Bony structure enhanced synthetic CT generation using Dixon sequences for pelvis MR‐only radiotherapy
Medical Physics · 2023-12 · 50 (12) · 7368-7382
DOI: 10.1002/mp.16556 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, online erschienen am 26.06.2023. Verfasst von Xiao Liang und 15 weiteren Autor:innen. Beteiligte Einrichtung: Medical Artificial Intelligence and Automation Laboratory and Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dall…. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background
MRI‐only radiotherapy planning (MROP) is beneficial to patients by avoiding MRI/CT registration errors, simplifying the radiation treatment simulation workflow and reducing exposure to ionizing radiation. MRI is the primary imaging modality for soft tissue delineation. Treatment planning CTs (i.e., CT simulation scan) are redundant if a synthetic CT (sCT) can be generated from the MRI to provide the patient positioning and electron density information. Unsupervised deep learning (DL) models like CycleGAN are widely used in MR‐to‐sCT conversion, when paired patient CT and MR image datasets are not available for model training. However, compared to supervised DL models, they cannot guarantee anatomic consistency, especially around bone.
Purpose
The purpose of this work was to improve the sCT accuracy generated from MRI around bone for MROP.
Methods
To generate more reliable bony structures on sCT images, we proposed to add bony structure constraints in the unsupervised CycleGAN model's loss function and leverage Dixon constructed fat and in‐phase (IP) MR images. Dixon images provide better bone contrast than T2‐weighted images as inputs to a modified multi‐channel CycleGAN. A private dataset with a total of 31 prostate cancer patients were used for training (20) and testing (11).
Results
We compared model performance with and without bony structure constraints using single‐ and multi‐channel inputs. Among all the models, multi‐channel CycleGAN with bony structure constraints had the lowest mean absolute error, both inside the bone and whole body (50.7 and 145.2 HU). This approach also resulted in the highest Dice similarity coefficient (0.88) of all bony structures compared with the planning CT.
Conclusion
Modified multi‐channel CycleGAN with bony structure constraints, taking Dixon‐constructed fat and IP images as inputs, can generate clinically suitable sCT images in both bone and soft tissue. The generated sCT images have the potential to be used for accurate dose calculation and patient positioning in MROP radiation therapy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Medical Artificial Intelligence and Automation Laboratory and Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, Texas, USA.
Zitieren
APA
Liang, X., Yen, A., Bai, T., Godley, A., Shen, C., Wu, J., Meng, B., Lin, M. H., Medin, P., Yan, Y., Owrangi, A., Desai, N., Hannan, R., Garant, A., Jiang, S., & Deng, J. (2023). Bony structure enhanced synthetic CT generation using Dixon sequences for pelvis MR‐only radiotherapy. Medical Physics, 50(12), 7368-7382. https://doi.org/10.1002/mp.16556
Vancouver
Liang X, Yen A, Bai T, Godley A, Shen C, Wu J, et al. Bony structure enhanced synthetic CT generation using Dixon sequences for pelvis MR‐only radiotherapy. Medical Physics. 2023;50(12):7368-7382. doi:10.1002/mp.16556
BibTeX
@article{Liang2023,
author = {Liang, Xiao and Yen, Allen and Bai, Ti and Godley, Andrew and Shen, Chenyang and Wu, Junjie and Meng, Boyu and Lin, Mu-Han and Medin, Paul and Yan, Yulong and Owrangi, Amir and Desai, Neil and Hannan, Raquibul and Garant, Aurelie and Jiang, Steve and Deng, Jie},
title = {Bony structure enhanced synthetic CT generation using Dixon sequences for pelvis MR‐only radiotherapy},
journal = {Medical Physics},
year = {2023},
volume = {50},
number = {12},
pages = {7368--7382},
doi = {10.1002/mp.16556}
}
EUVIMED