Estimating follow‐up CTs from geometric deformations of catheter implants in interstitial breast brachytherapy: A feasibility study using electromagnetic tracking

Christopher Dürrbeck, Moritz Schulz, Leonie Pflaum, Karoline Kallis, Tobias Geimer, Nadin Abu-Hossin, Vratislav Strnad, Andreas Maier, Rainer Fietkau, Christoph Bert

Medical Physics · 2023

DOI: 10.1002/mp.16659 ↗

At a glance

Journal article in Medical Physics, 2023. Written by Christopher Dürrbeck and 9 co-authors. Institution: Department of Radiation Oncology, Universitätsklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany. Major MEDLINE topics: Brachytherapy. Cited 1 times so far (Crossref); 42 references. The full text is freely available (open access, CC BY 4.0).

AbstractSubjectsDati bibliograficiCiteRelated studies

Abstract (original language)

Background

Electromagnetic tracking (EMT) systems have been shown to provide valuable information on the geometry of catheter implants in breast cancer patients undergoing interstitial brachytherapy (iBT). In the context of an extended patient‐specific, pre‐treatment verification, EMT can play a key role in determining the potential need and, if applicable, the appropriate time for treatment adaptation. To detect dosimetric shortcomings the relative position between catheters, and target volume and critical structures must be known. Since EMT cannot provide the anatomical context and standard imaging techniques such as cone‐beam CT are not yet available in most brachytherapy suites, it is not possible to detect anatomic changes on a daily or fraction basis, so the need for adaptive planning cannot be identified.

Purpose

The aim of this feasibility study is to develop and evaluate a technique capable of estimating follow‐up CTs at any time based on the initial treatment planning CT (PCT) and surrogate information about changes of the implant geometry from an EMT system.

Methods

A deformation vector field is calculated from two different implant reconstructions acquired in treatment position through EMT, the first immediately after the PCT and the second at another time point during the course of treatment. The calculation is based on discrete displacement vectors of pairs of control and target points. These are extrapolated by means of different radial basis functions in order to cover the entire CT volume. The adequate parameters for the calculation of the deformation field were identified. By warping the PCT according to the deformation field, one obtains an estimated CT (ECT) that reflects the geometric changes. For the proof of concept, ECTs were computed for the time point of the clinical follow‐up CT (FCT) that is embedded in the treatment workflow after the fourth fraction.

Results

ECT and clinical FCTs of 20 patients were compared to each other quantitatively in terms of absolute Hounsfield unit differences in the planning target volume (PTV) and in a convex hull (CH) enclosing the catheters. The median differences were 31.2 and 29.5 HU for the CH and the PTV, respectively.

Conclusion

The proposed ECT approach was able to approximate the “anatomy of the day” and therefore, in principle, allows a dosimetric appraisal of the treatment plan quality before each fraction. In this way, it can contribute to a more detailed patient‐specific quality assurance in iBT of the breast and help to identify the timing for a potential treatment adaptation.

Subjects

MeSH terms (MEDLINE)

BrachytherapyHumansFeasibility StudiesFollow-Up StudiesRadiotherapy Planning, Computer-AssistedRadiotherapy DosageTomography, X-Ray ComputedCathetersElectromagnetic Phenomena

Author keywords

assurancebreast cancerelectromagnetic trackinginterstitial brachytherapy

Authors and institutions

  • Department of Radiation Oncology, Universitätsklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
  • Comprehensive Cancer Center Erlangen-EMN (CCC ER-EMN), Erlangen, Germany.
  • Pattern Recognition Lab, FAU, Erlangen, Germany.

Cite

APA

Dürrbeck, C., Schulz, M., Pflaum, L., Kallis, K., Geimer, T., Abu-Hossin, N., Strnad, V., Maier, A., Fietkau, R., & Bert, C. (2023). Estimating follow‐up CTs from geometric deformations of catheter implants in interstitial breast brachytherapy: A feasibility study using electromagnetic tracking. Medical Physics. https://doi.org/10.1002/mp.16659

Vancouver

Dürrbeck C, Schulz M, Pflaum L, Kallis K, Geimer T, Abu-Hossin N, et al. Estimating follow‐up CTs from geometric deformations of catheter implants in interstitial breast brachytherapy: A feasibility study using electromagnetic tracking. Medical Physics. 2023. doi:10.1002/mp.16659

BibTeX
@article{Drrbeck2023,
  author = {Dürrbeck, Christopher and Schulz, Moritz and Pflaum, Leonie and Kallis, Karoline and Geimer, Tobias and Abu-Hossin, Nadin and Strnad, Vratislav and Maier, Andreas and Fietkau, Rainer and Bert, Christoph},
  title = {Estimating follow‐up CTs from geometric deformations of catheter implants in interstitial breast brachytherapy: A feasibility study using electromagnetic tracking},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16659}
}
Start for free →