Evaluation of repeatability and reproducibility of radiomic features produced by the fan‐beam kV‐CT on a novel ring gantry‐based PET/CT linear accelerator
Medical Physics · 2023
DOI: 10.1002/mp.16399 ↗At a glance
Journal article in Medical Physics, 2023. Written by Trevor Ketcherside and 11 co-authors. Institution: City of Hope National Medical Center, Duarte, California, USA. Major MEDLINE topics: Positron Emission Tomography Computed Tomography, Image Processing, Computer-Assisted. Cited 5 times so far (Crossref); 35 references. The full text is freely available (open access, CC BY 4.0).
Abstract (original language)
Background
The RefleXion X1 is a novel radiotherapy delivery system on a ring gantry equipped with fan‐beam kV‐CT and PET imaging subsystems. The day‐to‐day scanning variability of radiomics features must be evaluated before any attempt to utilize radiomics features.
Purpose
This study aims to characterize the repeatability and reproducibility of radiomic features produced by the RefleXion X1 kV‐CT. Materials and Methods The Credence Cartridge Radiomics (CCR) phantom includes six cartridges of varied materials. It was scanned 10 times on the RefleXion X1 kVCT imaging subsystem over a 3‐month period using the two most frequently used scanning protocols (BMS and BMF). Fifty‐five radiomic features were extracted for each ROI on each CT scan and analyzed using LifeX software. The coefficient of variation (COV) was computed to evaluate the repeatability. Intraclass correlation coefficient (ICC) and concordance correlation coefficient (CCC) were used to evaluate the repeatability and reproducibility of the scanned images using 0.9 as the threshold. This process is repeated on a GE PET‐CT scanner using several built‐in protocols as a comparison.
Results
On average, 87% of the features on both scan protocols on the RefleXion X1 kVCT imaging subsystem can be considered repeatable as they met COV < 10% criteria. On GE PET‐CT, this number is similar at 86%. When we tighten the criteria to COV 0.9 respectively for BMS and BMF protocols on RefleXion X1. On the other hand, the percentage of features with ICC > 0.9 on GE PET‐CT ranges from 67% to 82%. The RefleXion X1 kVCT imaging subsystem showed excellent intra‐scanner reproducibility between the scanning protocols much better than the GE PET CT scanner. For the inter‐scanner reproducibility, the percentage of features with CCC > 0.9 ranged from 49% to 80%. between X1 and GE PET‐CT scanning protocols.
Conclusions
Clinically useful CT radiomic features produced by the RefleXion X1 kVCT imaging subsystem are reproducible and stable over time, demonstrating its utility as a quantitative imaging platform.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- City of Hope National Medical Center, Duarte, California, USA.
- Reflexion Medical Inc., Hayward, California, USA.
- UT MD Anderson Cancer Center, Houston, California, USA.
Cite
APA
Ketcherside, T., Shi, C., Chen, Q., Leung, D., Sundquist, A., Huntzinger, C., Court, L. E., Han, C., Watkins, T., Ladbury, C., Williams, T. M., & Liu, A. (2023). Evaluation of repeatability and reproducibility of radiomic features produced by the fan‐beam kV‐CT on a novel ring gantry‐based PET/CT linear accelerator. Medical Physics. https://doi.org/10.1002/mp.16399
Vancouver
Ketcherside T, Shi C, Chen Q, Leung D, Sundquist A, Huntzinger C, et al. Evaluation of repeatability and reproducibility of radiomic features produced by the fan‐beam kV‐CT on a novel ring gantry‐based PET/CT linear accelerator. Medical Physics. 2023. doi:10.1002/mp.16399
BibTeX
@article{Ketcherside2023,
author = {Ketcherside, Trevor and Shi, Chengyu and Chen, Quan and Leung, David and Sundquist, Andrew and Huntzinger, Calvin and Court, Laurence E and Han, Chunhui and Watkins, Tyler and Ladbury, Colton and Williams, Terence M and Liu, An},
title = {Evaluation of repeatability and reproducibility of radiomic features produced by the fan‐beam kV‐CT on a novel ring gantry‐based PET/CT linear accelerator},
journal = {Medical Physics},
year = {2023},
doi = {10.1002/mp.16399}
}
EUVIMED