EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

An anthropomorphic thyroid phantom for ultrasound‐guided radiofrequency ablation of nodules

Tim Boers, Wyger Brink, Leonardo Bianchi, Paola Saccomandi, Johan van Hespen, Germen Wennemars, Sicco Braak, Michel Versluis, Srirang Manohar

Medical Physics · 2024-02 · 51 (2) · 826-838

DOI: 10.1002/mp.16906 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, online erschienen am 23.12.2023. Verfasst von Tim Boers und 8 weiteren Autor:innen. Beteiligte Einrichtung: Multi-Modality Medical Imaging group, TechMed Centre, University of Twente, Enschede, the Netherlands. Hauptthemen laut MEDLINE: Thyroid Nodule, Catheter Ablation. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: ZGT Wetenschapsfonds and the Cooperation of Medical Specialists U.A. of the ZGT Hospital, Netherlands Organisation for Scientific Research (NWO)/the Netherlands Organisation for Health Research and Development (ZonMw), European Research Council (ERC), Horizon 2020 research and innovation program.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Needle‐based procedures, such as fine needle aspiration and thermal ablation, are often applied for thyroid nodule diagnosis and therapeutic purposes, respectively. With blood vessels and nerves nearby, these procedures can pose risks in damaging surrounding critical structures.

Purpose

The development and validation of innovative strategies to manage these risks require a test object with well‐characterized physical properties. For this work, we focus on the application of ultrasound‐guided thermal radiofrequency ablation.

Methods

We have developed a single‐use anthropomorphic phantom mimicking the thyroid and surrounding anatomical and physiological structures that are relevant to ultrasound‐guided thermal ablation. The phantom was composed of a mixture of polyacrylamide, water, and egg white extract and was cast using molds in multiple steps. The thermal, acoustical, and electrical characteristics were experimentally validated. The ablation zones were analyzed via non‐destructive T 2 ‐weighted magnetic resonance imaging scans utilizing the relaxometry changes of coagulated egg albumen, and the temperature distribution was monitored using an array of fiber Bragg grating sensors.

Results

The physical properties of the phantom were verified both on ultrasound as well as in terms of the phantom response to thermal ablation. The final temperature achieved (92°C), the median percentage of the nodule ablated (82.1%), the median volume ablated outside the nodule (0.8 mL), and the median number of critical structures affected (0) were quantified.

Conclusion

An anthropomorphic phantom that can provide a realistic model for development and training in ultrasound‐guided needle‐based thermal interventions for thyroid nodules has been presented.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Thyroid NoduleCatheter AblationHumansPhantoms, ImagingUltrasonography, InterventionalTreatment Outcome

Stichwörter der Autor:innen

MRIpolyacrylamide gelradiofrequency ablationthyroid phantomultrasound

Autor:innen und Institutionen

  • Multi-Modality Medical Imaging group, TechMed Centre, University of Twente, Enschede, the Netherlands.
  • Magnetic Detection and Imaging group, TechMed Centre, University of Twente, Enschede, the Netherlands.
  • Department of Mechanical Engineering, Politecnico di Milano, Milan, Italy.
  • Department of Radiology, Ziekenhuisgroep Twente, Almelo, the Netherlands.
  • Physics of Fluids group, TechMed Centre, University of Twente, Enschede, the Netherlands.

Förderung

  • ZGT Wetenschapsfonds and the Cooperation of Medical Specialists U.A. of the ZGT Hospital
  • Netherlands Organisation for Scientific Research (NWO)/the Netherlands Organisation for Health Research and Development (ZonMw)
  • European Research Council (ERC), Horizon 2020 research and innovation program

Zitieren

APA

Boers, T., Brink, W., Bianchi, L., Saccomandi, P., van Hespen, J., Wennemars, G., Braak, S., Versluis, M., & Manohar, S. (2024). An anthropomorphic thyroid phantom for ultrasound‐guided radiofrequency ablation of nodules. Medical Physics, 51(2), 826-838. https://doi.org/10.1002/mp.16906

Vancouver

Boers T, Brink W, Bianchi L, Saccomandi P, van Hespen J, Wennemars G, et al. An anthropomorphic thyroid phantom for ultrasound‐guided radiofrequency ablation of nodules. Medical Physics. 2024;51(2):826-838. doi:10.1002/mp.16906

BibTeX
@article{Boers2024,
  author = {Boers, Tim and Brink, Wyger and Bianchi, Leonardo and Saccomandi, Paola and van Hespen, Johan and Wennemars, Germen and Braak, Sicco and Versluis, Michel and Manohar, Srirang},
  title = {An anthropomorphic thyroid phantom for ultrasound‐guided radiofrequency ablation of nodules},
  journal = {Medical Physics},
  year = {2024},
  volume = {51},
  number = {2},
  pages = {826--838},
  doi = {10.1002/mp.16906}
}
Kostenlos starten →