule diagnosis and therapeutic purposes, respectively. With blood vessels and nerves nearby, these procedures can pose risks in damaging surrounding critical structures.

Purpose

The development and validation of innovative strategies to manage these risks require a test object with well‐characterized physical properties. For this work, we focus on the application of ultrasound‐guided thermal radiofrequency ablation.

Methods

We have developed a single‐use anthropomorphic phantom mimicking the thyroid and surrounding anatomical and physiological structures that are relevant to ultrasound‐guided thermal ablation. The phantom was composed of a mixture of polyacrylamide, water, and egg white extract and was cast using molds in multiple steps. The thermal, acoustical, and electrical characteristics were experimentally validated. The ablation zones were analyzed via non‐destructive T 2 ‐weighted magnetic resonance imaging scans utilizing the relaxometry changes of coagulated egg albumen, and the temperature distribution was monitored using an array of fiber Bragg grating sensors.

Results

The physical properties of the phantom were verified both on ultrasound as well as in terms of the phantom response to thermal ablation. The final temperature achieved (92°C), the median percentage of the nodule ablated (82.1%), the median volume ablated outside the nodule (0.8 mL), and the median number of critical structures affected (0) were quantified.

Conclusion

An anthropomorphic phantom that can provide a realistic model for development and training in ultrasound‐guided needle‐based thermal interventions for thyroid nodules has been presented.

Subjects

MeSH terms (MEDLINE)

Thyroid NoduleCatheter AblationHumansPhantoms, ImagingUltrasonography, InterventionalTreatment Outcome

Author keywords

MRIpolyacrylamide gelradiofrequency ablationthyroid phantomultrasound

Authors and institutions

  • Multi-Modality Medical Imaging group, TechMed Centre, University of Twente, Enschede, the Netherlands.
  • Magnetic Detection and Imaging group, TechMed Centre, University of Twente, Enschede, the Netherlands.
  • Department of Mechanical Engineering, Politecnico di Milano, Milan, Italy.
  • Department of Radiology, Ziekenhuisgroep Twente, Almelo, the Netherlands.
  • Physics of Fluids group, TechMed Centre, University of Twente, Enschede, the Netherlands.

Funding

  • ZGT Wetenschapsfonds and the Cooperation of Medical Specialists U.A. of the ZGT Hospital
  • Netherlands Organisation for Scientific Research (NWO)/the Netherlands Organisation for Health Research and Development (ZonMw)
  • European Research Council (ERC), Horizon 2020 research and innovation program

Cite

APA

Boers, T., Brink, W., Bianchi, L., Saccomandi, P., van Hespen, J., Wennemars, G., Braak, S., Versluis, M., & Manohar, S. (2024). An anthropomorphic thyroid phantom for ultrasound‐guided radiofrequency ablation of nodules. Medical Physics, 51(2), 826-838. https://doi.org/10.1002/mp.16906

Vancouver

Boers T, Brink W, Bianchi L, Saccomandi P, van Hespen J, Wennemars G, et al. An anthropomorphic thyroid phantom for ultrasound‐guided radiofrequency ablation of nodules. Medical Physics. 2024;51(2):826-838. doi:10.1002/mp.16906

BibTeX
@article{Boers2024,
  author = {Boers, Tim and Brink, Wyger and Bianchi, Leonardo and Saccomandi, Paola and van Hespen, Johan and Wennemars, Germen and Braak, Sicco and Versluis, Michel and Manohar, Srirang},
  title = {An anthropomorphic thyroid phantom for ultrasound‐guided radiofrequency ablation of nodules},
  journal = {Medical Physics},
  year = {2024},
  volume = {51},
  number = {2},
  pages = {826--838},
  doi = {10.1002/mp.16906}
}
Start for free →