EUVIMEDObjevte EUVIMED
Open access · CC BY-NC 4.0

Joint AAPM Task Group 282/EFOMP Working Group Report: Breast dosimetry for standard and contrast‐enhanced mammography and breast tomosynthesis

Ioannis Sechopoulos iD, David R Dance, John M Boone, Hilde T Bosmans, Marco Caballo, Oliver Diaz, Ruben van Engen, Christian Fedon, Stephen J Glick, Andrew M Hernandez, Melissa L Hill, Katie W Hulme and 4 more

Medical Physics · 2024-02 · 51 (2) · 712-739

DOI: 10.1002/mp.16842 ↗

At a glance

Journal article in Medical Physics, published online on 2023-11-29. Written by Ioannis Sechopoulos and 15 co-authors. Institution: Radboud University Medical Center, Nijmegen, The Netherlands. Major MEDLINE topics: Breast, Breast Neoplasms. Cited 33 times so far (Crossref); 36 references. The full text is freely available (open access, CC BY-NC 4.0).

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Currently, there are multiple breast dosimetry estimation methods for mammography and its variants in use throughout the world. This fact alone introduces uncertainty, since it is often impossible to distinguish which model is internally used by a specific imaging system. In addition, all current models are hampered by various limitations, in terms of overly simplified models of the breast and its composition, as well as simplistic models of the imaging system. Many of these simplifications were necessary, for the most part, due to the need to limit the computational cost of obtaining the required dose conversion coefficients decades ago, when these models were first implemented. With the advancements in computational power, and to address most of the known limitations of previous breast dosimetry methods, a new breast dosimetry method, based on new breast models, has been developed, implemented, and tested. This model, developed jointly by the American Association of Physicists in Medicine and the European Federation for Organizations of Medical Physics, is applicable to standard mammography, digital breast tomosynthesis, and their contrast‐enhanced variants. In addition, it includes models of the breast in both the cranio‐caudal and the medio‐lateral oblique views. Special emphasis was placed on the breast and system models used being based on evidence, either by analysis of large sets of patient data or by performing measurements on imaging devices from a range of manufacturers. Due to the vast number of dose conversion coefficients resulting from the developed model, and the relative complexity of the calculations needed to apply it, a software program has been made available for download or online use, free of charge, to apply the developed breast dosimetry method. The program is available for download or it can be used directly online. A separate User's Guide is provided with the software.

Subjects

MeSH terms (MEDLINE)

BreastBreast NeoplasmsHumansFemaleMammographyRadiometryMonte Carlo Method

Author keywords

dosimetrymammographytomosynthesis

Authors and institutions

  • Radboud University Medical Center, Nijmegen, The Netherlands.
  • Dutch Expert Centre for Screening (LRCB), Nijmegen, The Netherlands.
  • University of Twente, Enschede, The Netherlands.
  • National Co-ordinating Centre for the Physics of Mammography (NCCPM), Royal Surrey County Hospital, Guildford, UK.
  • University of California, Davis, California, USA.
  • KU Leuven, Leuven, Belgium.

ORCID profiles: Ioannis Sechopoulos

Cite

APA

Sechopoulos, I., Dance, D. R., Boone, J. M., Bosmans, H. T., Caballo, M., Diaz, O., van Engen, R., Fedon, C., Glick, S. J., Hernandez, A. M., Hill, M. L., Hulme, K. W., Longo, R., Rabin, C., Sanderink, W. B. G., & Seibert, J. A. (2024). Joint AAPM Task Group 282/EFOMP Working Group Report: Breast dosimetry for standard and contrast‐enhanced mammography and breast tomosynthesis. Medical Physics, 51(2), 712-739. https://doi.org/10.1002/mp.16842

Vancouver

Sechopoulos I, Dance DR, Boone JM, Bosmans HT, Caballo M, Diaz O, et al. Joint AAPM Task Group 282/EFOMP Working Group Report: Breast dosimetry for standard and contrast‐enhanced mammography and breast tomosynthesis. Medical Physics. 2024;51(2):712-739. doi:10.1002/mp.16842

BibTeX
@article{Sechopoulos2024,
  author = {Sechopoulos, Ioannis and Dance, David R and Boone, John M and Bosmans, Hilde T and Caballo, Marco and Diaz, Oliver and van Engen, Ruben and Fedon, Christian and Glick, Stephen J and Hernandez, Andrew M and Hill, Melissa L and Hulme, Katie W and Longo, Renata and Rabin, Carolina and Sanderink, Wendelien B G and Seibert, J Anthony},
  title = {Joint AAPM Task Group 282/EFOMP Working Group Report: Breast dosimetry for standard and contrast‐enhanced mammography and breast tomosynthesis},
  journal = {Medical Physics},
  year = {2024},
  volume = {51},
  number = {2},
  pages = {712--739},
  doi = {10.1002/mp.16842}
}
Start for free →